Introduction
Among the main advantages of endoscopic ultrasound (EUS), tissue acquisition (TA) represents a valuable and accurate diagnostic technique for cytopathological and histological characterization of several abdominal lesions such as pancreatic masses, lymph nodes, or subepithelial lesions (SELs).
In spite of good results observed with EUS-guided fine-needle aspiration (EUS-FNA) and recent developments in this field, such as use of rapid on-site evaluation (ROSE) [1], contrast-enhanced guided FNA or tissue elastography [2], diagnostic sensitivity still remains an issue. Thus, the most important pitfall associated with this procedure is false-negative diagnosis, which has the potential to delay patient care and negatively affect patient outcomes.
Development of EUS-guided fine-needle biopsy (EUS-FNB) needles has generated a great deal of interest in the field of EUS-TA primarily based on proposed advantages over EUS-FNA needles concerning diagnostic accuracy. Furthermore, EUS-FNB seems to improve procurement of samples with preserved tissue architecture, thus allowing for immunohistochemistry required for certain diagnoses, obviating ROSE and obtaining results in fewer passes [2].
Two recent meta-analyses reached the conclusion that EUS-FNB performed with the reverse bevel needle (ProCore, Cook Medical, Limerick, Ireland) shows comparable diagnostic accuracy and sample adequacy in comparison to EUS-FNA [3]
[4]; even the previous finding of a lower number of passes through the lesion needed to obtain adequate samples with FNB was recently challenged [4].
Two newer FNB needles were recently introduced in endoscopic practice: one with fork-tip design with two leading sharp tips on the opposite side of the lumen (SharkCore, Medtronic, Minneapolis, Minnesota, United States), and another with Franseen tip design with three symmetric cutting edges (Acquire, Boston Scientific Corp, Natick, Massachusetts, United States).
Although these novel needle designs are thought to improve tissue capture and several studies have been published testing these novel devices, there is limited evidence on their diagnostic performance in terms of diagnostic yield and histology core procurement; hence the pressing need to systematically assess the increasing body of evidence in the field to better define their optimal role and safety in endoscopy.
The aim of this meta-analysis was to provide a pooled estimate of the diagnostic performance and safety profile of these two newer needles, namely Franseen and Fork-tip, thus attempting to determine their comparative efficacy and potential utility in EUS-TA.
The primary endpoint was sample adequacy. Secondary outcomes were diagnostic accuracy, optimal histological core procurement, mean number of needle passes, pooled specificity and sensitivity. Safety data were also analyzed.
Patients and methods
Inclusion and exclusion criteria
Only studies meeting the following criteria were included: 1) full-text articles recruiting patients undergoing EUS-TA of solid lesions with Franseen or Fork-tip needles; 2) studies published in English; and 3) articles reporting at least one of the following data: sample adequacy, diagnostic accuracy (or data useful for its calculation), histologic core procurement. Case reports, non-endoscopic studies, review articles, and animal models were excluded.
Search strategy
[Fig. 1] reports the search strategy followed in the meta-analysis.Bibliographic research was conducted on PubMed, EMBASE, Cochrane Library and Google Scholar including all studies fulfilling inclusion criteria published until March 2019. Details of the search strategy adopted are reported in Supplementary Table 1.
Fig. 1 Flowchart of included studies.
Relevant reviews and meta-analyses on use of EUS-TA with newer biopsy needles were examined for potential suitable studies. Authors of included studies were contacted to obtain full text or further information when needed.
Data extraction was conducted by two reviewers (AF and VDP) using a standardized approach (PRISMA Statement). The quality of included studies was assessed by two authors independently (AF, VDP) according to the Cochrane Collaboration’s tool for assessing risk of bias [5] for randomized controlled trials (RCTs) and the Newcastle-Ottawa scale [6] for non-randomized studies. Disagreements were solved by discussion and following a third opinion (NM).
Outcomes
The primary outcome was sample adequacy, defined as ability to procure cytological and/or histological samples adequate for interpretation; secondary outcomes were diagnostic accuracy (defined as true positive + true negative/total number of patients), optimal core histological tissue (defined as samples with high cellularity and quality enabling appropriate core assessment in terms of tissue architecture), specificity (true negative/true negative + false positive), sensitivity (true positive/true positive + false negative), number of needle passes, and safety.
