Thromb Haemost 1980; 43(01): 053-057
DOI: 10.1055/s-0038-1650011
DOI: 10.1055/s-0038-1650011
Original Article
Simplified Thromboplastin Calibration: Further Experience of the Dutch Reference Laboratory for Anticoagulant Control
Further Information
Publication History
Received 21 January 1980
Accepted 24 January 1980
Publication Date:
13 July 2018 (online)


Summary
The modified (and thereby simplified) thromboplastin calibration procedure using pooled lyophilized normal and patient plasmas is shown to equal, for batch-to-batch control of Thrombotest, the calibration procedure described by Biggs and Denson, which makes use of individual freshly prepared patient plasmas.