CC BY-NC-ND 4.0 · Endosc Int Open 2022; 10(06): E905-E909
DOI: 10.1055/a-1789-0238
Innovation forum

Use of modified therapeutic upper endoscope for ERCP in patients post pancreaticoduodenectomy

Nicholas M. McDonald
1   University of Minnesota Medical Center, Division of Gastroenterology, Hepatology, and Nutrition, Minneapolis, Minnesota, United States
,
Mohamed Abdallah
1   University of Minnesota Medical Center, Division of Gastroenterology, Hepatology, and Nutrition, Minneapolis, Minnesota, United States
,
Dharma Sunjaya
1   University of Minnesota Medical Center, Division of Gastroenterology, Hepatology, and Nutrition, Minneapolis, Minnesota, United States
2   Minneapolis VA Health Care System, Division of Gastroenterology and Hepatology, Minneapolis, Minnesota, United States
,
Mohammad Bilal
1   University of Minnesota Medical Center, Division of Gastroenterology, Hepatology, and Nutrition, Minneapolis, Minnesota, United States
2   Minneapolis VA Health Care System, Division of Gastroenterology and Hepatology, Minneapolis, Minnesota, United States
› Author Affiliations
 

Introduction

Since the invention of endoscopic retrograde cholangioscopy (ERCP) in 1968, the procedure has continued to evolve and has become integral to modern management of pancreaticobiliary disease [1]. While ERCP is commonly performed, in patients with gastric or small bowel surgically altered anatomy (SAA), the procedure may pose substantial technical challenges, resulting in failure [2] [3] [4] [5] [6]. Common types of SAA posing unique challenges to ERCP include Billroth I gastrectomy, Billroth II gastrectomy, Roux-en-Y gastric bypass, and pancreaticoduodenectomy, also known as a Whipple procedure [2] [3] [4] [5] [6]. Anatomical challenges in ERCP for SAA are often associated with difficulties in intubation of the anastomosis, length of the biliopancreatic (afferent) limb, inability to cannulate the biliary tree through the anastomosis, lack of an elevator mechanism, or endoscope incompatibility with the required endoscopic accessories [2].

In cases of ERCP in patients with pancreaticoduodenectomy anatomy, multiple types of endoscopes have been used, with a traditional side-viewing duodenoscope being the most common [7]. Chahal et al have previously evaluated outcomes of ERCP in patients with pancreaticoduodenectomy. In their series of 88 patients with pancreaticoduodenectomy who underwent ERCP, they found a 51 % technical success rate and low rates of adverse events (AEs) [7]. In this series, a conventional side-viewing duodenoscope was used in all the cases; in the 14.8 % of cases in which the conventional duodenoscope failed in successful completion for ERCP, an adult or pediatric colonoscope was utilized.

Conventional ERCP with a side-viewing duodenoscope allows for therapeutic interventions including sphincteroplasty, stone extraction, tissue sampling, stent placement, and more [2]. Adult or pediatric colonoscopes offer the advantage of being forward-viewing, and the longer length makes it possible to reach the biliary limb. Other options include using balloon enteroscopy-assisted ERCP [8] [9]. When a conventional duodenoscope cannot be used, each approach (adult and pediatric colonoscope, and balloon enteroscopy-assisted ERCP) has its advantages and potential limitations. Here we present our preliminary experience utilizing the newer modified therapeutic upper endoscope (1T; GIF-1TH190 Olympus, Center Valley, Pennsylvania, United States) for ERCP in patients with pancreaticoduodenectomy after a failed attempt with a conventional duodenoscope.


