Validity and Reliability of the Axiom Powertrain Cycle Ergometer When Compared with an SRM Powermeter

W. Bertucci1 , S. Duc1 , V. Villerius1 , F. Grappe1
  • 1Laboratoire de Mécanique Appliquée (U.M.R. C.N.R.S. 6604), Université de Franche Comté, Besançon, France
Weitere Informationen

Publikationsverlauf

Accepted after revision: January 6, 2004

Publikationsdatum:
30. Juli 2004 (online)

Abstract

The purpose of this study was to determine the validity and the reliability of a stationary electromagnetically-braked cycle ergometer (Axiom PowerTrain) against the SRM power measuring crankset. Nineteen male competitive cyclists completed four tests on their bicycle equipped with a 20-strain gauges SRM crankset: a maximal aerobic power (MAP) test and three 10-min time trials (TTs) with three different simulated slopes (0, 3, and 6 %). The Axiom ergometer overestimated (p < 0.05) the SRM power output during the last stage of the MAP test and during TTs, by 5 % and 12 %, respectively. Power output (PO) of the Axiom ergometer drifted significantly (p < 0.05) with the time during TT. These findings indicate that the Axiom ergometer does not provide a valid measure of PO compared with SRM. However, the small coefficient of variation (2.2 %) during the MAP test indicates that the Axiom provides a reliable PO and that it can be used e.g. for relative PO comparisons with competitive cyclists during a race season.

References

  • 1 Atkinson G, Nevill A M. Statistical methods for assessing measurement error (reliability) in variables relevant to sports medicine.  Sports Med. 1998;  26 217-238
  • 2 Balmer J, Davison R C, Coleman D A, Bird S R. The validity of power output recorded during exercise performance tests using a Kingcycle air-braked cycle ergometer when compared with an SRM powermeter.  Int J Sports Med. 2000;  21 195-199
  • 3 Balmer J, Davison R C, Bird S R. Reliability of an air-braked ergometer to record peak power during a maximal cycling test.  Med Sci Sports Exerc. 2000;  32 1790-1793
  • 4 Bland J M, Altman D G. Statistical methods for assessing agreement between two methods of clinical measurement.  Lancet. 1986;  1 307-310
  • 5 Candau B, Grappe F, Ménard M, Barbier B, Millet G Y, Hoffman M, Belli A, Rouillon J D. Simplified deceleration method for assessment of resistive forces in cycling.  Med Sci Sports Exerc. 1999;  31 1441-1447
  • 6 Coyle E F, Feltner M E, Kautz S A, Hamilton M, Montain S, Baylor A, Abraham L, Petrek G. Physiological and biomechanical factors associated with elite endurance cycling performance.  Med Sci Sports Exerc. 1991;  23 93-107
  • 7 Finn J P, Maxwell R, Withers R T. Air-braked cycle ergometers: Validity of the correction factor for barometric pressure.  Int J Sports Med. 2000;  21 488-491
  • 8 Grappe F, Candau R, Barbier B, Hoffman M, Belli A, Rouillon J D. Influence of tyre pressure and vertical load on coefficient of rolling resistance and simulated cycling performance.  Ergonomics. 1999;  10 1361-1371
  • 9 Hickey M S, Costill D L, McConee G K, Widrick J J, Tanaka H. Day to day variation in time trial cycling performance.  Int J Sports Med. 1992;  13 467-470
  • 10 Hopkins W G, Hawley J A, Burke L M. Design and analysis of research on sport performance enhancement.  Med Sci Sports Exerc. 1999;  31 472-485
  • 11 Jones S M, Passfield L. Dynamic calibration of bicycle power measuring cranks. Haake SJ The Engineering of Sport. Oxford; Blackwell Science 1998: 265-274
  • 12 Lawton E W, Martin D T, Lee H. Validation of SRM Power Cranks Using Dynamic Calibration. Fifth IOC World Congress, Oct 31 - Nov 5, 1999. Sydney; International Olympic Committee 1999
  • 13 Martin J C, Milliken D L, Cobb J E, McFadden K L, Coggan A R. Validation of a mathematical model for road cycling power.  J Appl Biomech. 1998;  14 276-291
  • 14 Millet G P, Tronche C, Fuster N, Bentley D J, Candau R. Validity and reliability of the Polar®S710 mobile cycling powermeter.  Int J Sports Med. 2003;  24 156-161
  • 15 Nevill A. Why the analysis of performance variables recorded on a ratio scale will invariably benefit from a log transformation.  J Sports Sci. 1997;  15 457-458
  • 16 Nevill A M, Atkinson G. Assessing agreement between measurements recorded on a ratio scale in sports medicine and sports science.  Br J Sports Med. 1997;  31 314-318
  • 17 Paton C D, Hopkins W G. Tests of cycling performance.  Sports Med. 2001;  31 489-496
  • 18 Perrey S, Grappe F, Girard A, Bringard A, Groslambert A, Bertucci W, Rouillon J D. Physiological and metabolic responses of triathletes to a simulated 30 min time trial in cycling at self-selected intensity.  Int J Sports Med. 2003;  24 138-143
  • 19 Reiser M, Meyer T, Kindermann W, Daugs R. Transferability of workload measurements between three different types of ergometer.  Eur J Appl Physiol. 2000;  82 245-249
  • 20 Vandewalle H, Peres G, Heller J, Panel J, Monod H. Force-velocity relationship and maximal power on a cycle ergometer.  Eur J Appl Physiol. 1987;  56 650-656
  • 21 Van Praagh E, Bedu M, Roddier P, Coudert J. A simple calibration method for mechanically braked cycle ergometers.  Int J Sports Med. 1992;  13 27-30

W. Bertucci

Laboratoire des Sciences du Sport

Place Saint-Jacques

25000 Besançon

France

Telefon: + 33381665681

Fax: + 33 3 81 66 56 92

eMail: wil.ber@oreka.com