Statistical analysis
Diagnostic outcomes were computed overall and then separately pooled for each needle through a random-effects model based on DerSimonian and Laird test, and summary estimates were expressed in terms of rate and 95 % Confidence Interval (CI). Chi-square and I² tests were used for across studies comparison of the percentage of variability attributable to heterogeneity beyond chance. P < 0.10 for chi-square test and I² < 20 % were interpreted as low-level heterogeneity.
Probability of publication bias was assessed using funnel plots and with Begg and Mazumdar’s test.
Safety data were inconsistently reported, hence they were analyzed descriptively.
Pooled diagnostic outcomes of the two needles were compared using the bivariate approach [7].
Sensitivity analysis was conducted according to location of the lesion sampled (pancreas vs lymph nodes vs SEL), quality of included studies, use of ROSE, and study design (single-cohort versus comparative). A separate comparative analysis was conducted with studies directly comparing newer needles to a control device (FNA or reverse-bevel FNB).
All statistical analyses were conducted using RevMan version 5 from the Cochrane collaboration and OpenMeta[Analyst] software. For all calculations a two-tailed P value of less than 0.05 was considered statistically significant.
Results
Characteristics of included studies
As shown in [Fig. 1], of 3323 studies initially identified, after preliminary exclusion of papers not fulfilling inclusion criteria, 26 potentially relevant articles were examined. Among these studies, two were excluded because they were focused exclusively on liver lesions [8] or tested the impact of metal stents on tissue acquisition [9].
Finally, 24 studies [10]
[11]
[12]
[13]
[14]
[15]
[16]
[17]
[18]
[19]
[20]
[21]
[22]
[23]
[24]
[25]
[26]
[27]
[28]
[29]
[30]
[31]
[32]
[33] with 6641 patients were included in the meta-analysis. Main characteristics of included studies are reported in [Table 1]. Four RCTs were conducted in Asia [14]
[17]
[18]
[19] and all comparative studies presented two well-balanced arms in terms of lesion size and clinical-demographical characteristics ([Table 1]). Quality was deemed mainly moderate/high with only three studies assessed as low-quality articles [21]
[24]
[31].
Table 1
Characteristics of included studies.
Study
|
Needle
|
Sample size
|
Study period/design
|
Country
|
Age
|
Gender male
|
Lesion size (cm)
|
Location
|
ROSE
|
Franseen
|
Adler 2018 [10]
|
Acquire
|
200
|
2016/Retrospective
|
USA
|
63 ± 14.5
|
121 (60.5 %)
|
3.6 (0.3 – 10)
|
Pancreas: 55 % Nodes: 23 % SE: 17 % Other: 4 %
|
Yes
|
Bang 2017 [11]
|
Acquire
|
