Thromb Haemost 2002; 88(05): 773-780
DOI: 10.1055/s-0037-1613301
Review Article
Schattauer GmbH

Rapid-SF: A Rapid Whole-Blood Screen for Soluble Fibrin Monomer[*]

Karen L. Hay
1   School of Medicine, Department of Pathology and Human Anatomy, Loma Linda University, Loma Linda, California, USA
,
Brian S. Bull
1   School of Medicine, Department of Pathology and Human Anatomy, Loma Linda University, Loma Linda, California, USA
› Institutsangaben
Weitere Informationen

Publikationsverlauf

Received 19. März 2002

Accepted after resubmission 02. Juli 2002

Publikationsdatum:
08. Dezember 2017 (online)

Zoom Image

Summary

Thrombosis accounts for a high proportion of disability and death in the West. Although soluble fibrin (SF) assays have been shown to be good predictors of thrombosis, current quantitative assays are too complex or lengthy to provide timely results, while simpler methods are qualitative and lack sensitivity. We here describe a rapid, new, protamine-based whole-blood screening method for SF which is quantitative and suitable for point-of-care use. Citrated whole blood is mixed with reagent under controlled conditions and the time until development of an SF precipitate is measured. Negative samples remain precipitate-free for 300 seconds. Strongly positive samples develop precipitate in as little as 10 seconds. SF times are mathematically converted to arbitrary SF units (SFU). This Rapid-SF test provides a simple and reliable means of detecting the presence of SF, and is well-suited for whole-blood rapid screening in the emergency department, operating room or clinic.

* Funded in part by a grant from the Montgomery Street Foundation