Introduction
In recent years, image-enhanced endoscopy (IEE), such as narrow-band imaging (NBI)
or flexible spectral imaging color enhancement (FICE), has been used to improve visualization
of the vascular patterns in the surface of the mucosa for enhanced diagnosis, such
as histological diagnosis [1]
[2], distinguishing between neoplasia and non-neoplasia [3], and depth of infiltration [4].
NBI is a commonly used IEE method that enhances the visualization of the surface vascular
pattern using optical filters against a xenon lamp that allows narrow-band light to
pass at wavelengths of 415 and 540 nm. NBI utilizes the wavelength-dependent absorption
rate of hemoglobin and the scattering characteristics of tissue to enhance the visible
vascular pattern. Short wavelength light, or blue light (415 nm), is strongly absorbed
by hemoglobin in the shallow depth of the mucosa that is permeated with surface blood
vessels, and the long wavelength light (540 nm) goes through to the middle depth of
the mucosa and the middle-depth blood vessels. Shorter wavelength light can depict
only the surface information on the mucosa such as surface blood vessels and surface
structure patterns instead of the mixed information as seen with the deeper area visualized
by white light (WL) illumination. NBI is also effective for detection of neoplasia
in the head and neck region and esophagus [5]. In the stomach, non-magnifying NBI does not work for detection because it is too
dark [6]. In colon polyp detection, NBI does not improve the adenoma detection rate during
primary colonoscopy compared to results with WL imaging [7].
Recently, a new IEE method named “Blue Laser Imaging” (BLI) has been developed by
FUJIFILM Corporation (Tokyo, Japan) [8]. The BLI system has a unique feature of illumination using two lasers and a white
light phosphor to accomplish the visual enhancement of surface vessels and structures.
One of the lasers, with a wavelength of 450 nm, stimulates the phosphor to irradiate
a white-color illumination, and the other laser, with a wavelength of 410 nm, is used
to enhance the blood vessels at shallow depth in the mucosa. The BLI system switches
between WL imaging mode and BLI mode by changing the intensity of these lasers.
In this study, we evaluated a prototype of the BLI endoscopy system that is equipped
for two BLI observation modes by changing the intensity balance of these two kinds
of laser. One mode is designed for high-contrast imaging (BLI-contrast) to observe
blood vessels in detail by near field or magnified view. The other (BLI-bright) is
designed for a brighter image and yet maintains the vascular contrast that is expected
with use of a far field of view; this may help to detect lesions in screening endoscopy.
These two BLI modes and the WL mode can be easily and instantly switched by pressing
a button on the handle of the scope.
BLI with magnification can predict histopathological diagnosis and invasion depth
with well-matched diagnostic effectiveness compared with NBI magnification [9]. However, there has been no report that evaluated the effect regarding observable
distance in BLI-bright mode. It is difficult to evaluate the efficacy between BLI
and NBI because visibility is affected by the location of the lesion, observable distance,
features of the lesion, and so on.
In this study, we evaluated the clinical utility of the BLI-bright mode by measuring
the observable distance from lesions and comparing this to similar data from the BLI-contrast
and commonly available NBI techniques. We performed endoscopic examinations using
BLI and NBI systems in a tandem manner in patients who had neoplastic lesions containing
early cancers in the pharynx, esophagus, stomach, or colorectum. With this process,
we observed the same lesion on the same day with both endoscopy systems. Then, brightness
and contrast in these images of the lesions were evaluated quantitatively according
to a general formula.
Patients and methods
Patients
Patients who had scheduled endoscopic examinations for either suspected neoplastic
lesions including adenoma, or endoscopic resection for intramucosal cancer in the
middle or lower pharynx, esophagus, stomach, or left colon and rectum from February
to August 2011 at the National Cancer Center Hospital East in Japan were assigned
to this study. They were all over 20 years of age, and their Eastern Cooperative Oncology
Group (ECOG) performance status was 0 or 1. Patients who had received surgical resection,
chemotherapy, or radiotherapy were excluded from this study. All patients gave written
informed consent, after clear and thorough explanation, to undergo tandem endoscopy
in a day, after the study was approved by the institutional review board of our hospital.
This study has been registered in the University Hospital Medical Information Network
Clinical Trials Registry (UMIN-CTR) as study #UMIN (ID: 000004983).