#

Results

Case 1

A 72-year-old man with a history of extrahepatic cholangiocarcinoma who underwent classic (non-pylorus sparing) pancreaticoduodenectomy 1 year prior and currently was on chemotherapy with capecitabine and oxaliplatin presented with jaundice. Laboratory evaluation revealed a total bilirubin of 31.1 mg/dL (reference range 0.2–1.2 mg/dL) and direct bilirubin of 22.6 mg/dL (reference range < 0.5 mg/dL) which was elevated from 1 month prior, at which, time the patient had normal bilirubin levels. Magnetic resonance cholangiopancreatography revealed intrahepatic biliary dilation with an abrupt cut-off at the level of the hepaticojejunostomy suggestive of stricture or tumor recurrence, as well as peritoneal carcinomatosis. Given this, the patient was referred for ERCP. A conventional side-viewing duodenoscope was utilized; however, it could not be advanced to the biliary limb. A 1 T therapeutic upper endoscope was then used. The biliary limb was identified and confirmed with the presence of suture material. However, despite aggressive interrogation, the hepaticojejunal anastomosis could not be identified. A tattoo was placed for identification of biliary limb, 3 to 4 cm into the biliary limb. The decision was made to perform an endoscopic ultrasound-guided rendezvous to obtain biliary access. A linear echoendoscope (GF-UCT180 Olympus, Center Valley, Pennsylvania, United States) was used and a left intrahepatic biliary radical was identified and punctured using a 19-gauge fine-needle aspiration needle. Then, a 0.025-inch and 450-cm-long straight guidewire was placed across the hepaticojejunostomy stricture and the echoendoscope was removed and the guidewire left in place. The 1 T therapeutic upper endoscope with a clear distal attachment cap was then inserted into the biliary limb and the hepaticojejunostomy was identified by viewing the guidewire entry point into the jejunum. The hepaticojejunal anastomosis was subsequently cannulated adjacent to the wire using a biliary balloon extraction catheter loaded with a guidewire. A cholangiogram demonstrated intrahepatic biliary dilation and focal stenosis at the level of the hepaticojejunostomy consistent with recurrent cholangiocarcinoma ([Fig. 1]). An uncovered metal stent (10 mm in diameter and 4-cm length) was placed, and the jaundice had resolved 8 weeks later at the time of last follow-up ([Fig. 2]). The patient was admitted post procedure for observation and was discharged the following day without any procedure- or anesthesia-related AEs.

Zoom Image
Fig. 1 Fluoroscopic view of dilation of intrahepatic biliary dilation extending to the hepaticojejunostomy.
Zoom Image
Fig. 2 Uncovered metal stent placed across the hepaticojejunostomy.

#

Case 2

A 64-year-old woman with a history of Gardner syndrome requiring colectomy with Kock pouch and classic (non-pylorus sparing) pancreaticoduodenectomy for an ampullary adenoma 10 years prior presented for follow-up of adenomatous tissue at the hepaticojejunal (HJ) anastomosis. Total bilirubin was mildly elevated at 1.5 mg/dL (reference range 0.2 to 1.2 mg/dL). She had previously been found to have adenomatous tissue at the hepaticojejunal anastomosis and had undergone endoscopic mucosal resection and intraductal radiofrequency ablation with resultant HJ stricture ([Fig. 3]). Due to the stricture, the bile duct could not be cannulated and required placement of an internal-external biliary drain by interventional radiology. When the patient returned for follow-up, after failure of a conventional side-viewing duodenoscope, a 1 T therapeutic upper endoscope was advanced to the HJ anastomosis and the biliary drain was visualized within the jejunum and fluoroscopically ([Fig. 4] and [Fig. 5]). The drain was removed by an interventional radiologist and a wire was placed across the anastomosis into the jejunum. Then, the wire was grasped with a snare and pulled through the working channel of the endoscope. A biliary balloon catheter was loaded over the wire and then advanced through the endoscope, allowing for successful cannulation. After balloon dilation of a 5-mm area of focal stenosis, the hepatico-jejunostomy was explored using cholangioscopy (Spyglass DS, Boston Scientific, Marlborough, Massachusetts, United States) through the 1 T therapeutic upper endoscope working channel. The cholangioscope was advanced to the hepatic duct bifurcation. Abnormal-appearing mucosa was seen from 2 cm distal from the hilum extending all the way to the HJ anastomosis ([Fig. 6]). SpyBite max forceps (Boston Scientific, Marlborough, Massachusetts, United States) were used and biopsies were obtained. Pathology of the biopsy specimen ultimately revealed tubular adenoma without dysplasia.

Zoom Image
Fig. 3 Endoscopic view using a 1 T therapeutic gastroscope with clear distal attachment cap showing the hepaticojejunal anastomosis.
Zoom Image
Fig. 4 Endoscopic view using a 1 T therapeutic gastroscope with clear distal attachment cap showing biliary drain from Interventional Radiology.
Zoom Image
Fig. 5 Fluoroscopic view of biliary cannulation through the hepaticojejunostomy via the modified 1 T therapeutic upper endoscope.
Zoom Image
Fig. 6 Cholangioscopy showing intraductal adenoma extending to the anastomosis.

