Keywords arthroplasty, replacement, knee - knee prosthesis - prosthesis design - prosthesis failure
Introduction
The aging of the population and the higher prevalence of patients with osteoarthritis has increased the frequency of indications for total knee arthroplasty (TKA).[1 ]
[2 ] Total knee arthroplasty can be defined as a highly-complex surgical procedure for the treatment of arthrosis that is capable of demonstrating satisfactory and long-lasting data on the improvement in pain, quality of life and patient functional outcomes, as well as in the correction of deformities and instabilities with origins related to degenerative processes that compromise the knee joint.[3 ] This procedure has excellent postoperative results in relation to implant survival, with rates higher than 95% in at least 10 years of follow-up.[4 ]
Polyethylene wear can produce debris that influences the release of prosthetic components. Total knee arthroplasty with a rotating platform has theoretical biomechanical advantages over a fixed-platform design.[5 ] These advantages include an improvement in kinematics by increasing the range of motion, facilitating axial rotation, better distribution of stress between the femoral and tibial components, and reduction in release forces at the implant interface with the bone.[6 ]
[7 ]
[8 ] Many variables can influence the frictional wear behavior of polyethylene, like the design of the prosthesis, the raw material used, and the surgical technique applied and the patient's morbidities, such as the activity level and body mass. The objective of the present study was to evaluate the wear of polyethylene of a Brazilian ultracongruent knee prosthesis with a rotating platform (Rotaflex, Víncula, São Paulo, SP, Brasil).
Materials and Methods
We used the test method with the standard parameters for loading and preparation listed in the norm ISO 14243-1:2009–Implants for Surgery – Wear of Total Knee Joint Prostheses – Part 1: Loading and Displacement Parameters for Wear Testing Machines with Load Control and Corresponding Environmental Conditions for Tests (Accredited) , and the measurement methods mentioned in the norm ISO 14243–2:2009 Implants for Surgery – Wear of Total Knee Joint Prostheses – Part 2: Methods of Measurement (Accredited) . For the evaluation of the wear behavior, the Rotaflex prosthesis was used.[9 ]
[10 ]
In total, 3 simultaneous tests were performed in the ISO 14243–1 knee joint gait simulator (EndoLab, Riedering, Germany) in 5 systems, totaling 15 components ([Figures 1 ] and [2 ]).
Fig. 1 Representation of the individual test chamber.
Fig. 2 Qualitative virtual analysis of the wear.
In the simulations, the implant was attached to the extension device. A cyclical flexion-extension variation from 0° to 58° was applied ([Table 1 ]). An axial force ranging from 168 N to 2,600 N was also applied, depending on the degree of flexion, simulating a normal human walk ([Table 1 ]). The tibial base was free to accommodate to the femoral component under the influence of applied contact forces, with this movement having all degrees of freedom, except the flexion-extension angle, which followed the specified cyclic variation. With this simulation, the applied contact force actions were: axial force, anteroposterior (AP) force, and tibial rotation torque. The femoral and tibial metal components, as well as the polyethylene, were immersed in a fluid medium simulating human synovial fluid throughout the test, which was carried out in a controlled environment, simulating the physiological conditions.
Table 1
Parameter
Values according to ISO 14243–1
Flexion/Extension
0° to 58°
Axial force
168 N to 2,600 N
Anteroposterior force
-265 N to 110 N
Torque
−1 Nm to 6 Nm
Frequency
1 Hz
Test fluid
Calf serum
Movement restriction - anteroposterior* (contrary to the positive anteroposterior movement)
9.3 N/mm
Movement restriction - anteroposterior* (contrary to the negative anteroposterior movement)
44 N/mm
Restriction of tibial rotation**
0.36 Nm/°
The wear assessment followed the ISO 14243–2:2009 norms, with 10 million cycles and measurements taken at every millionth cycle. In accordance with the aforementioned norms, the wear was assessed by analyzing the loss of mass.
