Introduction
Colorectal endoscopic submucosal dissection (ESD) has enabled high en bloc resection
rates regardless of tumor size [1]
[2]
[3]. However, colorectal ESD is a more challenging procedure than conventional endoscopic
mucosal resection (EMR) because of its technical difficulties, longer procedure time,
and higher rates of perforation [4]
[5]. Generally, endoscopists manipulate both ESD knives and endoscopes using their right
hand during colorectal ESD. In an unstable situation, however, maneuvers with the
right hand make it difficult for endoscopists to maintain the endoscopic operative
view once they release their right hand from the endoscopes to manipulate the ESD
knives. To overcome this, we developed a novel device called the Nelaton Attachment,
which allows endoscopists to manipulate the ESD knives using two fingers (index and
middle fingers) on their left hand, dubbed the two-fingers method, while holding the
endoscopes with their right hand [6]
[7].
Some endoscopists use their left fingers to manipulate ESD knives during colorectal
ESD, especially in unstable situations where they cannot release their right hand
from the endoscopes. The concept behind our device is that even in a stable situation,
using left fingers to manipulate an ESD knife may facilitate colorectal ESD because
the endoscopists can keep their right hand on the endoscopes. In fact, using the Nelaton
Attachment, endoscopists can always use their left fingers to manipulate ESD knives
without releasing their right hand from the endoscopes. We believe that the Nelaton
Attachment will help endoscopists perform the two-fingers method more accurately,
which will reduce procedure times and enhance safety.
Although we have used the Nelaton Attachment for colorectal ESDs in clinical practice,
its safety and efficacy have not been evaluated. Therefore, we retrospectively investigated
the feasibility and safety of the Nelaton Attachment for colorectal ESD.
Patients and methods
We retrospectively reviewed the records of 36 consecutive patients with 37 colorectal
tumors who had undergone ESD at Kishiwada Tokushukai Hospital and Naritatomisato Tokushukai
Hospital between April 2016 and September 2018. The Nelaton Attachment was used for
22 of the 37 colorectal ESDs. Patient and tumor characteristics including sex, age,
and tumor location were investigated. Furthermore, treatment results including the
tumor and resected specimen size, procedure time, en bloc resection rate, R0 resection
rate, histological diagnosis, and complications such as perforation rate were analyzed.
We compared the characteristics of patients, tumors, and clinical outcomes of colorectal
ESDs between patients that underwent colorectal ESD with use of the Nelaton Attachment
(Nelaton Attachment group) and those that underwent colorectal ESD without use of
the Nelaton Attachment (non-Nelaton Attachment group). We also investigated the effect
of the Nelaton Attachment in an ex vivo experiment. This study was approved by the
institutional review board, and all patients met the indication criteria recommended
by the Japanese colorectal ESD/EMR guidelines [2], and written, informed consent for the ESD procedure was obtained from all patients.
Nelaton Attachment
The Nelaton Attachment was prepared by cutting a 10-Fr transurethral catheter (inner
diameter: 2.1 ± 0.1 mm, SAFEED Nelaton Catheter; Terumo, Tokyo, Japan) to a length
of 8 to 10 cm. The catheter was slit along its length and attached to the sheath of
a FlushKnife BT-S (Fujifilm Co., Tokyo, Japan) ([Fig. 1] and [Fig. 2]).
Fig. 1 Nelaton Attachment on the sheath of a FlushKnife BT-S.
Fig. 2 Control of the FlushKnife BT-S with the Nelaton Attachment.
