Key words CT - CT-angiography - cardiac
Abbreviations
AUC:
Area under the curve
BMI:
Body mass index
BPM:
Beats per minute
CAD:
Coronary artery disease
CAD-RADS:
Coronary Artery Disease Reporting and Data System
CCTA:
Coronary computed tomography angiography
CT:
Computed tomography
C FFR:
CT-based calculation of fractional flow reserve
ECG:
Electrocardiogram
FDA:
US Food and Drug Administration
HIPAA:
Health Insurance Portability and Accountability Act
ICA:
Invasive catheter angiography
LAD:
Left anterior descending artery
LCX:
Left circumflex artery
MACE:
Major cardiac event
NPV:
Negative predictive value
PGMI:
Perfect, Good, Moderate, Inadequate
PPV:
Positive predictive value
RCA:
Right coronary artery
Introduction
Coronary artery disease (CAD) is the major cause of mortality and morbidity worldwide. It is challenging for the health care system and its budget [1 ]
[2 ]. Several noninvasive tests can be used to stratify CAD [3 ]
[4 ].
Coronary CT angiography (CCTA) is widely accepted as a reliable noninvasive modality for the exclusion of obstructive coronary stenosis [5 ]. It is a robust diagnostic test to rule out severe coronary stenosis. However, CCTA has limited specificity in determining the functional significance of coronary stenoses [6 ]
[7 ]
[8 ].
For this reason, its main application is in patients with a low-to-intermediate pre-test risk [9 ].
CCTA identified many patients with CAD requiring invasive catheter angiography (ICA). However, no significantly obstructed coronary vessels could be identified subsequently [10 ]. More than 50 % of lesions that appear to be obstructive on CCTA do not cause ischemia [11 ]. For this reason, the invasive FFR during catheter angiography is considered the gold standard for testing for ischemia [12 ]. The convincing scientific basis for broad use of the invasive FFR has so far been in strong contrast to the rather limited use in the clinical routine. With the constant technical development of modern computed tomography and the associated post-processing, there is the possibility of computed tomography (CT)-based calculation of the fractional flow reserve C FFR. Several studies showed that a C FFR value below 0.75–0.80 indicates a relevant stenosis which should be treated by revascularization [13 ]
[14 ]
[15 ]. cFFR using advanced computational analytic approaches allows hemodynamic assessment of a coronary lesion with a single noninvasive test, so that this can be combined with the anatomical information from the CCTA scan [16 ]. C FFR has already been proven to be superior to CCTA in the detection of hemodynamically significant stenoses. It is an independent prognostic risk factor for cardiac events and survival [17 ]
[18 ]. The technical aspect of the feasibility of C FFR has been proven in several multi-center studies, both on the basis of using scanners with at least 64 detector rows and mostly on high-end CT devices [17 ]
[19 ].
Computational fluid dynamics modeling techniques allow a precise calculation of cFFR noninvasively from standard CCTA images [20 ]. In this context, the C FFR provides superior diagnostic information for the identification of stenoses but has to be further investigated in individual cases and with respect to the degree of severity [21 ]
[22 ].
Nevertheless, cFFR models are complex and use different information regarding geometric, hemodynamic, and material parameters [23 ]. To challenge this, on-site machine-learning (ML)-based approaches for calculating cFFR were developed. Currently, the use of investigational artificial intelligence with deep machine learning algorithm software is well established as a potential time-efficient approach [24 ]
[25 ]
[26 ].
The aim of this retrospective single-center study is to evaluate the everyday practicability of the method on the scanner on-site, both in prospectively triggered and retrospectively gated examinations. The analysis not only focusses on comparing diagnostic accuracy with coronary computed tomographic angiography (CCTA) and invasive catheter angiography (ICA) but also encompasses post-processing, as well as technical and formal aspects of implementation, in conjunction with clinical parameters
Material and methods
Study population
In a single-center retrospective analysis of the database of the radiological department, 230 patients (74 female; 156 male, mean age 63.8 ± 9.4 years) with CCTA performed within 21 months between 01/2018 and 09/2019 and pathological findings were identified. The case selection is explained in detail in [Fig. 1 ]. Exclusion criteria were coronary anomalies, stents, CT scans with reduced quality (step artifacts, image noise), datasets with software problems, coronary artery aneurysm, chronic obstruction of coronary arteries, and significant stenosis of the main branch.
