Abstract
The purpose of this report is to describe, based on the experience with 50 patients,
transcranial colour-coded duplex ultrasonography and to evaluate its practical potential
in the routine examination in comparison with conventional transcranial Doppier ultrasonography.
Under Visual control, the Dopplcr angle between the ultrasonic beam and the flow direction
can be adjusted. The mean angle of ineidence in the middle cerebral artery was 33°
+ 15°, in the anterior cerebral artery 35° ± 17°, in the posterior cerebral artery
45° ± 18° and in the basilar artery 15° ± 14°. The measurements of blood flow velocity
under consideration of the angle of ineidence resulted in 15%, 18%, 30% and 3% respectively
higher velocity values. Examples of pathologic findings demonstrate the practical
potential of this noninvasive method.
Zusammenfassung
Bei 50 Patienten wurden die bei der transkraniellen farbkodierten Duplexsonographie
gemessenen Strömungsgeschwindigkeiten der basalen Hirnarterien mit den Ergebnissen
der konventionellen transkraniellen Dopplersonographie verglichen. Der Beschallungswinkel
wurde unter visueller Kontrolle korrigiert und betrug in der A. cerebri media 33°
± 15°, 35° ± 17° in der A. cerebri anterior, 45° ± 18° in der A. cerebri posterior
und 15° ± 14° in der Arteria basilaris. Die maximalen Strömungsgeschwindigkeiten,
gemessen nach der Winkelkorrektur, waren in diesen Arterien um 15%, 18 %, 30 % bzw.
3 % höher als bei der konventionellen TCD-Metho-de. Anhand von pathologischen Beispielen
werden die Anwendungsmöglichkeiten dieser neuen Untersuchungsmethode diskutiert.
Key words
Transcranial Doppier ultrasonography - Transcranial colour-coded duplex ultrasonography
- Duplex Ultrasonography - Angle of ineidence - Intracranial arteries
Schlüsselwörter
Transkranielle Dopplersonographie - Transkranielle Farbduplexsonographie - Duplexsonographie
- Beschallungswinkel - Intrakranielle Arterien