ESD procedures using the Nelaton Attachment
Patients were sedated with intravenous flunitrazepam, pethidine hydrochloride and
dexmedetomidine
during the procedures. Normal saline was first injected into the submucosa to facilitate
subsequent injections of sodium hyaluronate solution (MucoUp; SEIKAGAKU Co., Tokyo,
Japan) into
an appropriate submucosal plane. Sodium hyaluronate was then injected to create adequate
protrusion of the targeted mucosa. A mucosal incision around the tumor and submucosal
dissection
of the tumor were both performed using a 1.5-mm FlushKnife BT-S until the tumor was
completely
excised. Carbon dioxide insufflation was used instead of air insufflation to reduce
patient
discomfort. When the tumor was located at the rectum or distal colon, a GIF-HQ290
(Olympus
Optical Co., Tokyo, Japan) was mainly used. On the other hand, when the tumor was
located at the
proximal colon, a PCF-260AZI (Olympus Optical Co.), CF-HQ290 (Olympus Optical Co.),
or PCF-H290I
(Olympus Optical Co.) was used. A transparent hood was attached to the endoscopic
tip in all procedures to make the lesion more visible. A VIO300 D (Erbe Elektromedizin,
Tubingen, Germany) was used as the electrosurgical generator. After reaching the tumor,
a 1.5-mm FlushKnife BT-S was inserted inside the forceps channel of the endoscope,
setting the tip of the knife at the usual position to incise and dissect the tumor,
and then the Nelaton Attachment was attached to the sheath of the 1.5-mm FlushKnife
BT-S. Medical tape was placed on the sheath of the FlushKnife BT-S just above the
distal part of the Nelaton Attachment to prevent positional dislocation. All colorectal
ESDs were carried out by a single endoscopist (T. Y.) with experience performing more
than 100 ESDs. The procedure time was defined as the time from the first submucosal
injection to the completion of dissection. Insertion or withdrawal of the 1.5-mm FlushKnife
BT-S was controlled using the two-fingers method ([Video 1], [Video 2]) [7].
Video 1 How to set up the Nelaton Attachment.
Video 2 How to use the Nelaton Attachment. We show a case of gastric ESD in which the endoscopist
encountered an unstable situation during the procedure, requiring the use of his left
fingers many times. The tumor was type IIa, 6 mm in size, and located at the prepylorus.
Ex vivo experiment
The ex vivo experiment involved inserting and withdrawing an ESD knife 2 cm using
the two-fingers method with and without the Nelaton Attachment. Before starting the
experiment, a 1.5-mm FlushKnife BT-S was inserted inside the same endoscope (PCF-260AZI;
Olympus Optical Co.) until the tip of the knife could be seen in the endoscopic view.
The knife was then withdrawn slightly until the tip of the knife could not be seen
in the endoscopic view. An endoscopist inserted the knife 2 cm and subsequently withdrew
the knife using the two-fingers method. A cycle comprised one insertion and one withdrawal.
Two different endoscopists (with 4 and 6 years of ESD experience) performed five cycles
with and without the Nelaton Attachment. The experiment was repeated five times, and
the average time to perform five cycles was calculated.
Statistical analysis
Sex, tumor location, morphology, histology, and perforation rate were compared using
the Fisher’s exact test. Age was compared using Student’s t-test. Tumor size, resected specimen size, procedure time, and average time for the
2-cm insertion and withdrawal for five cycles were compared using the Mann-Whitney
U test. The threshold for significance was P < 0.05. SPSS version 22.0 for windows (IBM Japan Ltd., Tokyo, Japan) was used for
statistical analyses.
Results
Participants and descriptive data
Twenty-one patients (8 men and 13 women) with 22 colorectal tumors were treated with
the Nelaton Attachment (Nelaton Attachment group), whereas 15 patients (9 men and
6 women) with 15 colorectal tumors were treated without the Nelaton Attachment (non-Nelaton
Attachment group). No significant differences in patient and tumor characteristics
were found between the two groups. The characteristics of the patients, tumor location,
and morphology are shown in [Table 1].