Results
After 10 million cycles, the qualitative analysis of the polyethylene surface showed an appearance of regular wear, with polished and matte areas ([Figure 2 ]). This regular wear pattern indicates an intrinsic stability of the prosthetic components ([Figure 3 ]).
Fig. 3 X-UHMWPE insert wear versus number of cycles.
The quantitative result of mass wear for every millionth cycle is found in [Table 1 ]. [Figure 3 ] expresses the results mentioned in [Table 2 ].
Table 2
Coupling
1.1
1.2
1.3
Mass
Mass
Mass
Cycles (million)
X-UHMWPE 1.1 insert (g)
X-UHMWPE 1.1 insert (mg)
X-UHMWPE 1.2 insert (g)
X-UHMWPE 1.2 insert (mg)
X-UHMWPE 1.3 insert (g)
X-UHMWPE 1.3 insert (mg)
0.0
44.71242
0.00
44.48231
0.00
44.66329
0.00
0.5
44.70981
3.91
44.48121
2.41
44.65973
4.86
1.0
44.71073
4.24
44.48138
3.48
44.66000
5.84
2.0
44.71047
6.47
44.48102
5.81
44.65979
8.01
3.0
44.71210
6.71
44.48202
6.69
44.66085
8.83
4.0
44.71149
9.18
44.47986
10.70
44.65949
12.06
5.0
44.71378
9.30
44.48183
11.14
44.66343
10.52
6.0
44.71097
14.74
44.47747
18.13
44.65948
17.09
7.0
44.71004
15.47
44.47531
20.08
44.65842
17.95
7.5
44.71357
17.77
44.47810
23.13
44.66167
20.54
8.0
44.70607
22.38
44.47417
24.17
44.65625
23.06
9.0
44.70556
24.50
44.47504
24.90
44.65551
25.40
10.0
44.70660
25.23
44.47497
26.75
44.65645
26.24
The mean wear rate was 2.56 mg per millionth cycle, which was determined after 10 million cycles ([Figure 1 ]).
Discussion
Total knee arthroplasty aims to promote pain relief and improve function in a lasting way. However, surgery can fail for a number of reasons, such as loosening of the components, infection, instability, and persistent pain, for example.[11 ] In order to reduce the wear of the polyethylene and consequently the production of debris, a tibial component with a rotating platform was created, in which the polyethylene can move rotationally over the tibial component, hypothetically reducing its friction and wear.[6 ]
[7 ]
[12 ]
In a study,[11 ] the authors state that prostheses with ultracongruent rotational support have the advantage of standardizing the contact pressures between components, thus reducing the formation of polyethylene particles and, consequently, osteolysis, in addition to the better adaptation of the extensor mechanism to possible imperfections in the rotational positioning of the tibial component.[11 ] An in-vivo video-fluoroscopic study, followed by three-dimensional reconstruction of the images obtained, comparing prostheses with fixed and mobile bases, with the same origin and design, showed that the femorotibial contact surface is twice as large in prostheses with rotational support when compared to those with fixed support.[13 ] In this study, the authors noted that the good results were similar in both models, but, both objectively and subjectively, the mobile platform was judged to be the closest to the normal knee.[13 ]
The high durability of prostheses with a rotating polyethylene component is well elucidated in the literature, and the prostheses can last for more than 20 years in 97.7% of cases.[14 ]
Schmidt et al[15 ] studied the wear rate of polyethylene in different models of prostheses already commercialized and established in the market, and they found values of volumetric wear that ranged from 1.9 mg/mc to 14.6 mg/mc. In the present study, the result obtained of 2.67 mg/mc of volumetric wear after 10 million test cycles, compared to the values found by Schmidt et al,[15 ] demonstrated that the Rotaflex system approaches the lowest rate found (1.9 mg/mc). In addition, the 2.67 mg/mc of wear of the polyethylene component measured in this Brazilian prosthesis obtained a wear resistance performance 5.47 times higher than the maximum published values.
Conclusion
The wear of the polyethylene of the evaluated prosthesis was minimal after the tests performed with safety limits higher than those recommended by biomechanical engineering.