30
|
2016/Retrospective
|
USA
|
71.5 (56 – 79)
|
21 (70 %)
|
3.4 (2.8 – 4)
|
Pancreas: 60 % Nodes: 10 % SE: 16.6 % Other: 13.4 %
|
Yes
|
Haseeb 2018 [12]
|
Acquire
|
132
|
2016 – 2017/Retrospective
|
USA
|
63.5 ± 13
|
87 (66 %)
|
NR
|
Pancreas: 73 % Nodes: 21 % SE: 4 % Other: 1 %
|
Yes
|
Leung Ki 2019 [13]
|
Acquire
|
54
|
2016/Retrospective
|
France
|
70 (61 – 78)
|
34 (74 %)
|
2.4 ± 1.38
|
Pancreas: 57 % Nodes: 17 % SE: 9 % Other: 17 %
|
No
|
Sugiura 2019 [14]
|
Acquire
|
100
|
2017 – 2018/Prospective
|
Japan
|
70 (31 – 87)
|
57 (57 %)
|
2.3 (0.7 – 8.2)
|
Pancreas: 81 % Nodes: 13 % SE: 3 % Other: 3 %
|
No
|
Franseen versus FNA
|
Bang 2017 [15]
|
Acquire FNA
|
46 46
|
NR/Crossover RCT
|
USA
|
67.9 ± 14.7
|
28 (60.9 %)
|
2.9 ± 0.8
|
Pancreas: 100 %
|
Yes
|
El Hajii 2018 [16]
|
Acquire Expect
|
51 50
|
2013 – 2017/Retrospective
|
USA
|
NR
|
NR
|
2.98
2.9
|
Pancreas: 45 % Pancreas: 40 %
|
Yes
|
Fujita 2018 [17]
|
Acquire Expect
|
17 44
|
2013 – 2017/Retrospective
|
Japan
|
72 (58.5 – 74.5) 67 (55 – 74.8)
|
10 (58.8 %) 27 (61.3 %)
|
2.67 (1.9 – 4) 2.39 (1.6 – 3)
|
SE: 100 % SE: 100 %
|
No
|
Ishikawa 2019 [18]
|
Acquire EZShot
|
50 36
|
2016 – 2018/Retrospective
|
Japan
|
70.5 (60 – 75) 65 (57 – 77)
|
30 (60 %) 21 (58.3 %)
|
3 (2 – 3.9) 2.3 (2.3 – 3.4)
|
Pancreas: 100 % Pancreas: 100 %
|
No
|
Mukai 2018 [19]
|
Acquire Expect
|
30 30
|
2016 – 2017/Retrospective
|
Japan
|
64.7 ± 12.5 69 ± 9.5
|
19 (63.3 %) 20 (66.6 %)
|
3 ± 1.07 2.7 ± 0.7
|
Pancreas: 100 % Pancreas: 100 %
|
No
|
Fork-Tip
|
DiMaio 2016 [20]
|
SharkCore
|
226
|
2014 – 2015/Retrospective
|
USA
|
66 (18 – 92)
|
113 (50 %)
|
2.6 (0.2 – 15)
|
Pancreas: 60 % Nodes: 22 % SE: 12 % Other: 6 %
|
Yes
|
Ishikawa 2018 [21]
|
SharkCore
|
85
|
2015 – 2016/Retrospective
|
Canada
|
62.3 ± 15
|
48 (60.7 %)
|
3.8 ± 3.7
|
Pancreas: 56 % Nodes: 15 % SE: 26 % Other: 3 %
|
No
|
Larsen 2018 [22]
|
SharkCore
|
41
|
2015 – 2016/Prospective
|
Denmark
|
68 ± 11
|
22 (54 %)
|
2.8 ± 1.1
|
Pancreas: 100 %
|
No
|
Fork-Tip versus FNA
|
El Chafic 2017 [23]
|
SharkCore EchoTip Ultra/Expect
|
15 91
|
2011 – 2016/Retrospective
|
USA
|
65 ± 12.7 64.8 ± 15.7
|
9 (60 %) 44 (48.3 %)
|
2.5 ± 0.9 2.8 ± 1.65
|
SE: 100 % SE: 100 %
|
Yes
|
Jovani 2017 [24]
|
SharkCore FNA
|
51 51
|
2015 – 2016/Retrospective
|
USA
|
63.5 ± 12.7 62.1 ± 11.6
|
22 (43.1 %) 27 (52.9 %)
|
3.16 ± 1.8 2.67 ± 1.1
|
Pancreas: 57 % Pancreas: 57 %
|
NR
|
Kandel 2016 [25]
|
SharkCore FNA
|
39 117
|
2012 – 2015/Retrospective
|
USA
|
66 (26 – 85) 70 (17.91)
|
19 (49 %) 65 (56 %)
|
1.8 (0.2 – 20) 2.3 (0.4 – 11)
|
Pancreas: 56 % Pancreas: 56 %
|
NR
|
Naveed 2018 [26]
|
SharkCore EchoTip
|
115 973
|
2009 – 2015/Retrospective
|
USA
|
66.1 66.7