Table 1
Characteristics of patients and tumors.
|
NA (+) N = 22
|
NA (–) N = 15
|
P value
|
|
Age, years, mean ± SD
|
69.7 ± 12.3
|
68.9 ± 10.9
|
0.82
|
|
Male/female, n (%)
|
8/14 (36/64)
|
9/6 (60/40)
|
0.19
|
|
Tumor location, n (%)
|
|
|
0.84
|
|
|
5 (23)
|
5 (33)
|
|
|
|
3 (14)
|
0 (0)
|
|
|
|
1 (4)
|
1 (7)
|
|
|
|
6 (27)
|
4 (27)
|
|
|
|
4 (18)
|
3 (20)
|
|
|
|
3 (14)
|
2 (13)
|
|
|
|
|
|
> 0.99
|
|
|
6 (27)
|
4 (27)
|
|
|
|
11 (50)
|
7 (47)
|
|
|
|
5 (23)
|
4 (27)
|
|
NA, Nelaton Attachment; SD, standard deviation; LST-G, laterally spreading tumor-granular
type; LST-NG, laterally spreading tumor-nongranular type
Outcomes data
Median sizes of the resected specimens were 30 mm (range 23 – 46 mm) and 27 mm (range
17 – 94 mm) in the Nelaton Attachment and non-Nelaton Attachment groups, respectively.
Median procedure time was significantly shorter in the Nelaton Attachment group (38 min
[range 6 – 195 min]) compared to the non-Nelaton Attachment group (75 min [range 17 – 198
min]; P = 0.030). There were two cases of perforation in the non-Nelaton Attachment group. The
Nelaton Attachment was successfully used in the Nelaton Attachment group. No technical
or mechanical problems such as positional dislocation of the Nelaton Attachment occurred
during any of the procedures. Clinical outcomes of the colorectal ESDs are shown in
[Table 2].
Table 2
Clinical outcomes.
|
NA (+) N = 22
|
NA (–) N = 15
|
P value
|
|
Tumor size, median (range), mm
|
21 (13 – 35)
|
20 (8 – 85)
|
0.27
|
|
Resected specimen size, median (range), mm
|
30 (23 – 46)
|
27 (17 – 94)
|
0.15
|
|
Procedure time, median (range), min
|
38 (6 – 195)
|
75 (17 – 198)
|
0.030
|
|
Histology, n (%)
|
|
|
0.37
|
|
|
13 (59)
|
5 (33)
|
|
|
|
6 (27)
|
8 (53)
|
|
|
|
1 (5)
|
1 (7)
|
|
|
|
2 (9)
|
1 (7)
|
|
|
En bloc resection, n (%)
|
22 (100)
|
15 (100)
|
|
|
R0 resection, n (%)
|
22 (100)
|
15 (100)
|
|
|
Perforation, n (%)
|
0 (0)
|
2 (13)
|
0.15
|
NA, Nelaton Attachment.
Results of the ex vivo experiment
Median time to complete five cycles of the 2-cm insertion and withdrawal was significantly
faster with the Nelaton Attachment than without the Nelaton Attachment (P = 0.001). Results of the ex vivo experiment are shown in [Table 3].
Table 3
Results of the ex vivo experiment.
|
NA (+)
|
NA (–)
|
P value
|
|
Time (s), median (range)[1]
|
8.6 (7.1 – 17.2)
|
23.3 (11.0 – 32.6)
|
0.001
|
1 Time to insert and withdraw an ESD knife 2 cm for a total of five times.
Discussion
The aims of the current study were to retrospectively investigate feasibility and
safety of the Nelaton Attachment for performing colorectal ESD, and to evaluate the
effect of the Nelaton Attachment in an ex vivo experiment. Results of the ex vivo
experiment clearly show that it is possible to manipulate the knife faster with the
Nelaton Attachment than without the Nelaton Attachment using the two-fingers method.
Results of the clinical retrospective study demonstrate that the Nelaton Attachment
was successfully used in the Nelaton Attachment group without technical or mechanical
problems and that use of the device shortened procedure time. Perforation did not
occur in the Nelaton Attachment group, whereas two cases of perforation occurred in
the non-Nelaton Attachment group, suggesting that using the Nelaton Attachment does
not increase occurrence of adverse events in colorectal ESD.