Fig. 1 Patient recruitment.
CT examinations
CCTA was performed according to the guidelines of the Society of Cardiovascular Computed Tomography. CT examinations were performed on Somatom Definition AS + (Siemens Healthcare GmbH, Erlangen) with a 128-row detector. Image acquisition was prospectively triggered to the patients’ electrocardiogram at max. 30 % to 80 % of the R-R interval per our routine clinical practice for CCTA, with a slice thickness of 0.75 mm and a reconstruction increment of 0.5 mm. Either prospective or retrospective electrocardiography synchronization (with X-ray tube current modulation for retrospective electrocardiography-gated protocols) was used for image acquisition. Sublingual nitroglycerin was administered 5 min. prior to scanning in all patients unless a contraindication was present. A beta blocker was given in the case of a heart rate over 65 beats per minute unless a contraindication was present. The maximum beta blocker dose was 100 mg metoprolol tartrate per os or 20 mg metoprolol tartrate i. v. CT examinations included intravenous contrast media injection (Imeron 400, Bracco Imaging Germany) in a weight-adapted dose. Contrast agent was injected at a flow rate of 4–6 mL/sec by antecubital vein. Timing bolus or bolus tracking was used, depending on the protocol.
CT quality was ranked as perfect, good, moderate, or inadequate (PGMI).
Software technique
Analyses of fractional flow reserve from CCTA were performed using a dedicated AI-based research software prototype (cFFR, version 3.5; Siemens Healthcare GmbH, Erlangen) [25 ]
[26 ]. A patient-specific model of the coronary artery tree was calculated using a semiautomatic approach.
The step-and-shoot procedure (prospective electrocardiogram (ECG) triggering) was used in n = 141 patients (61.3 %) and the helix method (retrospective ECG triggering) in n = 89 patients (38.7 %).
First, centerlines were automatically extracted by the software and had to be accepted by the reader. After acceptance or correction of the luminal contour proposals, a three-dimensional mesh representing the coronary artery tree was calculated.
After correction and acceptance of the coronary tree geometry and the myocardial mass by the user, the software used a multilayer neural network architecture to compute the functional severity of the lesion based on the anatomic information and the calculated hemodynamic conditions [24 ]
[25 ]
[26 ]. By using this semi-automatic on-site software approach, there was only a minimal need for human input.
No side branches were evaluated. C FFR values ≤ 0.8 were regarded as significant stenosis. C FFR values were evaluated at two points: at the maximum point of the stenosis and 1.0 cm distal to the stenosis. The lesion was identified on the basis of the CCTA finding and the parallel review of the ICA.
The degree of maximal stenosis was categorized according to the CAD-RADS scheme: non-significant stenosis (0–49 % stenosis; CAD-RADS 0–2), moderate stenosis (50–70 %, CAD-RADS 3) and obstructive stenosis (70–99 %, CAD-RADS 4). Vessels with total occlusion (CAD-RADS 5) were excluded because of falsification of the anatomic model of the coronary artery tree.
Furthermore, the Agatston Score, stenosis degree, and stenosis area were evaluated (last two points were shown in supplemental material).
Postprocessing analysis
To assess post-processing, we evaluated the mean post-processing time from starting the patient in the software to the final result of the evaluation.
Additionally, a subgroup classification concerning technical post-processing was performed:
No changes necessary (post-processing time < 15 min.)
Minor corrections necessary (centerline corrections, post-processing time between 15 and 25 min.)
Major correction necessary (post-processing time > 25 min., > 2 corrections, centerline and lumen correction necessary)
Technical failure = no evaluation possible
Furthermore, the required time from starting the software to the final result was evaluated.
Invasive coronary angiography
Invasive coronary angiography was performed as the gold standard. Invasive FFR measurements were obtained in cases of intermediate stenoses or lesions suspicious for coronary artery disease. An FFR value of less than or equal to 0.80 was considered diagnostic for lesion-specific myocardial ischemia. Invasive FFR enrollment was correlated by senior cardiac consultants (BD, BP).
Comparison invasive coronary angiography and C FFR values
The cFFR measured for each lesion was compared point by point with the ICA report. Stenosis over 70 % was classified as obstructive according to the CAD-RAD classification. In general, the degree of stenosis was compared, i. e., according to the CAD-RADS scheme. Depending on the evaluation, a correct finding was the confirmation of the exclusion of a relevant CAD or the confirmation of the presence of an obstructive form of the same. The invasive FFR was given to us as a numerical value by the treating cardiologists. Here because of the small number a further evaluation was not performed.
