Abstract
Background:
Guidelines established by the American Academy of Audiology (AAA) currently recommend behavioral testing when fitting frequency-modulated (FM) systems to individuals with cochlear implants (CIs). A protocol for completing electroacoustic measures on CI sound processors has not yet been established or validated when fitting either FM or digital-modulated (DM) systems, mini microphones, or mini microphones coupled to DM systems. In response, professionals have used or altered the AAA (2008) electroacoustic verification steps for fitting FM systems to hearing aids when fitting FM/DM systems, mini microphones, or mini microphones coupled to FM/DM systems to CI sound processors.
Purpose:
The purpose of this research is to determine if the electroacoustic verification guidelines established by AAA (2008) for fitting FM systems to hearing aids are feasible and verifiable when fitting mini microphones and mini microphones coupled to DM systems to CI sound processors.
Research Design:
Electroacoustic measures were conducted on 51 Cochlear Nucleus 6/CP910 sound processors, one Cochlear Wireless Mini Microphone 2+ (MM2+), and one Phonak DM System (one Roger Inspiro transmitter and one Roger X receiver) using an adapted AAA (2008) protocol (Nair et al, 2017). Phonak’s recommended default receiver gain setting was used with the Roger X receiver and adjusted if necessary to achieve transparency. Transparency refers to when the signal output of the device is the same when coupled and when not coupled to remote microphone technology.
Study Sample:
Electroacoustic measures were conducted on 51 Cochlear Nucleus 6/CP910 sound processors. In this study, the 51 Cochlear Nucleus 6/CP910 sound processors were either streaming to the Cochlear MM2+ or streaming to the MM2+ coupled to a Phonak DM system.
Data Collection and Analysis:
In a clinical setting, using the AAA (2008) protocol for electroacoustic measurements when fitting FM systems to hearing aids, electroacoustic measurements using various equipment (MM2+ and Phonak DM system) were performed on 51 Cochlear Nucleus 6/CP910 sound processors using the Audioscan Verifit to determine transparency and verify DM advantage, comparing speech inputs (65 dB SPL) in an effort to achieve equal outputs. If transparency was not achieved when the CI sound processor was streaming to the MM2+ coupled to the Phonak DM system at the default receiver gain, adjustments were made to the Roger X receiver’s gain. The integrity of the signal was monitored with the manufacturer’s monitor earphones.
Results:
Using the AAA (2008) hearing aid protocol, when the Cochlear Nucleus 6/CP910 sound processor was streaming to the Cochlear MM2+, transparency was achieved for 50 of 51 CI sound processors. Again, using the AAA (2008) protocol when the Cochlear Nucleus 6/CP910 sound processor was streaming to the Cochlear MM2+ coupled to the Phonak DM system at Phonak’s recommended default receiver gain, 28 sound processors achieved transparency. After the receiver gain was adjusted, the remaining 23 sound processors met transparency.
Conclusion:
Electroacoustic measurements and transparency can be achieved for CI sound processors coupled to either a MM2+ only or to a MM2+ and a DM system by adapting the AAA (2008) guidelines.
-
REFERENCES
- American Academy of Audiology (AAA) 2008 Remote Microphone Hearing Assistance Technologies for Children and Youth from Birth to 21 Years. www.audiology.org/recources/documentlibrary/Documents/HATGuidelines.pdf . Accessed September 4, 2015.
- Anderson KL, Goldstein H, Colodzin L, Iglehart F. 2005; Benefit of S/N enhancing devices to speech perception of children listening in a typical classroom with hearing aids or cochlear implant. J Educ Audiol 12: 16-30
- Cochlear Americas 2016 Mini-Microphone 2+ Data Sheet. http://www.vlibank.be/scan.jsp?PID=E30955&TYPE=pdf&RANG=A . Accessed June 1, 2016.
- De Ceulaer G, Bestel J, Mulder HE, Goldbeck F, de Varebeke SPJ, Govaerts PJ. 2016; Speech understanding in noise with the Roger Pen, Naida CI Q70 processor, and integrated Roger 17 receiver in a multi-talker network. Eur Arch Otorhinolaryngol 273: 1107-1114
- Fitzpatrick EM, Seguin C, Schramm DR, Armstrong S, Chenier J. 2009; The benefits of remote microphone technology for adults with cochlear implants. Ear Hear 30 (05) 590-599
- Nair EL, Sousa R, Wannagot S. 2017; Verification of a proposed clinical electroacoustic test protocol for personal digital modulation receivers coupled to cochlear implant sound processors. J Am Acad Audiol 28: 625-635
- Phonak 2014 Roger Inspiro User Guide. https://www.phonak.com/content/phonak/ca/fr/hearing-aids/accessories/roger-inspiro/_jcr_content/mainpar/downloads/productdownloadspar/download_1/file.res/User_Guide_roger_inspiro_029-0264.pdf . Accessed May 12, 2017
- Schafer EC, Kleineck MP. 2009; Improvements in speech recognition performance using cochlear implants and three types of FM systems: a meta-analytic approach. J Educ Audiol 15: 4-14
- Schafer EC, Thibodeau LM. 2003; Speech recognition performance of children using cochlear implants and FM systems. J Educ Audiol 11: 15-26
- Schafer EC, Musgrave E, Momin S, Sandrock C, Romine D. 2013; A proposed electroacoustic test protocol for personal FM receivers coupled to cochlear implant sound processors. J Am Acad Audiol 24: 941-954
- Wolfe J. 2016 Hear Well or Hearsay? Do Modern Wireless Technologies Improve Hearing Performance for CI Users?. https://www.phonakpro.com/content/dam/phonakpro/gc_hq/en/events/2016/international_pediatric_audiology_conference_atlanta/3_3_Jace_Wolfe.pdf . Accessed May 5, 2017.
- Wolfe J, Morais M, Schafer E. 2015; Improving hearing performance for cochlear implant recipients with use of a digital, wireless, remote-microphone, audio-streaming accessory. J Am Acad Audiol 26: 532-539
- Wolfe J, Morais M, Schafer E, Mills E, Mulder HE, Goldbeck F, Marquis F, John A, Hudson M, Peters BR, Lianos L. 2013; Evaluation of speech recognition of cochlear implant recipients using a personal digital adaptive radio frequency system. J Am Acad Audiol 24 (08) 714-724