BLI endoscope system
The prototype of the BLI laser illumination endoscopic system developed by FUJIFILM
was used in this study. The system consists of the VP-4450HD processor, the laser
light source, and special scope series. The light source has two lasers; one of them
with a wavelength of 450 nm to stimulate the phosphor in the tip of the special scope
to irradiate white-color illumination that is similar to that of a xenon lamp, and
the other laser with a wavelength of 410 nm to enhance the visualization of blood
vessels.
This BLI system has three observation modes: WL mode, BLI-contrast, and BLI-bright.
The outward appearance of the visibility in WL mode is the same as that seen with
a conventional xenon light. BLI-contrast is accomplished with a strong 410 nm laser
intensity ratio for a high contrast image of blood vessels in the shallow depth of
the mucosa. BLI-bright is designed for a brighter image and yet maintains vascular
contrast with a suitable intensity balance of these lasers. While BLI-contrast is
made to observe blood vessels in detail with a near field or magnified view, the novel
feature of BLI-bright is its use with a far field of view that may help in the detection
of lesions in screening endoscopy. These three modes can be easily and instantly switched
by pressing a button on the handle of the scope.
Design of the clinical trial
This was the first in-human clinical trial using a novel BLI prototype. This study
was a prospective, comparative study using tandem endoscopic observation with both
BLI and NBI. The BLI system was a prototype of the BLI endoscopy system (FUJIFILM,
Tokyo, Japan), and the NBI system generally used in this study was the CV-260SL processor
and CLV-260SL light source with a GIF-H260Z or CF-H260AZI scope (Olympus, Tokyo, Japan).
Patients underwent tandem repeated endoscopic examinations with both the BLI system
and NBI system on the same day. For those patients who were scheduled for endoscopic
resection, endoscopy with the BLI system was performed first, and then endoscopy with
NBI was performed, because the existence and location of the individual lesions had
already been confirmed by initial endoscopy in our hospital. For those patients who
had suspected neoplastic lesions in the left side colon and rectum, colonoscopy with
the NBI system was performed first, and then colonoscopy with BLI was performed. When
no neoplastic lesions, including early cancer, were found at the first colonoscopy,
the patient’s enrollment was terminated without second colonoscopy. Only one lesion
was selected for each patient for assessment even though there were several lesions.
Images were acquired at distances of approximately 40, 30, 20, and 10 mm from the
lesion site without magnification, and at a distance of about 5 and 3 mm with magnification.
The following method was used to measure the distance between the lesion and the endoscope.
The depths of field of the BLI prototype scope, GIF-H260Z, and CF-H260AZI were 6 – 100 mm,
7 – 100 mm, and 7 – 100 mm without magnification, and 2 – 3 mm, 1.5 – 3 mm, and 2 – 3 mm
with magnification, respectively. At first, the image of the nearest position from
the surface of the mucosa where a clear image could be obtained without magnification,
i. e. just a few millimeters from the minimum depth of field of each scope, was acquired
and was defined as the image of “10 mm”. Then the scope was pulled out by hand in
steps of 10 mm and the images acquired were defined as 20 mm, 30 mm, and 40 mm, respectively.
Then, with the same technique, images from distances of about 5 mm with medium magnification,
and from about 3 mm with full magnification were obtained, respectively.
When the procedure was performed with the BLI system, images of BLI-bright and BLI-contrast
were acquired consequently by switching between these modes. Among the BLI and NBI
systems, the lesion was defined to be visible as a brownish area in individual organs
because there has been no official evaluation method regarding far-field observation
in IEE. The brownish area was defined as accumulation of neovascularity in the lesion
[10]. When the demarcation line of the brownish area could not be recognized, the judgment
was “negative” even though the lesion was detected by the morphological change.
All images were assessed at a later date by a single endoscopist (KK). Images were
displayed on a monitor, in the order of NBI, BLI-bright, and then BLI-contrast, and
in descending order of observation distance from 40 mm to 3 mm, to establish whether
it was “possible to distinguish regions exhibiting neoplastic changes” as a brownish
area. The farthest observation distance of the images where such regions could be
distinguished was defined as the “observable distance”. Each lesion was confirmed
by histopathologic analysis from the endoscopic resection or biopsy specimens according
to the Japanese classification for cancer of the head and neck [11], esophagus [12], stomach [13], and colon and rectum [14]. From resected specimens, the depth of infiltration in individual lesions was evaluated,
such as epithelial (EP), lamina propria mucosae (LPM), and muscularis mucosa (MM)
in the esophagus; mucosal (M) and submucosal (SM) cancer in the stomach; and adenoma,
mucosal (M), and submucosal (SM) cancer in the colorectum.
