Eur J Pediatr Surg 2006; 16(6): 392-395
DOI: 10.1055/s-2006-924738
Original Article

Georg Thieme Verlag KG Stuttgart, New York · Masson Editeur Paris

The Advantages and Disadvantages of Nerve Monitoring During Thyroid Surgery in Childhood

T. Meyer1 , 2 , W. Hamelmann2 , W. Timmermann2 , B. Meyer3 , B. Höcht1
  • 1Paediatric Surgery Unit, Centre of Operative Medicine, Julius-Maximilians-University, Würzburg, Germany
  • 2Department of General, Visceral, Vascular and Transplant Surgery, Centre of Operative Medicine, Julius-Maximilians-University, Würzburg, Germany
  • 3Department of Anaesthesiology, Centre of Operative Medicine, Julius-Maximilians-University, Würzburg, Germany
Further Information

Publication History

Received: May 1, 2006

Accepted after Revision: May 13, 2006

Publication Date:
08 January 2007 (online)

Abstract

Background: Intraoperative damage of the recurrent laryngeal nerve (RLN) and the external branch of the superior laryngeal nerve (EBSLN) are common and cause complications in thyroid surgery. The method of intraoperative electrophysiological neuromonitoring (IONM) of the RLN and the EBSLN has been used in adults for several years. The question is whether IONM can be used in paediatric surgery for the identification and functional control of the RLN and the EBSLN, and what the advantages and disadvantages of using it are. Methods: Surgical treatment of benign (n = 9) and malign (n = 2) thyroid diseases was carried out in eleven children (median age: 13 years) with a total of 18 nerves at risk (NAR). The function of the vocal cord was monitored in all children before and after surgery. Intraoperative identification and functional control of RLN and EBSLN was performed using the Neurosign 100®. All results before and after resection of the thyroid gland were documented. Results: A clear and reliable identification of the RLN and the EBSLN was possible in all cases. All NAR showed a constant physiological nerve signal before and after surgical resection of the thyroid gland. However, in one young patient the postoperative examination of the vocal cord revealed a partial paralysis on the left side after thyroidectomy, despite the constant intraoperative signal of the RLN. After a period of six months, normal functioning had resumed. Conclusion: The IONM of both RLN and EBSLN in adult surgery could be used successfully in paediatric surgery. In this study a clear and positive identification of both nerves was possible in all patients. Therefore the IONM of the RLN and EBSLN could develop into a successful and easy method for the prevention of intraoperative nerve damage during thyroid surgery in childhood and young adolescence. Assuming that a constant IONM signal represents a normal vocal cord, our evaluation showed that there is a small percentage of false negative and positive results.

References

  • 1 Beldi G, Kinsbergen T, Schlumpf R. Evaluation of intraoperative recurrent nerve monitoring in thyroid surgery.  World J Surg. 2004;  28 589-591
  • 2 Brauckhoff M, Gimm O, Thanh P N, Brauckhoff K, Ukkat J, Thomusch O, Dralle H. First experiences in intraoperative neurostimulation of the recurrent laryngeal nerve during thyroid surgery of children and adolescents.  J Pediatr Surg. 2002;  37 1414-1418
  • 3 Eisele D W. Intraoperative electrophysiologic monitoring of the recurrent laryngeal nerve.  Laryngoscope. 1996;  106 443-449
  • 4 Hamelmann W, Meyer T, Timm S, Timmermann W. Kritische Beurteilung und Fehlermöglichkeiten des intraoperativen Neuromonitoring (IONM) bei Operationen an der Schilddrüse.  Zentralbl Chir. 2002;  127 409-413
  • 5 Hermann M, Hellebart C, Freissmuth M. Neuromonitoring in thyroid surgery: prospective evaluation of intraoperative electrophysiological responses for the prediction of recurrent laryngeal nerve injury.  Ann Surg. 2004;  240 9-17
  • 6 Jonas J, Bähr R. Die intraoperative elektrophysiologische Identifikation des Nervus laryngeus recurrens.  Chirurg. 2000;  71 534-538
  • 7 Meyer T, Hamelmann W, Hoppe F, Timmermann W. Intraoperatives Neuromonitoring des Nervus laryngeus recurrens.  Chir Praxis. 2002;  60 615-622
  • 8 Meyer T, Hamelmann W, Schramm C, Hoppe F, Thiede A, Timmermann W. Is there a high risk for damaging the external branch of the superior laryngeal nerve during thyroid surgery?.  Eur Surg. 2003;  35 202-205
  • 9 Møller A R. Basis of intraoperative neurophysiologic monitoring. Møller AR Intraoperative Neurophysiologic Monitoring. New York; Harwood Academic Publishers GmbH 1995: 1-15
  • 10 Thomusch O, Sekulla C, Machens A, Neumann H J, Timmermann W, Dralle H. Validity of intra-operative neuromonitoring signals in thyroid surgery.  Langenbecks Arch Surg. 2004;  389 499-503
  • 11 Timmermann W, Dralle H, Hamelmann W, Thomusch O, Sekulla C, Meyer T, Timm S, Thiede A. Reduziert das intraoperative Neuromonitoring die Recurrensparese bei Schildddrüsenoperationen?.  Zentralbl Chir. 2002;  127 395-399
  • 12 Timmermann W, Hamelmann W, Meyer T, Timm S, Schramm C, Hoppe F, Thiede A. Der Ramus externus des Nervus laryngeus superior (RELS): Ein Stiefkind in der Schilddrüsenchirurgie.  Zentralbl Chir. 2002;  127 425-428
  • 13 Timmermann W, Hamelmann W H, Thiede A. Schilddrüsenchirurgie: Neuromonitoring zur Schonung des Nervus recurrens.  Dtsch Ärztebl. 2004;  101 1341-1345

Dr. med. T. Meyer

Paediatric Surgery Unit
Department of General Surgery
Zentrum Operative Medizin (ZOM)

Oberdürrbacher Straße 6

97080 Würzburg

Germany

Email: Meyer_t@chirurgie.uni-wuerzburg.de