Abstract
We compared cardiac output (CO) determined simultaneously by electrical impedance
cardiography method (COICG) and by the CO2 rebreathing technique (CO2REB) during three different steady-state exercises (target heart rate of 120, 140,
and 160 min-1) in 8 healthy fit young men. The mean difference correlation coefficient obtained
between the values of COICG and CO2REB was 0.85 and the mean difference (COICG-CO2REB) was 0.06 l/min (0.12 %). At 120 min-1, COICG was lower than CO2REB but the tendency was reversed at 140 and 160 min-1 where COICG was higher than CO2REB. This evolution may be explained by the difficultly of using CO2 rebreathing technique at the highest steady-state exercises and by the progressive
acidemia due to exercise. The present results suggest that electrical impedance
cardiography method provides acceptable evaluation of CO and may favourably replace
the CO2 rebreathing technique during mild (or moderate) to high steady-state exercises.
Key words
Impedance cardiography - CO2 rebreathing technique - indirect Fick principle - steady state exercise
References
- 1 Bland J M, Altman D G. Statistical methods for assessing agreement between two
methods of clinical measurement. Lancet 1986: 307-310
- 2
Bloch K E, Russi E W.
Comparison of impedance cardiography to invasive techniques for measurement
of cardiac output.
Am J Cardiol.
1997;
79
846
- 3
Charloux A, Lonsdorfer-Wolf E, Richard R, Lampert E, Oswald-Mammosser M, Mettauer B,
Geny B, Lonsdorfer J.
A new impedance cardiograph device for the non-invasive evaluation of cardiac
output at rest and during exercise; comparison with the ”direct“ Fick method.
Eur J Appl Physiol.
2000;
82
313-320
- 4
Christensen T B, Jensen B V, Hjerpe J, Kanstrup I L.
Cardiac output measured by electric bioimpedance compared with the CO2 rebreathing technique at different exercise levels.
Clin Physiol.
2000;
20
101-105
- 5
Defares J G.
Determination of PV·CO2 from the exponential CO2 rise during rebreathing.
J Appl Physiol.
1958;
13
159-164
- 6
Denniston J C, Maher J T, Reeves J T, Cruz J C, Cymerman A, Grover R F.
Measurement of cardiac output by electrical impedance at rest and during exercise.
J Appl Physiol.
1976;
40
91-95
- 7
Freund P R.
Transesophageal Doppler scanning versus thermodilution during general anesthesia.
An initial comparison of cardiac output techniques.
Am J Surg.
1987;
153
490-494
- 8
Jones N L, Robertson D G, Kane J W.
Difference between end-tidal and arterial PCO2 in exercise.
J Appl Physiol.
1979;
47
954-960
- 9
Gluer C C, Blake G, Lu Y, Blunt A, Jergas M, Genant H K.
Accurate assessment of precision errors: How to measure the reproducibility
of bone densitometry techniques.
Osteoporosis Int.
1995;
5
262-270
- 10
Kubicek W G, Karnegis J N, Patterson R P, Witsoe D A, Mattson R H.
Development and evaluation of an impedance cardiac output system.
Aerosp Med.
1966;
37
1208-1212
- 11
Lefrant J Y, Benbabaali M, Ripart J, Aya A G, Sassi G, Dauzat M, de la Coussaye J E,
Eledjam J J.
CO assessment by suprasternal Doppler in critically ill patients: comparison
with thermodilution.
Intensive Care Med.
2000;
26
693-697
- 12
Marik P E, Pendelton J E, Smith R.
A comparison of hemodynamic parameters derived from transthoracic electrical
bioimpedance with those parameters obtained by thermodilution and ventricular
angiography.
Crit Care Med.
1997;
25
1545-1550
- 13
Moore R, Sansores R, Guimond V, Abboud R.
Evaluation of cardiac output by thoracic electrical bioimpedance during exercise
in normal subjects.
Chest.
1992;
102
448-455
- 14
Newman D G, Callister R.
The non-invasive assessment of stroke volume and cardiac output by impedance
cardiography: a review.
Aviat Space Environ Med.
1999;
70
780-789
- 15
Nugent A M, McParland J, McEneaney D J, Steele I, Campbell N P, Stanford C F, Nicholls D P.
Non-invasive measurement of cardiac output by a carbon dioxide rebreathing method
at rest and during exercise.
Eur Heart J.
1994;
15
361-368
- 16
Pianosi P, Garros D.
Comparison of impedance cardiography with indirect Fick (CO2) method of measuring cardiac output in healthy children during exercise.
Am J Cardiol.
1996;
77
745-749
- 17
Pickett B R, Buell J C.
Validity of cardiac output measurement by computer-averaged impedance cardiography,
and comparison with simultaneous thermodilution determinations.
Am J Cardiol.
1992;
69
1354-1358
- 18
Richard R, Lonsdorfer-Wolf E, Charloux A, Doutreleau S, Buchheit M, Oswald-Mammosser M,
Lampert E, Mettauer B, Geny B, Lonsdorfer J.
Non-invasive cardiac output evaluation during a maximal progressive exercise
test, using a new impedance cardiograph device.
Eur J Appl Physiol.
2001;
85
202-207
- 19
Rose J S, Nanna M, Rahimtoola S H, Elkayam U, McKay C, Chandraratna P A.
Accuracy of determination of changes in cardiac output by transcutaneous continuous-wave
Doppler computer.
Am J Cardiol.
1984;
54
1099-1101
- 20
Royse C F, Royse A G, Blake D W, Grigg L E.
Measurement of cardiac output by transoesophageal echocardiography: a comparison
of two Doppler methods with thermodilution.
Anaesth Intens Care.
1999;
27
586-590
- 21
Sun X G, Hansen J E, Stringer W W, Ting H, Wasserman K.
Carbon dioxide pressure-concentration relationship in arterial and mixed venous
blood during exercise.
J Appl Physiol.
2001;
90
1798-1810
- 22
Warburton D E, Haykowsky M J, Quinney H A, Humen D P, Teo K K.
Reliability and validity of measures of cardiac output during incremental to
maximal aerobic exercise. Part I: Conventional techniques.
Sports Med.
1999;
27
23-41
- 23
Wasserman K, van Kessel A L, Burton G G.
Interaction of physiological mechanisms during exercise.
J Appl Physiol.
1967;
22
71-85
- 24
Wong D H, Onishi R, Tremper K K, Reeves C, Zaccari J, Wong A B, Miller J B, Cordero V,
Davidson J.
Thoracic bioimpedance and Doppler cardiac output measurement: learning curve
and interobserver reproducibility.
Crit Care Med.
1989;
17
1194-1198
- 25
Wong K L, Hou P C.
The accuracy of bioimpedance cardiography in the measurement of cardiac output
in comparison with thermodilution method.
Acta Anaesthesiol Sin.
1996;
34
55-59
T. Nicolas
Laboratoire des Sciences du Sport
Place St Jacques · 25030 Besançon Cedex · France ·
Phone: +33 381 66 56 08
Fax: +33 381 66 56 92
Email: nicolas.tordi@univ-fcomte.fr