Low Sampling Rates Bias Outcomes from the Wingate Test

E. L. Santos1 , J. S. Novaes2 , V. M. Reis3 , A. Giannella-Neto1
  • 1Federal University of Rio de Janeiro, Biomedical Engineering Program – COPPE, Rio de Janeiro, Brazil
  • 2Federal University of Rio de Janeiro, Graduate School of Physical Education, Rio de Janeiro, Brazil
  • 3UTAD, Sports, Vila Real, Portugal
Weitere Informationen

Publikationsverlauf

accepted after revision July 10, 2010

Publikationsdatum:
01. September 2010 (online)

Preview

Abstract

The purpose of this work was to apply a simple method for acquisition of power output (PO) during the Wingate Anaerobic Test (WAnT) at a high sampling rate (S R ) and to compare the effect of lower S R on the measurements extracted from the PO. 26 male subjects underwent 2 WAnTs on a cycle ergometer. The reference PO was calculated at 30 Hz as a function of the linear velocity, the moment of inertia and the frictional load. The PO was sampled at 0.2, 0.5, 1, 2 and 5 Hz. Both the peak (16.03±2.22 W·kg−1) and mean PO (10.34±1.01 W·kg−1) presented lower relative values when the S R was lower. Peak PO was attenuated by 0.29–42.07% for decreasing sampling rates, resulting in different values for 0.2 and 1 Hz (P<0.001). When the S R was 0.2 Hz, the time to peak was delayed by 53.81% (P<0.001) and the fatigue index was attenuated by 22.12% (P<0.001). In conclusion, due to the differences achieved here and the fact that the peak flywheel frequency is around 2.3 Hz, we strongly recommend that the PO be sampled at 5 Hz instead of 0.2 Hz in order to avoid biased errors and misunderstandings of the WAnT results.

References

Correspondence

Prof. Edil LuisSantos 

Federal University of Rio de Janeiro

Biomedical Engineering Program – COPPE

PO box 68510 Rio de Janeiro

21945-970 Rio de Janeiro

Brazil

Telefon: +55/21/2562 8575

Fax: +55/21/2562 8591

eMail: edil.luis@peb.ufrj.br