Background and Study Aims: This paper presents the results of a comparison between two different three-dimensional (3-D) video systems, one with single-channel optics, the other with bi-channel optics. The latter integrates two lens systems, each transferring one half of the stereoscopic image; the former uses only one lens system, similar to a two-dimensional (2-D) endoscope, which transfers the complete stereoscopic picture.
Material and Methods: In our training centre for minimally invasive surgery, surgeons were involved in basic and advanced laparoscopic courses using both a 2-D system and the two 3-D video systems. They completed analog scale questionnaires in order to record a subjective impression of the relative convenience of operating in 2-D and 3-D vision, and to identify perceived deficiencies in the 3-D system. As an objective test, different experimental tasks were developed, in order to measure performance times and to count pre-defined errors made while using the two 3-D video systems and the 2-D system.
Results and Conclusion: Using the bi-channel optical system, the surgeon has a heightened spatial perception, and can work faster and more safely than with a single-channel system. However, single-channel optics allow the use of an angulated endoscope, and the free rotation of the optics relative to the camera, which is necessary for some operative applications.
References
-
1
Becker H, Melzer A, Schurr M O, Buess G F.
3-D video techniques in endoscopic surgery.
Endosc Surg.
1992;
1
40-46
-
2
Buess G F, van Bergen P, Kunert W, Schurr M O.
Vergleichsstudie verschiedener 2-D und 3-D Sichtsysteme in der minimal invasiven Chirurgie.
Chirurg.
1996;
67
1041-1046
-
3
van Bergen P, Kunert W, Schurr M O, Buess G F.
Vergleichsstudie Endoskopischer 2-D und 3-D Sichtsysteme.
Langenbecks Archiv für Chirurgie.
1996;
II
634-637
-
4
van Bergen P, Kunert W, Buess G F.
Comparative study of different 2-D and 3-D vision systems for minimally invasive surgery.
Surg Endosc.
1998;
12
948-954
-
5
Birkett D H, Josephs L G, Este-McDonald J.
A new 3-D laparoscope in gastrointestinal surgery.
Surg Endosc.
1994;
8
1448-1451
-
6
Durrani A F, Preminger G M.
Advanced endoscopic imaging: 3-D laparoscopic endoscopy.
Surg Tech Int.
1994;
III
141-147
-
7
Hanna G B, Shimi S M, Cuschieri A.
Randomised study of influence of two-dimensional versus three-dimensional imaging on performance of laparoscopic cholecystectomy.
Lancet.
1998;
351
248-251
-
8
Jones D B, Brewer J D, Soper R N, Soper N J.
The influence of three-dimensional video systems on laparoscopic task performance.
Surg Laparosc Endosc.
1996;
6
191-197
-
9
McDougall E M, Soble J J, Wolf J S, et al.
Comparison of three-dimensional and two-dimensional laparoscopic video systems.
J Endourol.
1996;
10
371-374
-
10
Mitchell T N, Robertson J, Nagy A G, Lomax A.
Three-dimensional endoscopic imaging for minimal access surgery.
J R Coll Surg Edinburgh.
1993;
38
285-292
-
11
Peitgen K, Walz M V, Holtmann G, et al.
A prospective randomized experimental evaluation of three-dimensional imaging in laparoscopy.
Gastrointest Endosc.
1996;
44/3
262-267
-
12
Vry U, Krause-Bonte M.
3D-Endoskopie in der Chirurgie, aufrecht und dreidimensional in jeder Blickrichtung.
Zeiss Information mit Jenaer Rundschau 4.
1995;
6
6-8
-
13
Schwind Herbert GmbH, KG.
Optische Geräte Betriebshandbuch.
TITMUS.
1985;
II
15
-
14
Kunert W, Flemming E, Schurr M O, Buess G F.
Optik mit natürlicher Zusatzbeleuchtung.
Langenbecks Archiv für Chirurgie.
1997;
II
1232-1234
P. van BergenM.D.
Section for Minimally Invasive Surgery
Eberhard-Karls-Universität Tübingen
Waldhoernlestrasse 22
72072 Tübingen
Germany
Phone: + 49-7071-295569
Email: patrick@van-bergen.de