#
#

Discussion

Surgically altered anatomy including pancreaticoduodenectomy anatomy poses a unique challenge to ERCP. A variety of endoscopes have been employed for ERCP in pancreaticoduodenectomy, each with their own advantages and disadvantages. A traditional duodenoscope allows for a side-viewing angle which can be helpful to locate the hepaticojejunostomy. At 4.2 mm, a duodenoscope offers the largest channel size of the endoscopes reviewed here. In addition, most common ERCP tools are designed for a traditional duodenoscope ([Table 1]). However, in patients with SAA including pancreaticoduodenectomy, it is often challenging to reach the biliary limb and the hepatico-jejunal anastomosis with a conventional duodenoscope. Potential reasons for this include length of biliary limb and acute angulations in the small intestine that are challenging to navigate with a side-viewing and relatively stiffer duodenoscope. This has led to use of adult and pediatric colonoscopes and balloon enteroscopy-assisted ERCP for performing ERCP in patients after pancreaticoduodenectomy.

Table 1

Characteristics of various endoscopes used in ERCP in patients with pancreaticoduodenectomy.

Duodenoscope (Olympus TJF-Q190V)

Pediatric colonoscope (Olympus PCF-PH190)

Adult colonoscope (Olympus CF-HQ190 L/I)

Enteroscope (SIF-Q180)

1 T therapeutic endoscope (GIF-1TH190 Olympus)

Working length

124 cm

168 cm (L model)
133 (I model)

168 cm (L model)
133 (I model)

200 cm

103 cm

Direction of view

Backward side viewing 15°

Forward viewing

Forward viewing

Forward viewing

Forward viewing

Channel size

4.2 mm

3.2 mm

3.7 mm

2.8 mm

3.7 mm

Distal end

outer diameter

13.5 mm

9.7 mm

13.2 mm

9.2 mm

10 mm

Maximum angulation (upwards)

120°

180°

180°

180°

210°

Compatible accessories

Fully-covered and uncovered metal biliary stents (8 mm and 10 mm)
Plastic stents (8.5F and 10F)
Biliary stone extraction balloon
Single-operator cholangioscope

8.5F plastic stent
Biliary stone extraction balloon

Plastic stents (8.5 Fr and 10 Fr)
Uncovered and fully-covered metal biliary stents (8 mm and 10 mm) [ select manufacturers][*]
Biliary stone extraction balloon

No standard devices [Dedicated devices][1]

Fully-covered and uncovered metal biliary stents (8F and 10F)
Plastic stents (8.5F and 10F)
Biliary stone extraction balloon
Single-operator cholangioscope

* Selected devices include Quantum TTC Balloon Dilator, (Cook Medical, Bloomington, Indiana, United States), Classic Cotton CannulaTome (Cook Medical, Bloomington, Indiana, United States), and 8.5F plastic stent and OASIS One Action Stent Introduction System, (Cook Medical, Bloomington, Indiana, United States).


