Endosc Int Open 2016; 04(08): E922-E926
DOI: 10.1055/s-0042-111905
Original article
© Georg Thieme Verlag KG Stuttgart · New York

A novel mechanical simulator for cannulation and sphincterotomy after Billroth II or Roux-en-Y reconstruction

Eckart Frimberger
1   II. Medizinische Klinik und Poliklinik, Klinikum rechts der Isar der Technischen Universität München, Munich, Germany
,
Mohamed Abdelhafez
1   II. Medizinische Klinik und Poliklinik, Klinikum rechts der Isar der Technischen Universität München, Munich, Germany
2   Gastrointestinal Endoscopy Unit, Kasr Alainy Hospital, Cairo University, Cairo, Egypt
,
Roland M. Schmid
1   II. Medizinische Klinik und Poliklinik, Klinikum rechts der Isar der Technischen Universität München, Munich, Germany
,
Stefan von Delius
1   II. Medizinische Klinik und Poliklinik, Klinikum rechts der Isar der Technischen Universität München, Munich, Germany
› Institutsangaben
Weitere Informationen

Publikationsverlauf

submitted 31. Dezember 2015

accepted after revision 17. Mai 2016

Publikationsdatum:
08. August 2016 (online)

Zoom Image

Introduction: In patients with Billroth II (B II) or Roux-en-Y anatomy, endoscopic retrograde cholangiopancreatography (ERCP) is demanding. Here, we describe a novel simulator with simulated fluoroscopy for cannulation and sphincterotomy training in such situations.

Methods: A custom-made simulation system was built based upon a common chassis of a series of previously described ERCP simulators. The papilla is made out of organic material and can be cut by high frequency current. The advancement of guidewires and other instruments within transparent mock bile ducts can be viewed in the window of the simulator without the need for fluoroscopy. The ERCP B II/Roux-en-Y simulation system was first evaluated during an ERCP course.

Results: There were no technical problems related to the novel simulator during the course. After sphincterotomy, the organic papillae could easily be exchanged within a few seconds. Overall, the novel B II/Roux-en-Y simulator achieved favorable results by trainees and expert endoscopists in all categories assessed.

Conclusions: The new B II/Roux-en-Y mechanical simulator is simple and practicable. A first evaluation during an ERCP course showed promising results.