|
57 (50 %) 496 (51 %)
|
2.75 2.59
|
Pancreas: 100 % Pancreas: 100 %
|
Yes
|
Song 2018 [27]
|
SharkCore®
EchoTip/Expect
|
139 42
|
2013 – 2017/Retrospective
|
USA
|
64.7 ± 11.9 61.2 ± 12.6
|
47.5 % 54.8 %
|
2.8 ± 1.44 2.8 ± 1.8
|
Pancreas: 78 % Pancreas:79 %
|
No
|
Witt 2018 [28]
|
SharkCore EchoTip
|
10 10
|
2015 – 2016/Retrospective
|
USA
|
NR
|
NR
|
NR
|
Pancreas: 100 % Pancreas: 100 %
|
Yes
|
Fork-Tip versus Reverse bevel
|
Abdelfatah 2018 [29]
|
SharkCore ProCore
|
162 139
|
2014 – 2016/Retrospective
|
USA
|
67 ± 12 67 ± 11.8
|
70 (49 %) 65 (47 %)
|
2.5 ± 1.4 2.4 ± 1.3
|
Pancreas: 66 % Pancreas: 63 %
|
77 % 98.3 %
|
Nayar 2016 [30]
|
SharkCore ProCore
|
101 100
|
2013 – 2015/Retrospective
|
UK
|
66.4 68.1
|
58 (59 %) 49 (49 %)
|
3.4 (1.4 – 9) 3.3 (1 – 8.5)
|
Pancreas: 100 % Pancreas: 100 %
|
No
|
Franseen/Fork-Tip versus FNA
|
Bang 2019 [31]
|
Acquire/ SharkCore Expect
|
938 2082
|
2014 – 2017/Retrospective
|
USA
|
67.1 ± 12.9 65.8 ± 13.7
|
510 (54.4 %) 1181 (56 %)
|
2.88 ± 1.32 2.69 ± 1.39
|
Pancreas: 73 % Nodes: 9.4 % Pancreas: 71 % Nodes: 13 %
|
Yes
|
Franseen versus Fork-Tip
|
Abdelfatah 2018 [32]
|
Acquire SharkCore
|
97 97
|
2015 – 2017/Retrospective
|
USA
|
63.7 ± 10.8 62.8 ± 15.5
|
47 % 52 %
|
2.4 ± 1.3 2.5 ± 1.4
|
Pancreas: 55 % Pancreas: 48 %
|
14.5 %
|
Bang 2018 [33]
|
Acquire SharkCore
|
50 50
|
2016 – 2017/Crossover RCT
|
USA
|
71.3 ± 11
|
28 (56 %)
|
2.4 ± 0.6
|
Pancreas: 100 %
|
Yes
|
Data are reported as absolute numbers (percentages) or mean (± standard deviation or with interquartile range)
FNA, fine-needle aspiration; NR, not reported; RCT, randomized controlled trial; ROSE, rapid on-site evaluation; SE: subepithelial lesion
Details on methodological characteristics and quality of included articles are shown in Supplementary Table 2.
Sample adequacy
As reported in [Table 2] and Supplementary Fig. 1, overall sample adequacy with the two newer needles was 94.8 % (93.1% – 96.4 %; I2 = 87 %). Subanalysis performed according to needle design showed superiority of Franseen needle over Fork-tip (96.1 %, 93.7 % to 98.4 % versus 92.4 %, 88.8 % to 96 %; P < 0.001), although this finding should be interpreted with caution due to the high heterogeneity observed (I2 = 82.9 % and 88 %, respectively; [Table 2] and Supplementary Fig. 2a and Supplementary Fig. 2b). There was no evidence of publication bias (data not shown).
Table 2
Overall and subgroup analysis of main diagnostic outcomes. Subgroup analysis was performed based on a) location of the target lesion (pancreas, nodes, subepithelial lesion), and b) needle used (Franseen versus Fork-Tip). Numbers in parentheses indicate 95 % confidence intervals.