Advantages of using pediatric and adult colonoscopes include having the largest endoscopic working length of the endoscopes reviewed here, which range from 133 cm to 168 cm depending on the model, a forward-viewing camera to aid in reaching the anastomosis, and variable stiffness, which can reduce looping. Disadvantages of the pediatric and adult colonoscopes include a smaller working channel (3.2 mm versus 3.7 mm with the pediatric and adult colonoscope, respectively), inability to perform short wire technique, or passage of several accessories utilized in ERCP. While plastic stents and biliary stone extraction balloons are compatible with pediatric and adult colonoscopes, it is critical to consider endoscope working length as well as the distance from the biopsy port to the insertion tube of the endoscope; while a device may be long enough to reach the distal end of the colonoscope, it may have limited device working length compared to when used through a duodenoscope, which potentially could lead to situations in which the device might not be long enough to reach the proximal biliary tree. Similarly, the single balloon enteroscope has the advantage of a forward-viewing camera, ability to be advanced to longer lengths, and tip angulation similar to an adult or pediatric colonoscope. The disadvantages include less maneuverability, inability to perform short wire technique during ERCP, and a relatively smaller working channel (2.8 mm) that does not allow for utilization of several accessories needed in ERCP such as stone extraction balloons, plastic biliary stents, or metal biliary stents. However, it is important to note that there are special dedicated accessories available that are compatible with the single balloon enteroscope. Those special devices allow for performing most interventions except for the use of a biliary stone extraction catheter and placement of a 10F plastic stent. [Table 1] summarizes the common accessories compatible with the previously mentioned endoscopes. The recent development of the modified therapeutic upper endoscope (1 T endoscope) offers several advantages when performing ERCP in patients with pancreaticoduodenectomy anatomy. The previous iteration of the therapeutic gastroscope was limited by its lack of maneuverability due to its stiffness and larger diameter [10]. The 1 T endoscope offers a working length of 103 cm, a channel size of 3.7 mm, and it is forward-viewing. One of the main advantages is the increased maneuverability and a maximal upward angulation of 210°, the highest of all the endoscopes reviewed here ([Table 1]). Due the endoscope length and working channel size, the 1 T endoscope can accommodate almost all the accessories that a conventional side-viewing duodenoscope can, including biliary stone extraction balloons, plastic biliary stents, and both uncovered and covered metal biliary stents. The conventional upper endoscope has previously been employed for ERCP in patients with surgically altered anatomy [11]. However, the conventional upper endoscope has a 2.8-mm working channel, and several accessories used during ERCP are not compatible with that. In contrast, the 1 T endoscope’s 3.7-mm working channel easily accommodates majority of accessories using in ERCP. The 1 T endoscope offers increased flexibility, which is helpful in navigating the small intestine. In addition, it can allow for using a short wire technique for performing ERCP. Finally, this is the only endoscope (apart from the conventional duodenoscope) from the ones mentioned previously that can be used to perform single-operator cholangioscopy (Boston Scientific, Marlborough, Massachusetts, United States) in patients with pancreaticoduodenectomy. Currently, in most patients with SAA who need single-operator cholangioscopy for exploration of the biliary tree, the most commonly utilized approach is cholangioscopy through a percutaneous biliary drain placed by an interventional radiologist prior to endoscopic intervention. The main limitation of the 1 T endoscope is the lack of an elevator and occasional inability to reach the biliary limb and hepaticojejunal anastomosis due to its length. While the 1 T endoscope can be helpful in these cases, it may not be able to reach the hepaticojejunostomy in some patients. In cases in which the 1 T endoscope fails, alternatives such as balloon-assisted ERCP or using a colonoscope for ERCP can be tried.


#

Conclusions

Here, we described two cases of patients with pancreaticoduodenectomy anatomy requiring ERCP for different indications successfully performed using a 1 T therapeutic gastroscope. In our practice, we use a distal attachment clear cap on the 1 T therapeutic gastroscope, which helps with engagement and cannulation of the hepaticojejunostomy. This approach has previously been described for balloon enteroscopy-assisted ERCP [9]. We prefer using this endoscope first because it allows us to determine if we can reach the biliary limb and hepaticojejunostomy in a short amount of time. Other tips that have been helpful in our practice include using minimal air or CO2 insufflation similar to colonoscopy, abdominal pressure in certain situations, and the use of suction to help in advancing the 1 T endoscope to the HJ anastomosis. We also use glucagon as necessary to help with small intestinal peristalsis. While the 1 T endoscope is not always successful in reaching the hepaticojejunostomy, it is useful to attempt, as almost all the accessories that are compatible with the traditional duodenoscope are compatible with the 1 T endoscope including 10F plastic and metal stents. Our preliminary data suggest that using the modified therapeutic upper endoscope is a useful addition to the endoscopist toolbox while performing ERCP in patients post pancreaticoduodenectomy.


#
#

Competing interests

The authors declare that they have no conflict of interest.


Corresponding author

Michael McDonald
Division of Gastroenterology, Hepatology, and Nutrition
University of Minnesota Medical Center
500 SE Harvard St, MMC 36
Minneapolis, MN 55454-1400
United States   
Fax: +1-612-625-5620   

Publication History

Received: 08 October 2021

Accepted after revision: 03 March 2022

Accepted Manuscript online:
04 March 2022

Article published online:
10 June 2022

© 2022. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/)

Georg Thieme Verlag KG
Rüdigerstraße 14, 70469 Stuttgart, Germany


Zoom Image
Fig. 1 Fluoroscopic view of dilation of intrahepatic biliary dilation extending to the hepaticojejunostomy.
Zoom Image
Fig. 2 Uncovered metal stent placed across the hepaticojejunostomy.
Zoom Image
Fig. 3 Endoscopic view using a 1 T therapeutic gastroscope with clear distal attachment cap showing the hepaticojejunal anastomosis.
Zoom Image
Fig. 4 Endoscopic view using a 1 T therapeutic gastroscope with clear distal attachment cap showing biliary drain from Interventional Radiology.
Zoom Image
Fig. 5 Fluoroscopic view of biliary cannulation through the hepaticojejunostomy via the modified 1 T therapeutic upper endoscope.
Zoom Image
Fig. 6 Cholangioscopy showing intraductal adenoma extending to the anastomosis.