Compliance with ethical standards/ethical approval
This study received no funding. Our local ethics board approved this retrospective study and waived the written informed consent requirement (No. 408/19). All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. The study was HIPAA compliant.
Statistics
Statistical analysis was performed using dedicated software (IBM SPSS 27.0, Armonk, USA). All results are expressed as mean with standard deviation or median with interquartile ranges. The Kolmogorov-Smirnov test and Shapiro-Wilk test were used for testing normality including Lilliefors significance correction. Paired t-test was used for testing between cFFR values at different levels of the stenosis.
The Kruskal-Wallis test was used for comparison between the different post-processing times.
An invasive FFR of 0.80 or less was the reference standard. Bonferroni adjustment was made for multiple comparisons. The sensitivity, specificity, positive predictive value, and negative predictive values of CCTA, coronary angiography, and cFFR were calculated.
Results
Cohort characteristics
An overview of patient characteristics is given in [Table 1 ], [2 ]. All patient characteristics did not show a normal distribution (height, weight, BMI, diabetes, hypertonus, hypercholesterolemia, smoking, each p < 0.001*). The mean height was 173 cm, the mean weight was 84.25 kg, and the mean BMI was 28.10. Within the subgroup of three-vessel CAD, hypercholesterolemia was significantly higher than in the subgroup of two-vessel CAD (p < 0.001*) and in single-vessel CAD (p = 0.048*). Diabetes was significantly more common in the subgroup of three-vessel CAD than in the subgroup of two-vessel CAD (p = 0.038*).
Table 1
Overview of patient characteristics distributed to vessel CAD. Abbreviations: CAD = coronary artery disease; IQR = interquartile range.
Patient characteristics and risk factors
Overall
Single-vessel CAD
Two-vessel CAD
Three-vessel CAD
n (%)
n (%)
n (%)
n (%)
230/230(100 %)
115/230 (50.0 %)
67/230 (29.1 %)
58/230 (20.9 %)
Previously known CAD
126/230 (54.7 %)
46/115 (40.0 %)
36/67 (53.7 %)
44/58 (75.9)
Diabetes
31/230 (13.4 %)
13/115 (11.3 %)
7/67 (10.4 %)
11/58 (18.9 %)
Hypertonus
138/230 (60.0 %)
69/115 (60.0 %)
40/67 (59.7 %)
29/58 (50.0 %)
Hypercholesterolemia
76/230 (33.0 %)
34/115 (29.5 %)
18/67 (26.8 %)
24/58 (41.3 %)
Smoking
32/230 (13.9 %)
18/115 (15.6 %)
9/67 (13.4 %)
5/58 (8.6 %)
Agatston Score
328.3
IQR:69.0–617.3
142.2
IQR:16.3–150.3
360.5
IQR:79.4–486.1
731.2
IQR:265.1–1085.5
Table 2
Overview of patient characteristics distributed to cFFR. Comparison of Agatston values showed a significant difference (p = < 0.001*). However, other characteristics (diabetes, hypertonus, hypercholesterolemia, and smoking) did not show a significant difference (n.s.) tested by chi square post hoc test.
cFFR< 0.80
cFFR> 0.80
n (%)
n (%)
50/230 (21.7 %)
180/230 (78.3 %)
Diabetes
9/50 (18.0 %)
22/180 (12.2 %)
Hypertonus
31/50 (62.0 %)
107/180 (59.4 %)
Hypercholesterolemia
15/50 (33.3 %)
61/180 (33.9 %)
Smoking
7/50 (14.0 %)
25/180 (13.9 %)
Agatston Score
693.56
IQR:128.89–1330.59
310.62
IQR:67.99–561.98
The three-vessel CAD Agatston score was 731.2 (IQR 265.1–1085.5) and thus significantly higher than in two-vessel CAD with 360.5 (IQR 79.4–486.1) (p < 0.001) and in single-vessel CAD with 142.2 (IQR: 16.3–150.3) (p < 0.001). The correlation between Agatston Score and distal FFR is shown in [Fig. 2 ] (r = 0.046 p < 0.001*).
Fig. 2 Correlation of Agatston score with cFFR distal to the stenosis (r = 0.046, p < 0.001*).