Quantitative evaluation of image data
Images taken from the stomach were used for the quantitative evaluation of brightness
and contrast. The image was selected with the following conditions: 1) images were
available for all conditions (BLI-bright/BLI-contrast/NBI, distance of 40/30/20/10 mm),
2) the imaging angle was a downward view, 3) the lesion was distinguished without
magnification in at least one of the conditions, and 4) the periphery of the lesion
as a brownish area was clearly distinguishable from normal mucosa.
Five sampling areas near the periphery of the lesion, and five corresponding sampling
areas on the surrounding normal mucosa were selected manually in each image ([Fig. 1]). Each sampling area included 100 pixels (10 × 10 pixels), and was selected as the
same position on the mucosa, avoiding the specular reflection area in the set of images
of the same lesion.
Fig. 1 Example of the sampling areas for image analysis observed by a BLI-bright, b BLI-contrast, c NBI. Red points are sampling areas on the lesion, and green points are those on the
normal mucosa.
The brightness Ym was defined as an average of the luminance values in the sampling
areas on the surrounding mucosa according to the formula (1) as shown below [15]:
Ym = 0.30 × R + 0.59 × G + 0.11 × B ... (1)
where R, G, and B indicate average signal intensities of red, green and blue in the
sampling area, respectively.
Contrast was defined by the following formula (2):
Contrast = (Ym – Yv)/Ym (2)
where Yv is the average brightness in the sampling area on the lesion, and Ym is the
average brightness in the corresponding sampling area on the surrounding mucosa. The
brightness and contrast of each image were defined as an average of each value from
the five sampling areas.
Statistical analysis
Patient characteristics and the observable distance as a result of evaluation were
summarized according to the disease site. Observable distances for either BLI-contrast
or NBI were compared with those of BLI-bright using the Wilcoxon signed-rank test.
All statistical tests were performed with SPSS version 11 (SPSS Inc, Chicago, IL,
United States). All P-values were reported as two-sided without adjusting for the multiple testing problem.
Results
Patient characteristics
After 51 patients had been recruited, these patients underwent the first endoscopy.
Eleven patients were removed from analysis of the study after the first endoscopy
because no lesion was found, and 40 patients immediately underwent the second examination.
Of the 40 lesions, 39 were histologically evaluated for endoscopic resection and the
remaining one lesion was evaluated by biopsy specimen. One stomach lesion could not
be distinguished without magnification, neither with BLI nor with NBI. This case was
diagnosed by histopathology as signet ring cell carcinoma. Hence, this was regarded
as indistinguishable and was excluded from subsequent analysis. Finally, a total of
39 lesions from 39 patients were analyzed in this study ([Table 1]).
Table 1
Analyzed patient characteristics (n = 39)
Age
|
|
Mean, years
|
71.0
|
Range, years
|
49 – 85
|
Sex, n (%)
|
|
Men
|
32 (82 %)
|
Women
|
7 (18 %)
|
Location, n (%)
|
|
Pharynx
|
3 (8 %)
|
Esophagus
|
8 (21 %)
|
SCC EP
|
(2)
|
LPM
|
(4)
|
Adenocarcinoma
|
(2)
|
Stomach
|
14 (36 %)
|
M
|
(9)
|
SM
|
(5)
|
Colorectum
|
14 (36 %)
|
EP, epithelium; LPM, lamina propria mucosae; M, mucosa; SCC, squamous cell carcinoma;
SM, submucosa.
Observable distance among BLI-bright, BLI-contrast and NBI images
An example set of images for the colon adenoma case is shown in [Fig. 2]. In this case, the whole lesion could be distinguished at up to a distance of 40 mm
in the series of BLI-bright images. In the series of BLI-contrast images and NBI images,
the clarity of the lesion decreased as the observation distance increased; the lesion
could not be distinguished at an image distance of 40 mm. Therefore, the observation
distances for BLI-bright, BLI-contrast, and NBI in this case were read as 40, 20,
and 10 mm, respectively. Since this lesion was composed of flat and nodular portions,
the nodule was visible at 40 mm in all three modes. However, the whole lesion composed
of flat and nodular portions was visible in BLI-bright alone. Visualization of the
flat portion from a far field of view has clinical impact.