Location
|
Needle
|
No. of Cohorts
|
No. of patients
|
Summary Estimate (95 % CI)
|
Within-group heterogeneity (I2)
|
Sample Adequacy
|
Overall
|
Overall
|
25
|
2894
|
94.8 % (93.1 %-96.4 %)
|
87 %
|
Franseen
|
12
|
857
|
96.1 % (93.7 %-98.4 %)
|
82.9 %
|
Fork-Tip
|
13
|
1099
|
92.4 % (88.8 %-96 %)
|
88 %
|
Pancreas
|
Overall
|
22
|
1987
|
95.6 % (94 %-97.3 %)
|
48.9 %
|
Franseen
|
10
|
567
|
97 % (94.8 %-99.3 %)
|
13.4 %
|
Fork-Tip
|
11
|
736
|
92.6 % (88.8 %-96.4 %)
|
34 %
|
Nodes
|
Overall
|
8
|
223
|
97.2 % (94 %-100 %)
|
52.6 %
|
Franseen
|
4
|
95
|
98.4 % (96 %-100 %)
|
0 %
|
Fork-Tip
|
3
|
54
|
77.6 % (46.2 %-100 %)
|
39 %
|
Subepithelial Lesion
|
Overall
|
9
|
218
|
95.6 % (92.2 %-98.9 %)
|
27 %
|
Franseen
|
5
|
68
|
97.6 % (94 %-100 %)
|
0 %
|
Fork-Tip
|
4
|
64
|
90.5 % (79.6 %-100 %)
|
38.9 %
|
Histological Core Procurement
|
Overall
|
Overall
|
17
|
1932
|
92.5 % (89.8 %-95.3 %)
|
79.8 %
|
Franseen
|
7
|
455
|
93.5 % (89.8 %-97.2 %)
|
65 %
|
Fork-Tip
|
9
|
539
|
90.8 % (85.5 %-96.2 %)
|
84 %
|
Pancreas
|
Overall
|
11
|
1192
|
93.7 % (90.7 %-96.8 %)
|
35.3 %
|
Franseen
|
5
|
225
|
94 % (89 %-99 %)
|
36.3 %
|
Fork-Tip
|
5
|
283
|
93.1 % (87.3 %-99 %)
|
33.8 %
|
Diagnostic Accuracy
|
Overall
|
Overall
|
15
|
945
|
95 % (93.5 %-96.5 %)
|
11 %
|
Franseen
|
8
|
492
|
95 % (92.5 %-97.5 %)
|
37 %
|
Fork-Tip
|
7
|
453
|
94.4 % (92.3 %-96.5 %)
|
0 %
|
Pancreas
|
Overall
|
11
|
631
|
96.2 % (94.7 %-97.7 %)
|
0 %
|
Franseen
|
7
|
324
|
96.8 % (94.9 %-98.7 %)
|
0 %
|
Fork-Tip
|
4
|
307
|
95.2 % (92.8 %-97.6 %)
|
0 %
|
Diagnostic Sensitivity
|
Overall
|
Overall
|
10
|
755
|
92.8 % (89.8 %-95.7 %)
|
64.2 %
|
Franseen
|
4
|
255
|
93.3 % (89.2 %-97.4 %)
|
45 %
|
Fork-Tip
|
6
|
500
|
92.2 % (87.9 %-96.6 %)
|
73 %
|
Pancreas
|
Overall
|
6
|
445
|
94.1 % (90.7 %-97.6 %)
|
40.4 %
|
Franseen
|
2
|
81
|
95.3 % (90.7 %-99.9 %)
|
0 %
|
Fork-Tip
|
4
|
364
|
93.4 % (88.3 %-98.4 %)
|
45.6 %
|
CI, confidence interval.
The study by Bang et al [31] did not report a subgroup analysis based on the needle design, therefore data from this study were considered only in the overall analysis.
The findings of main analysis were confirmed in sensitivity analysis performed according location of sampled lesions, study quality and use of ROSE. Sample adequacy in targeting pancreatic masses was 95.6 % (94 % – 97.3 %; I2 = 48.9 %) with significantly higher rates of adequate samples obtained with Franseen needle (97 %, 94.8 % to 99.3 % versus 92.6 %, 88.8 % to 96.4 %; P = 0.006). Of note, there was evidence of only low-moderate heterogeneity in sensitivity analysis for pancreatic masses (I2 = 13.4 % and 34 %, respectively; [Table 2] and [Fig. 2a] and [Fig. 2b]).
Fig. 2 Pooled analysis assessing rates of sample adequacy of a Franseen and b Fork-tip fine-needle biopsy in targeting pancreatic lesions. Sample adequacy in targeting pancreatic masses was 95.6 % (94 % – 97.3 %; I2 = 48.9 %) with significantly higher rates of adequate samples obtained with Franseen needle (97 %, 94.8 % to 99.3 % versus 92.6 %, 88.8 % to 96.4 %; P = 0.006).
When considering lymph nodes and SELs, findings were in keeping with the aforementioned results (adequacy 97.2 %, 94% – 100 % and 95.6 %, 92.2 % – 98.9 % with lymph nodes and SELs, respectively) with significant superiority of Franseen over Fork-tip needle (98.4 %, 96 % – 100 % versus 77.6 %, 46.2 % – 100%, P < 0.001 for lymph nodes and 97.6 %, 94 % – 100 % versus 90.5%, 79.6 % – 100 %, P = 0.006 for SELs). Heterogeneity was 0% and 39 % with the two needles, respectively ([Table 2]).