The two-fingers method involves using the index finger and middle finger of the left
hand to manipulate the treatment device while holding the endoscope with the right
hand. Although this method is effective, it does require some training to master.
Moreover, the thin sheaths of the ESD knives can be difficult to manipulate using
the two-fingers method, and moisture can make the sheaths slippery. Therefore, we
designed a novel device, the Nelaton Attachment, which helps the endoscopists perform
this technique easily.
The Nelaton Attachment increases the thickness of the FlushKnife BT-S sheath from
2.2 to 3.3 mm, which allows endoscopists to perform the two-fingers method quickly.
The Nelaton Attachment is made of polyvinyl chloride, a waterproofing material, which
prevents the fingers from slipping under wet conditions. Moreover, the stopper effect
of the Nelaton Attachment enables endoscopists to insert the FlushKnife BT-S into
the endoscope faster [7].
In our hospital, we mainly use the FlushKnife BT-S for colorectal ESD. This knife
has some merits in terms of its water jet function, which means that endoscopists
can inject normal saline into the submucosal layer using the knife without the need
to exchange it with other needles [8]
[9]
[10].
Frequent injection of saline into the submucosal layer keeps the layer expanded, which
helps endoscopists dissect the submucosal layer safely. Although this water jet function
is effective, injection of saline using the FlushKnife BT-S moves the endoscope in
the opposite direction, which compromises the endoscopic view. With the Nelaton Attachment,
however, endoscopists can grip the endoscope with their right hand steadily while
injecting normal saline with their left two fingers, thereby preventing movement of
the endoscope. In addition, they can push (or pull) the ESD knives using their left
two fingers while pushing (or pulling) the endoscope with their right hand simultaneously.
Because of these merits, the Nelaton Attachment enables endoscopists to control the
ESD knives using their left fingers accurately and quickly (as if using their right
hand) while holding the endoscopes with their right hand, thereby maintaining a good
endoscopic operative view, which is one of the most important factors for quick and
safe ESD.
There is another way that ESD knives can be manipulated. Kikuchi et al. reported the
utility of the Smart Shooter, a device that enables endoscopists to manipulate ESD
knives using their right thumb while holding the endoscope with their right hand [11]
[12]. Use of this device, however, requires endoscopists to learn how to control the
treatment device with their thumb, which they rarely have a chance to practice. In
contrast, endoscopists can practice the two-fingers method during several different
procedures, including biopsy, EMR snare, and clipping. Furthermore, the Smart Shooter
has some structural issues and use limitations [13]. Unlike the Smart Shooter, the Nelaton Attachment can be used with several different
types of ESD knives and endoscopes. In fact, a 10-Fr Nelaton Attachment can be used
with the Dual J Knife, and a 12-Fr Nelaton Attachment can be used with the Splash
M Knife, IT2 and Hook J Knife [7] ([Table 4]). Other advantages of the Nelaton Attachment are that it is inexpensive and available
worldwide.
Table 4
Nelaton Attachment for different types of ESD knives.
|
Knife
|
Sheath diameter (mm)
|
NA (Fr)
|
|
Dual J
|
2.0
|
10
|
|
Splash M
|
2.7
|
12
|
|
IT 2/ nano
|
2.6
|
12
|
|
Hook J
|
2.6
|
12
|
ESD, endoscopic submucosal dissection; NA, Nelaton Attachment.
Limitations of the current study include all colorectal ESDs being performed by a
single endoscopist, small sample size, and the research’s retrospective nature. In
addition, the non-Nelaton Attachment group was treated during the initial phase of
this study while the Nelaton Attachment group was treated during the late phase. The
possibility that a learning curve related to the two-fingers method, if any, that
could affect the outcomes of the study cannot be denied. Prospective studies, a larger
number of patients, and use of the Nelaton Attachment by different endoscopists are
required to verify our results.
Conclusion
In conclusion, the Nelaton Attachment is feasible and safe for colorectal ESD, and
may facilitate this procedure.