Fig. 2 Example of the distance/IEE mode-dependent images of the same case. Left: BLI-bright,
middle: BLI-contrast, right: NBI. Images are shown at 40 mm/30 mm/20 mm/10 mm observation
distances. Red arrows indicate the lesion. One nodule in the lesion was visible in
three modes at 40 mm observation distance, however, the whole lesion, composed of
flat and nodular portions, was visible in BLI-bright alone.
[Table 2] shows the number of cases by individual observable distances, and [Table 3] shows the mean observable distances. The overall mean observable distances of BLI-bright,
BLI-contrast, and NBI were 31.3 mm, 24.7 mm, and 23.8 mm, respectively. There was
a significant difference for BLI-bright compared to BLI-contrast and NBI (P < 0.01, and P < 0.01, respectively).
Table 2
Observable distances of individual organ neoplasia
|
All
(n = 39)
|
Pharynx
(n = 3)
|
Esophagus
(n = 8)
|
Stomach
(n = 14)
|
Colorectum
(n = 14)
|
BLI-bright
|
|
|
|
|
|
40 mm
|
21
|
2
|
5
|
6
|
8
|
30 mm
|
9
|
1
|
1
|
3
|
4
|
20 mm
|
4
|
0
|
1
|
1
|
2
|
10 mm
|
1
|
0
|
0
|
1
|
0
|
5 mm
|
4
|
0
|
1
|
3
|
0
|
3 mm
|
0
|
0
|
0
|
0
|
0
|
BLI-contrast
|
|
|
|
|
|
40 mm
|
10
|
2
|
2
|
1
|
5
|
30 mm
|
10
|
1
|
2
|
2
|
5
|
20 mm
|
9
|
0
|
2
|
5
|
2
|
10 mm
|
7
|
0
|
2
|
3
|
2
|
5 mm
|
2
|
0
|
0
|
2
|
0
|
3 mm
|
1
|
0
|
0
|
1
|
0
|
NBI
|
|
|
|
|
|
40 mm
|
9
|
2
|
0
|
1
|
6
|
30 mm
|
10
|
1
|
3
|
3
|
3
|
20 mm
|
9
|
0
|
4
|
2
|
3
|
10 mm
|
7
|
0
|
1
|
4
|
2
|
5 mm
|
4
|
0
|
0
|
4
|
0
|
3 mm
|
0
|
0
|
0
|
0
|
0
|
Table 3
Mean observable distance
|
BLI-bright
|
BLI-contrast
|
P value[1]
|
NBI
|
P value[2]
|
All
|
31.3
|
24.7
|
< 0.01
|
23.8
|
< 0.01
|
Pharynx
|
36.7
|
36.7
|
1.00
|
36.7
|
1.00
|
Esophagus
|
31.9
|
25
|
0.13
|
22.5
|
0.06
|
Stomach
|
26.8
|
17.4
|
0.04
|
16.4
|
0.03
|
Colorectum
|
34.3
|
29.3
|
0.09
|
29.3
|
0.17
|
1
P value between BLI-bright vs. BLI-contrast.
2
P value between BLI-bright vs. NBI.
For both BLI and NBI, the observable distance was different depending on the organ
where the lesion was located. The longest distance was seen in the pharynx, and the
shortest was in the stomach. An example of a set of images in the pharynx, esophagus,
and stomach, when the observable distances were 40 mm and 20 mm, is shown in [Fig. 3]. An example of colon distances is shown in Fig. [2]. The maximum observable distance in the stomach for BLI-bright (26.8 mm) was significantly
longer than that for NBI (16.4 mm, P = 0.03). There were no significant differences in maximum observable distances for
lesions in other organs and IEE methods.
Fig. 3 Endoscopic images of lesions in the a pharynx, b esophagus, c stomach, illuminated by BLI-bright (left), BLI-contrast (middle), NBI (right) from
observation distances of 1) 40 mm, 2) 20 mm. Red arrows indicate the lesion.
[Fig. 4] shows that the observable distances depended on the depth of infiltration of histopathological
findings in the esophagus and stomach. Because of the limited number of cases in our
study, no significant statistical difference could be described, however, BLI-bright
showed that observable distances tended to be longer in esophageal cases of the LPM
than in the EP, and were longer in gastric SM cases than in mucosal cases.