The above reported results were confirmed in sensitivity analysis restricted to high quality studies and those using ROSE (Supplementary Table 3). As depicted in Supplementary Fig. 3a, five studies [15]
[16]
[17]
[18]
[19] directly compared Franseen FNB needle to FNA, showing a clear advantage in terms of sample adequacy of Franseen needle (OR 4.29, 1.49 – 12.35; P = 0.007 and I2 = 0 %). Likewise, Fork-tip needle appeared to clearly overperform FNA in the pairwise meta-analysis of 5 studies [23]
[24]
[25]
[26]
[27] (OR 1.79, 1.01 – 3.19, P = 0.05 and I2 = 0 %; Supplementary Fig. 3b).
Technical details about handling of the sampled tissue are reported in Supplementary Table 4. Handling protocols were relatively homogeneous with no difference between Franseen and Fork-tip groups.
Optimal histologic core procurement, diagnostic accuracy, sensitivity and specificity
The rate of histological optimal core procurement obtained with the new FNB needles was 92.5 % (89.8 %-95.3 %) with evidence of high heterogeneity (I2 = 79.8 %). No difference between the two FNB needles was observed neither in main analysis (93.5 %, 89.8 % – 97.2 % versus 90.8 %, 85.5 % – 96.2 %; P = 0.4, [Table 2] and Supplementary Fig. 4) nor in subanalysis restricted to pancreatic masses (94 %, 89 % – 99 % versus 93.1 %, 87.3% – 99 %; P = 0.7, [Table 2]).
Moderate evidence of heterogeneity (I2 = 33 % to 35 %) and no evidence of publication bias was observed, as confirmed with Begg and Mazumdar’s test (P = 0.64). Overall diagnostic accuracy was 95 % (93.5 % – 96.5 %; I2 = 11 %) with no difference between Franseen and Fork-tip needle (95 %, 92.5 % – 97.5 % versus 94.4 %, 92.3 % – 96.5 %; P = 0.71, [Table 2] and Supplementary Fig. 5). Similarly, subanalysis of pancreatic lesions confirmed the above reported findings (96.8 %, 94.9 % – 98.7 % versus 95.2 %, 92.8 % – 97.6 %; P = 0.8 and I2 = 0 %).
Overall sensitivity of new FNB needles was 92.8 %, 89.8 % – 95.7 %, again with no difference between Franseen and Fork-tip needles (93.3 %, 89.2 % – 97.4 % versus 92.2 %, 87.9 % – 96.6%, respectively; P = 0.3, [Table 2]). Similar results were registered in subanalysis of pancreatic masses ([Table 2]). Heterogeneity was low to moderate (0 % to 45.6 %).
As expected, specificity was 100 % with both needles.
Number of passes and adverse events
Analysis of number of needle passes needed to obtain adequate sample showed favorable results with the new needles in comparison to standard FNA (mean difference: –0.42, –0.94 to 0.09, and –1.60, –2.23 to –0.98 with Franseen and Fork-tip needle, respectively; Supplementary Fig. 6).
Both needles were significantly superior to FNA when restricting the analysis to pancreatic masses (–0.44, –0.87 to –0.01, and –1.82, – 2.20 to –1.43 with the two needles, respectively; [Fig. 3]). Heterogeneity was moderate to high (I2 = 48 % to 92 %) and no evidence of publication bias was detected.
Fig. 3 Meta-analysis comparing mean number of needle passes of a Franseen and b Fork-tip needle in comparison to fine-needle aspiration. Both needles resulted in significantly superior to fine-needle aspiration (FNA) when restricting the analysis to pancreatic masses (–0.44, –0.87 to –0.01, and –1.82, –2.20 to –1.43 with the two needles, respectively). Heterogeneity was moderate to high (I2 = 48 % to 92 %).
Details on safety profile of the two devices are reported in Supplementary Table 5. Of note, only a small number of patients experienced adverse events and all of these complications (mainly bleeding) were mild and did not impact on patient outcomes.
Discussion
EUS-TA plays a pivotal role in the diagnostic algorithm of solid masses but its diagnostic accuracy is strictly dependent on a series of lesion-related features (such as size, number, histological type) and technical variables like needles adopted, number of passes or availability of ROSE. To at least partially overcome these limitations and given the pressing need for adequate histological samples for molecular analysis, biopsy needles have been developed and introduced in clinical practice [34].