There were a total of 116/448 (25.9 %) mild, 223/448 (49.8 %) moderate, and 109/448 (24.3 %) obstructive stenoses. Of these, 91/448 (20.3 %) were located in the RCA, 96/448 (21.4 %) in the LCX, and 261/448 (58.3 %) in the LAD.
CT parameter/quality
CT datasets with inadequate quality (n = 2) were excluded before statistical analysis. 32/230 (13.9 %) datasets had perfect quality, 122/230 (53.0 %) had good quality, and 74/230 (32.2 %) had moderate quality. 89/230 (38.7 %) CT scans were acquired retrospectively, 141/230 (61.3 %) prospectively.
The mean overall heart rate per minute was 60.5 ± 9.5 bpm (min. heart rate 55.4 ± 8.8 bpm, max. heart rate 67.9 ± 19.4 bpm). The RR distance was 1009.5 ± 140.0 ms.
In the subgroup of prospectively acquired CT scans, the mean heart rate was 58.3 ± 6.9 bpm vs. in the subgroups of retrospectively acquired CTs 64.6 ± 11.9 bpm, p = 0.001*. The min. heart rate of prospectively acquired CT scans was 54.4 ± 6.6 bpm, which was significantly lower than in the retrospectively acquired scans (58.0 ± 11.1 bpm), p = 0.010*. The max. heart rate showed no significant difference (66.9 ± 24.1 bpm vs. 72.8 ± 19.7 bpm, p = 0.054). The RR distance was significantly lower in the subgroup of prospectively acquired CT scans with 948.8 ± 158.0 ms vs. 1033.3 ± 116.1 ms (p = 0.001*).
C FFR values and correlation with invasive catheter angiography
The mean C FFR at the maximum point of the stenosis was 0.92 ± 0.09 and was significantly higher than the C FFR value of 0.89 ± 0.13 distal to the stenosis (p < 0.001*). A significant correlation between the two values was registered (r = 0.966, p = 0.001*). An example is given in [Fig. 3 ].
Fig. 3 Evaluation of a 60-year-old male patient with two-vessel CAD. The patient underwent CCTA for the evaluation of coronary status; in CCTA, multiple significant stenoses were detected in LAD (A) and in the proximal LCX. By evaluating the cFFR, significant stenosis can be verified: values of 0.72 and 0.64 were evaluated at the proximal stenosis and 1.0 cm above (B). In LCX complementary values of 0.79 and 0.72 were evaluated. One month after CCTA, the patient underwent invasive catheter angiography with quadruple stent supply of the LAD.
A detailed overview of stenosis parameters is given in supplemental material Table 1 .
Overall, in 57/230 patients (24.7 %), a catheter angiography examination was performed, while an invasive measurement of FFR was performed in only 2/48 patients. In ICA, more than one lesion was not suspected to be hemodynamically relevant with a necessity for FFR measurement in any patient.
No significant difference was detected between the degree of maximal stenosis (CAD-RADS 0–2/3/4) in CCTA and ICA (p = 0.171).
Overall, the degree of stenosis in CCTA correlated significantly with the degree of stenosis in ICA (r = 0.076, p < 0.001*). In subgroups of single-vessel CAD, it did not correlate significantly (r = 0.162, p = 0.092). However, a low significant correlation could be registered in two- and three-vessel CAD (r = 0.122, p < 0.001*/r = 0.126 p < 0.001*).
Overall, the C FFR values correlated significantly with the degree of stenosis in ICA (distal r = 0,084, p < 0.001*, at maximum point of stenosis r = 0,08, p = 0.003*).
Univariate linear regression analyses are additionally given in [Table 3 ].
Table 3
Univariate linear regression analyses for predictors of cFFR (distal). A p-value < 0.05 was considered statistically significant.
cFFR (distal) and
r2
p-value
Degree of stenosis in CCTA
r = 0.180
p < 0.001*
Degree of stenosis in ICA
r = 0.41
p = 0.018*
Agatston Score
r = 0.051
p < 0.001*
Post-processing time
r = 0.012
p = 0.011*
cFFR proximal
r = 0.924
p < 0.001*
Sensitivity/specificity/positive predictive value/negative predictive value
With a combination of CCTA and cFFR, a per vessel sensitivity/specificity/PPV/NPV of 70.6 %, 67.7 %, 64.3 %, and 73.7 %, respectively, were achieved. The detailed results are given in supplemental material Tables 2–4 .