Fig. 4 Average observable distances depending on the characteristics of each lesion. a EP, epithelial cancer; LPM, cancer infiltrating into the lamina propria mucosa; b M, mucosal cancer; SM, submucosal cancer.
Quantitative evaluation
To confirm the results from the endoscopist’s assessments, quantitative evaluation
of image data was conducted using representative cases in the stomach because the
differences between BLI-bright and other methods were the most pronounced in the stomach
lesion cases. Seven cases, which could be estimated in all observable distances from
40 mm to 3 mm, were selected from the 13 stomach cases that could be used for the
image analysis according to the criteria. Mean observable distances from the selected
seven cases were 37.1 mm, 17.6 mm, and 22.1 mm in BLI-bright, BLI-contrast, and NBI,
respectively.
The contrast curve depending on the observable distance is shown in [Fig. 5 a]. It is normalized at 10 mm of observable distance. While the contrast curve for
BLI-bright and BLI-contrast is stable from 10 mm to 40 mm, the contrast curve for
NBI decreases as the distance increases to more than 20 mm.
Fig. 5 a Contrast, and b brightness of BLI-bright, BLI-contrast, and NBI depending on the observation distances
calculated from image analysis. Each line is normalized by the average value at 10 mm.
The brightness curve dependent on observation distance is shown in [Fig. 5 b]. The brightness curve for each mode degraded depending on the observable distance.
Compared to the brightness curve of BLI-bright or NBI, the degradation of BLI-contrast
decreased rapidly at distances greater than 30 mm. This means that the image of BLI-contrast
became darker more rapidly with increasing distance than it did in the other modes,
making it difficult to find the lesions from a distance.
Longer observation distances made it difficult to distinguish the lesions by NBI because
of the degradation of contrast, although the image looked bright. Therefore, the BLI-bright
mode provided the best image to distinguish lesions in this situation.
Discussion
IEE is an effective method to detect neoplasia, distinguish neoplasia from non-neoplasia,
and to diagnose precursor or early cancer. NBI uses narrow band light by filtering
a xenon light source. It is suitable to visualize microvessels, however, the image
becomes dark when it is used for a far field of view. The new IEE “BLI” device is
a unique imaging method using laser light illumination. Because of this unique method,
it is able not only to achieve a high contrast image for the enhancement of surface
information but also to achieve a bright image.
This study revealed analytically the relationship of brightness, contrast, and observable
distance for BLI-bright, BLI-contrast, and NBI in neoplasia of the human gastrointestinal
tract. From the assessment by an endoscopist, it was revealed that BLI-bright had
significantly longer observable distances than the other methods. It was also shown
that BLI-bright mode had significantly longer observable distances than BLI-contrast
mode as seen in the same system by changing the intensity balance of the lasers.
Examination of the endoscopic images showed that they became darker in wider space
locations. A part of the illuminated light in the gastrointestinal tract reflects
off the surface of the mucosa, increasing the overall illumination in the location.
However, while this is particularly effective in a smaller space such as the pharynx,
in a wider area such as the stomach, the reflected illumination diffuses more readily
and provides much less benefit.
The statistical results corroborated this aspect. All cases in the pharynx could be
distinguished from 30 or 40 mm by all methods. However, a significant difference between
BLI-bright and the other methods was seen in the stomach, and an intermediate difference
was seen in the colorectum and esophagus. Therefore, it can be said that BLI-bright
will most benefit observations done in the stomach. Owing to the limited number of
cases, it is unclear whether it can be more effective than the other methods in the
colorectum and esophagus; however, BLI-bright appears effective in the detection of
lesions in these organs based on the results of this pilot study.
It was shown that visibility changes according to the characteristics of the lesion.
It seems that BLI-bright has a longer observable distance than does NBI in cases where
the lesions infiltrate into deeper layers, such as esophageal LPM or gastric SM cancers,
rather than in cases with a low degree of differentiation. In our previous study,
microvessel density had a correlation with tumor infiltration in esophageal squamous
cell carcinomas [16]. Because the number and diameter of microvessels on the surface of the tumor increased
according to invasion depth, we hypothesized that esophageal LPM or gastric SM cancers
would be visible as a brownish area in BLI-bright from the far field of view. BLI-bright
might be able to detect slightly infiltrated tumors more effectively than would the
other modes, and this capability is critical for establishing a correct diagnosis.