Given recent development of novel FNB needle designs (such as Franseen and Fork-tip needles), there is a clear need to systematically evaluate the impact of these newer devices on EUS-TA. The current manuscript represents the first attempt to systematically assess diagnostic performance of newer FNB needles, namely Franseen and Fork-tip needles, in a variety of abdominal masses.
With a meta-analysis of 24 studies selectively evaluating Franseen and Fork-tip needles (alone or in comparison to a control group), we made several key observations.
First, the two newer needles showed striking results in terms of sample adequacy (94.8 %), rate of histological optimal core procurement (92.5 %), diagnostic accuracy (95 %), and sensitivity (92.8 %). Although a significant increase in sample adequacy was observed with Franseen needle as compared to Fork-tip (96.1 % versus 92.4 %, P < 0.001), the other diagnostic outcomes were perfectly comparable between the two needle designs with no significant difference observed in terms of histological core procurement, diagnostic accuracy, or sensitivity. Second, all sensitivity analyses confirmed findings from the main analysis, in particular both Franseen and Fork-tip needles performed well in targeting all abdominal lesions (sample adequacy 95.6 % for pancreatic masses, 97.2 % for lymph nodes, and 95.6 % for SELs). As expected, use of ROSE did not impact significantly on diagnostic yield. Third, meta-analysis of head-to-head studies directly comparing FNB to FNA showed a clear advantage in terms of sample adequacy of both FNB needles (P = 0.007 and P = 0.05, respectively). Fourth, as already pointed out in previous meta-analyses [3]
[4], FNB required a lower number of needle passes through the lesion to obtain adequate samples in comparison to standard FNA. Fifth, both FNB needles resulted to be safe with only a small number of patients experiencing mainly mild pancreatitis or bleeding events.
These findings, which are considerably more favorable as compared to those reported in previous meta-analyses on a reverse bevel FNB needle [3], are likely to be related to the different design of newer needles with higher number of cutting points (3 in Franseen and 6 in Fork-tip needle) designed to provide improved control at the puncture site and stability at the tip, allowing for enhanced penetration [35]. As tissue moves into the tip of the Franseen needle, suction is applied to push the tissue into three symmetrical cutting heels which cut tissue from three different angles to make a more circular cut. In the case of the Fork-tip device, two opposing catch beveled tips allow funneling of target tissue into the lumen of the needle, while a distal cutting design minimizes stacking and fracturing of collected samples to produce more intact tissue architecture [35].
As expected, the aforementioned results were not influenced by use of ROSE, a tool not routinely adopted in non-American centers, thus confirming that FNB may obviate the need for an on-site pathologist to obtain optimal diagnostic outcomes. Moreover, increased costs of newer needles may be at least partially compensated for by obviating the need for an on-site pathologist.
Because most of the included studies were single-cohort non-comparative series, results regarding the higher sample adequacy observed with the Franseen needle as compared to the Fork-tip device should be interpreted with caution, considering that the single head-to-head trial directly comparing the two newer FNB needles did not report significant differences between the two devices [15].
There are some limitations to our study. First, the limited number of case-control or randomized studies does not allow a strong comparison between newer needles and standard FNA or reverse bevel FNB. This aspect requires particular caution in interpreting our comparative findings due to the high risk of indirectness. In fact, the validity of indirect comparisons rests on the assumption that factors in the design of the included studies (patients, co-interventions, measurement of outcomes) and methodological quality are not sufficiently different to result in different effects (the so-called “similarity assumption” [36]). Because this assumption is always in some doubt, indirect comparisons always warrant rating down by one level in quality of evidence. Second, as most included studies used 22G needles, subanalysis based on needle caliper was not feasible. Therefore, absent of studies specifically testing different calipers, our results should be considered applicable only to the 22G FNB device, which is indeed the most frequently used FNB needle worldwide.
Conclusion
In conclusion, despite these weaknesses, our meta-analysis speaks in favor of use of newer biopsy needles as safe and effective tools in EUS-guided tissue acquisition. In particular, based on the extremely high rates of optimal histological core procurement, they may represent a valuable option in conditions that require assessment of tissue architecture (for example, in oncologic studies that require core biopsies for personalized medicine or benign conditions such as autoimmune pancreatitis). Further RCTs comparing the two newer needles to reverse bevel FNB or each other are warranted to confirm these promising results.