Postprocessing analysis
Additionally, a subgroup classification concerning technical post-processing was performed. An overview is given in [Table 4 ].
Table 4
Overview of post-processing analysis classification distributed to vessel CAD. Additionally, a subgroup classification concerning criteria of technical post-processing was performed.
I
II
III
IV
Overall
No changes necessary
Minor corrections necessary
Major correction necessary
Technical failure, no evaluation possible
Single-vessel CAD
72
37
7
0
115
Two-vessel CAD
6
36
24
0
66
Three-vessel CAD
1
17
28
2
49
79
90
59
2
230
A significant, but weak correlation between the increasing subgroup and the number of vessels in CAD was registered (r = 0.041, p < 0.001*), also with C FFR values (r = 0.061, p < 0.001*). In contrast to this, no significant correlations could be registered between the degree of stenosis in CCTA and patient characteristics (height, weight, BMI, diabetes, hypertonus, hypercholesterolemia, smoking, each p = ns). The rejection rate was 16.3 % with 45 rejected examinations, e. g., for inadequate image quality.
Post-processing time
Significantly different mean postprocessing times were registered with 8.34 ± 4.66 min. in single-vessel CAD vs. 12.91 ± 3.92 min. in two-vessel CAD vs. 21.80 ± 5.94 min. in three-vessel CAD (each p < 0.001*). An overview is given in [Fig. 4 ].
Fig. 4 Distribution of post-processing time in minutes compared to vessel CAD and gender. The scatter in men is significantly greater than in women. In particular, the range of times required is greater among men with three-vessel CAD, but there is no significant difference between men and women in the three subgroups (p = 0.506; p = 0.433, and p = 0.219, respectively, for single/two/three-vessel CAD).
The range in men is significantly greater than in women. In particular, the range of times required is greater in men with three-vessel CAD, but there is no significant difference between men and women in the three subgroups (p = 0.506, p = 0.433 and p = 0.219, respectively for single/two/three-vessel CAD).
Discussion
This study demonstrates the technical feasibility of semiautomatic on-site software for C FFR in the daily routine on a 128-row CT scanner with minimal human input and corrections. A specificity of 82.3 % for C FFR could be achieved and a significant increase in sensitivity was achieved by combining C FFR and CCTA compared to a sensitivity of C FFR alone (70.6 % vs. 31.4 %). Our results also show that a measurement 1.0 cm distal to the stenosis is suitable for precise detection of C FFR values.
The diagnostic performance of C FFR has been validated in several major studies, both for the HeartFlow technique and for a research prototype by Siemens Healthineers GmbH
[27 ]
[28 ]
[29 ]. A C FFR value < 0.8 was the cut-off level for lesion-specific ischemia and shows a good correlation between C FFR and invasive FFR. Catheter-based invasive fractional flow reserve still tends to be used, and with technical progress and artificial intelligence, C FFR measurement was tested as an on-site approach [27 ]. Although there is evidence supporting their clinical use, C FFR techniques are not routinely applied in each patient up to now in clinical practice [30 ]. Here our study proved the feasibility in everyday use with minimal observer interaction in a routine real-world setting on a standard 128-row CT scanner.
1. Patient characteristics
Within the subgroup with three-vessel CAD, the prevalence of hypercholesterolemia and diabetes was significantly higher than in the other subgroups. Diabetes and hypercholesteremia are known risk factors for the development of CAD and damage to the small vessels in particular. These results are in line with the study by Zhu X. et al in 2022, who showed that the percentage of patients with a history of diabetes is higher in subgroups with major cardiac events (MACE) (70.9 % vs. 66.3 %, p < 0.05) [31 ]. Xue Y et al. showed that diabetes has no negative impact on the diagnostic accuracy of machine learning–based C FFR [32 ]
. Especially patients with type two diabetes mellitus have a high burden of CAD already in an early stage of the disease, and an advantage from hemodynamic evaluation of significant CAD by C FFR is proven [33 ]
.
2. C FFR evaluation
The C FFR was significantly higher at the maximum point of stenosis than distal to the stenosis. These results are in line with the studies by Nozaki Y. et al and Kueh S. et al., who showed that a C FFR measured 1–2 cm distal to a target stenosis is more consistent with the decision for revascularization than measured at the maximum point of the stenosis [34 ]
[35 ].
