Subscribe to RSS
DOI: 10.1055/s-0034-1376188
An Irrigation Suction System Using a Pressurized Fluid High-Flow Pump System in Comparison with a Standard Cuff Pressure System: Technical Note
Publication History
02 August 2013
03 March 2014
Publication Date:
27 April 2015 (online)
Abstract
Objective In neurosurgical operations, such as clipping of ruptured cerebral aneurysms, irrigation of the operative field is a basic and extremely important surgical technique. Various irrigation-suction (IS) systems have been marketed for irrigation of operative fields in neurosurgery. However, with these devices, we often must increase the cuff pressure to maintain fluid power and change the fluid bag relatively frequently. We report the usefulness of pressurized irrigation fluid with the Niagara high-flow irrigation system (NI).
Method Using both the NI system and an IS, we altered pressure and duration to investigate differences in flow rate.
Results There was no change in flow rate 80 minutes after pressurization of the NI system, and flow rates decreased slightly. Additionally, flow rates using NI were sensitive to changes in pressure.
Conclusion This system has reduced the workload of assisting nurses and surgeons during surgery. Thus we consider the technique to be very useful for neurosurgical operations that require delicate procedures.
-
References
- 1 Bleasel K, Harrison J. Irrigating sucker dissector: technical note. J Neurosurg 1980; 52 (5) 736-738
- 2 Lüdecke DK, Treige W. Pressure-irrigation-suction system. Technical note. Acta Neurochir (Wien) 1982; 66 (1–2) 123-126
- 3 Miyake H, Tsuji M, Hiramoto N, Ukita T, Ohmura T. Development of a new irrigation sucker for microneurosurgery—technical note. Neurol Med Chir (Tokyo) 2004; 44 (8) 442-445 ; discussion 445
- 4 Okada Y, Kawamata T, Kawashima A, Yamaguchi K, Hori T. Pressure-controlled dual irrigation-suction system for microneurosurgery: technical note. Neurosurgery 2009; 65 (3) E625; discussion E625
- 5 Day JD, Fukushima T. Two simple devices for microneurosurgery: automatic drip irrigating needle and a suction retractor. Neurosurgery 1993; 32 (5) 867-868
- 6 Evans DM, Weightman B, Deane G. A new suction-irrigation device: uses in microsurgery and plastic surgery. Br J Plast Surg 1983; 36 (3) 273-277
- 7 Johnson JP, Becker DP. A continuous microneurosurgical irrigation and suction system: technical note. Neurosurgery 1996; 39 (2) 409-411
- 8 Murai Y, Adachi K, Takagi R, Koketsu K, Matano F, Teramoto A. Intraoperative Matas test using microscope-integrated intraoperative indocyanine green videoangiography with temporary unilateral occlusion of the A1 segment of the anterior cerebral artery. World Neurosurg 2011; 76 (5) e7 , e10
- 9 Oertel J, Gen M, Krauss JK, Zumkeller M, Gaab MR. The use of waterjet dissection in endoscopic neurosurgery. Technical note. J Neurosurg 2006; 105 (6) 928-931
- 10 Sano H, Kato Y, Zhou J , et al. New jet irrigation bipolar system. Neurosurgery 1996; 38 (6) 1251-1253
- 11 Gul S, Bahadir B, Hanci V , et al. Effect of vardenafil on cerebral vasospasm following experimental subarachnoid hemorrhage in rats. J Clin Neurosci 2010; 17 (8) 1038-1041