Endoscopy 2015; 47(09): 820-824
DOI: 10.1055/s-0034-1391438
Innovations and brief communications
© Georg Thieme Verlag KG Stuttgart · New York

Gastric endoscopic submucosal dissection using novel 2.6-mm articulating devices: an ex vivo comparative and in vivo feasibility study

Ryu Nakadate
1   Department of Research and Development for Surgical Support Systems, Center for Advanced Medical Innovation, Kyushu University, Fukuoka, Japan
,
Shotaro Nakamura
1   Department of Research and Development for Surgical Support Systems, Center for Advanced Medical Innovation, Kyushu University, Fukuoka, Japan
2   Department of Medicine and Clinical Science, Graduate School of Medical Sciences, Kyushu University, Fukuoka, Japan
,
Tomohiko Moriyama
2   Department of Medicine and Clinical Science, Graduate School of Medical Sciences, Kyushu University, Fukuoka, Japan
,
Hajime Kenmotsu
1   Department of Research and Development for Surgical Support Systems, Center for Advanced Medical Innovation, Kyushu University, Fukuoka, Japan
,
Susumu Oguri
3   Department of Advanced Medical Initiatives, Graduate School of Medical Sciences, Kyushu University, Fukuoka, Japan
,
Jumpei Arata
1   Department of Research and Development for Surgical Support Systems, Center for Advanced Medical Innovation, Kyushu University, Fukuoka, Japan
,
Munenori Uemura
3   Department of Advanced Medical Initiatives, Graduate School of Medical Sciences, Kyushu University, Fukuoka, Japan
,
Kenoki Ohuchida
4   Department of Surgery and Oncology, Graduate School of Medical Sciences, Kyushu University, Fukuoka, Japan
,
Tomohiko Akahoshi
3   Department of Advanced Medical Initiatives, Graduate School of Medical Sciences, Kyushu University, Fukuoka, Japan
,
Tetsuo Ikeda
3   Department of Advanced Medical Initiatives, Graduate School of Medical Sciences, Kyushu University, Fukuoka, Japan
,
Makoto Hashizume
1   Department of Research and Development for Surgical Support Systems, Center for Advanced Medical Innovation, Kyushu University, Fukuoka, Japan
3   Department of Advanced Medical Initiatives, Graduate School of Medical Sciences, Kyushu University, Fukuoka, Japan
› Author Affiliations
Further Information

Publication History

submitted01 September 2014

accepted after revision14 December 2014

Publication Date:
10 February 2015 (online)

Background and study aims: The conventional procedure of endoscopic submucosal dissection (ESD) is technically demanding. This study investigated the efficiency of novel articulating devices (maximum diameter 2.6 mm), which can be used with commercially available, standard endoscopes.

Patients and methods: In an ex vivo comparative study, eight endoscopists were divided into novices and experienced operators, and performed ESD using new devices and the conventional setup. An in vivo animal experiment was performed by two experts. Procedure times for incision and dissection were recorded, and unit times for circumferential length and area of specimens were calculated.

Results: All procedures were successfully completed with en bloc resection. In the ex vivo study, the unit procedure times for incision and dissection by novices were significantly shorter using the new system (P < 0.01 and P < 0.05), whereas there was no significant difference for experienced endoscopists. Perforation occurred during one procedure in which the new system was used. The in vivo experiments were successfully completed without adverse events.

Conclusions: ESD using novel articulating devices was feasible. These devices were able to reduce the procedure time for novices.

 
  • References

  • 1 Gotoda T, Kondo H, Ono H et al. A new endoscopic mucosal resection procedure using an insulation-tipped electrosurgical knife for rectal flat lesions: report of two cases. Gastrointest Endosc 1999; 50: 560-563
  • 2 Yahagi N, Fujishiro M, Kakushima N et al. Endoscopic submucosal dissection for early gastric cancer using the tip of an electrosurgical snare (thin type). Dig Endosc 2004; 16: 34-38
  • 3 Deprez PH, Bergman JJ, Meisner S et al. Current practice with endoscopic submucosal dissection in Europe: position statement from a panel of experts. Endoscopy 2010; 42: 853-858
  • 4 Fukami N. What we want for ESD is a second hand! Traction method. Gastrointest Endosc 2013; 78: 274-276
  • 5 Wang Z, Phee SJ, Lomanto D et al. Endoscopic submucosal dissection of gastric lesions by using a master and slave transluminal endoscopic robot: an animal survival study. Endoscopy 2012; 44: 690-694
  • 6 Thompson CC, Ryou M, Soper NJ et al. Evaluation of a manually driven, multitasking platform for complex endoluminal and natural orifice transluminal endoscopic surgery applications (with video). Gastrointest Endosc 2009; 70: 121-125
  • 7 Fuchs KH, Breithaupt W. Transgastric small bowel resection with the new multitasking platform EndoSAMURAI™ for natural orifice transluminal endoscopic surgery. Surg Endosc 2012; 26: 2281-2287
  • 8 Dallemagne B, Marescaux J. The ANUBIS™ project. Minim Invasive Ther Allied Technol 2010; 19: 257-261
  • 9 Nakadate R, Kenmotsu H, Nakamura S et al. 2.6 mm articulating endoscopic submucosal dissection device insertable into standard endoscope. Int J CARS 2014; 9: S181-S182
  • 10 Sakurazawa N, Kato S, Miyashita M et al. An innovative technique for endoscopic submucosal dissection of early gastric cancer using a new spring device. Endoscopy 2009; 41: 929-933
  • 11 Akahoshi K, Okamoto R, Akahane H et al. Endoscopic submucosal dissection of early colorectal tumors using a grasping-type scissors forceps: a preliminary clinical study. Endoscopy 2010; 42: 419-422
  • 12 Homma K, Otaki Y, Sugawara M et al. Efficacy of novel SB knife jr examined in a multicenter study on colorectal endoscopic submucosal dissection. Dig Endosc 2012; 24: 419-422