The visibility of lesions can usually be described by the relationship between the
brightness and contrast of images. In general, brightness is needed to detect the
color changes of the lesions, and contrast is needed to identify the demarcation of
the lesion from normal mucosa. The light intensity of both systems was controlled
via the automatic iris by default. Therefore, both brightness and contrast are needed
to maintain effective visibility. Stable contrast such as that from the BLI-bright
and BLI-contrast seen in [Fig. 5 a], is desired. Contrast is independent of brightness when the illumination intensity
is suitable, but when the illumination is too weak, the contrast degrades because
the noise of the sensor device becomes relatively strong compared to the signal from
the object. Even though the image brightness can be adjusted by auto-exposure or image
processing, degradation of contrast may still be present. The brightness curve is
determined by the balance between distance-dependent degradation, auto-light control
function, and other adjustment functions of the system such as the auto-exposure function
or image processing. The differences in these functions between the BLI and NBI systems
caused the differences between the luminosity curves of BLI-bright and NBI as seen
in [Fig. 5 b].
The visibility provided by BLI-contrast mode was limited in the far field of view
owing to the lack of brightness as shown in [Fig. 5 b], although the contrast curve remained stable as seen in [Fig. 5 a]. Alternatively, the contrast of the commonly available NBI, as seen in [Fig. 5 a], decreased when the observation distance was more than 20 mm. BLI-bright had a good
balance of both contrast and brightness to allow for the detection of lesions from
the far field of view. Both BLI-bright and BLI-contrast mode in this prototype seemed
to have the potential to improve the visibility through adjustments because the observable
distance was limited by the image brightness, although the contrast was stable up
to 40 mm in both modes of BLI.
The most important goal of endoscopic screening is to enable the chance to identify
an early cancer or precursor. In conventional endoscopy, the chance or motive of detection
for neoplasia relies on the reddish areas, irregular mucosa, and elevations and depressions.
In contrast, IEE lesions are detected through observation of a brownish area. However,
magnified observation of suspected lesions is required in IEE, apart from pharyngeal
and esophageal cancers. Because the target lesion can be visible in BLI-bright without
magnification from the far field of view, we suggest that BLI-bright might be used
as a new modality for cancer screening.
There were some limitations in our study. It was performed in a single institute with
a small number of cases, and the observable distance among the three modes was evaluated
by a single endoscopist. Observation became feasible at a distance a few millimeters
short of 10 mm without magnification, demonstrating that the 10-mm distance could
serve as a basis. However, images were obtained at distances of 20 mm, 30 mm, and
40 mm to minimize the observational error, with the scope withdrawn manually in increments
of 10 mm. Physical measures, such as forceps, which require precise placement, were
not used, given the potential risk of hemorrhage through their contact with the lesion
and the greater patient burden due to prolongation of the study time.
The endoscopist (KK) had been informed about the lesions that needed to be assessed.
Therefore, a “detectable” lesion was defined as a brownish area distinguishable in
each image evaluated, instead of a tumor lesion as detected based on its morphological
changes. Furthermore, it was necessary to perform final diagnosis and endoscopic treatment
during NBI endoscopy used generally. Therefore, when endoscopy with a BLI system was
performed first, it was followed by NBI endoscopy. The fact that the order of the
images to be displayed was fixed under these circumstances might be limited as biased
estimates of performance. This study was conducted using the commonly available NBI
device because the new-generation NBI, which provides brighter imaging than the one
used in this study, only became available after study registration. However, the aim
of this present study was a preliminary performance evaluation using a novel endoscope
system, with a first in-human clinical trial of this type. We feel that IEE with brighter
technology has a greater potential for benefit in screening procedures than do current
methods, and would be useful in clinical practice. Hence, performance assessment with
a large scale clinical trial regarding the detection of neoplasia with this new BLI-bright
mode would be desirable.
Conclusion
The BLI system is a novel IEE using laser illumination that enables operators to change
the brightness and contrast just by changing the laser intensity. BLI-bright mode
gives the ability to capture a high-contrast image from a long distance, and BLI-contrast
mode allows for magnified images. These two modes will facilitate both the detection
of suspected lesions and the making of accurate diagnoses with magnified IEE images,
especially in organs with wide internal spaces such as the stomach.