Similarly, in invasive FFR, ischemia evaluation is typically performed by establishing the ratio of the coronary pressure within 1–2 cm distal to an anatomical stenosis and pressure at the level of the aorta at maximal hyperemia [35 ].
Overall, the degree of stenosis in CCTA and the C FFR values correlated significantly with the degree of stenosis in ICA. This is consistent with the results of two large studies (DeFACTO and NXT), where the investigators demonstrated the superiority of C FFR with a good correlation between C FFR and invasive FFR [27 ]
[36 ].
Overall and concordant to the literature, CCTA showed a strong negative predictive value and a specificity up to 72.5 %, which is a little less than in other current single-center studies [37 ]
[38 ]. Chua A. et al showed a sensitivity of 75.1 % in a comparable Australian collective [38 ]. The cFFR results were slightly below average compared to other current studies with a specificity of 82.3 % and a sensitivity of 31.4 %. The lower results are most likely due to the group of patients who received invasive angiography as the gold standard. On the one hand, due to the location, patients with relevant suspicion of significant CAD receive direct ICA. On the other hand, the patients we saw only received secondary angiography in the case of unclear CT findings. A systematic correlation by means of catheter angiography does not take place due to the location, so that the collective is more inhomogeneous compared to prospectively planned studies. Another reason can be the relatively high overall Agatston Score in the patient collective, which could have a negative effect on the accuracy of the C FFR. The diagnostic performance of the recently introduced machine-learning algorithm still requires further investigation [40 ].
The mean overall Agatston Score in our study was 330.2 ± 488.8. Mickley H et al. already showed that most patients with an Agatston Score > 399 had at least one stenosis with a cFFR ≤ 0.80 [39 ]. Tesche et al. showed also a high performance in low-to-intermediate Agatston scores (0 to 400) with a statistically significant difference (AUC 0.71 vs. 0.85, p = 0.04) [40 ].
Hamilton et al. showed that in “real-world” practice C FFR does not improve standard radiological assessment of coronary disease graded by CAD-RADS alone [41 ]. In their analysis of 1145 coronary CT examinations the use of C FFR in CAD-RADS 2–4 resulted in reduced accuracy and specificity without a significant increase in sensitivity [41 ]. C FFR reduced the accuracy of the CAD-RADS grade significantly from 91 % to 78.4 % in their study [41 ].
3. CT parameter/quality
At 16.3 % the rejection rate in our study was equal to or slightly lower than in other single-center study in the current literature where it ranges from 13 % to 33 % [2 ]
[27 ]. The main reason for C FFR rejection in our study was motion artifacts, as in the study by Pontone G. et al. [2 ].
Additionally, a novel subgroup classification concerning criteria of technical postprocessing was established. A significant correlation between the degree of necessary post-processing steps and the degree of CAD was registered.
The time required for post-processing increased significantly with the degree of CAD (each p < 0.001*). As expected, the post-processing effort is greater in patients with multivessel CAD than in patients with only single lesions. In patients with 3-vessel CAD, the processing time increased significantly, so that in this patient group with a high pre-test probability of a relevant stenosis, the use of cFFR must be critically questioned. The overall time consumption of post-processing is little less than other single-center studies.
4. Limitations
There are several limitations to this study. First, not all patients underwent invasive catheter angiography and only a minute number underwent an additional invasive FFR. This might be due to the clinical stage of CAD in the included patients.
Second, the applied cFFR software prototype in this study is not FDA-approved compared to other commercial applications. Third, no side branches of the coronaries were evaluated. Fourth, the evaluation of cFFR and CCTA was not performed double-blinded, and the reader of the cFFR dataset knew the CCTA results in each case. Fifth, the cFFR evaluation was only performed with one software/vendor. A comparison to the HeartFlow software solution was not performed, which limited the significance of the results regarding daily practice. Six, it is a single-center study with patients from a circumscribed spatial area. There is a selection bias with respect to patients and CT quality. Seventh, tandem stenoses were not explicitly excluded. These are included in the description statistics because they were included in the original CCTA assessment.
Conclusion
Noninvasive on-site quantification of C FFR is feasible with acceptable observer interaction in a routine real-world setting on a 128-row scanner. DL-based algorithms allow a robust and semi-automatic on-site determination of C FFR using data from standard CT scanners.