Klin Monbl Augenheilkd 2016; 233(06): 722-726
DOI: 10.1055/s-0042-108653
Klinische Studie
Georg Thieme Verlag KG Stuttgart · New York

Intrakorneale Ringsegmente bei Keratektasien – Zwischenergebnisse und potenzielle Komplikationen

Intracorneal Ring Segments to Treat Keratectasia – Interim Results and Potential Complications
M. El-Husseiny
1   Klinik für Augenheilkunde, Universitätsklinikum des Saarlandes, Homburg
,
L. Daas
1   Klinik für Augenheilkunde, Universitätsklinikum des Saarlandes, Homburg
,
A. Langenbucher
2   Experimentelle Ophthalmologie, Universitätsklinikum des Saarlandes, Homburg
,
B. Seitz
1   Klinik für Augenheilkunde, Universitätsklinikum des Saarlandes, Homburg
› Author Affiliations
Further Information

Publication History

eingereicht 15 March 2016

akzeptiert 12 May 2016

Publication Date:
17 June 2016 (online)

Zusammenfassung

Hintergrund und Ziele: Durch die minimalinvasive Operationstechnik mittels Femtosekundenlaser und die potenzielle Reversibilität stellen intrakorneale Ringsegmente (ICRS) eine zusätzliche reliable Methode zur Erweiterung des Spektrums der stadiengerechten Therapie des Keratokonus bei Kontaktlinsenintoleranz sowie bei der Behandlung von Post-LASIK-Keratektasie und pelluzider marginaler Degeneration (PMD) dar. Methoden: Von August 2011 bis September 2015 wurden im Homburger Zentrum für refraktive Chirurgie konsekutiv 84 Augen von 69 Patienten mit ICRS versorgt. Bei 74 Augen implantierten wir INTACS-SK (Intacs, Addition Technology, Inc.) und bei 10 Augen KeraRinge SI6 (Mediphacos Ltda.). Die Indikation darf nur nach einer hochspezifischen Diagnostik bei Hornhäuten mit einer narbenfreien optischen Achse erfolgen. Die Tunnelkreation sollte heute ausschließlich mit dem Femtosekundenlaser erfolgen, um intra- und postoperative Komplikationen zu vermeiden. Ergebnisse: Drei Monate postoperativ zeigten sich in der INTACS-SK-Gruppe ein Anstieg des mittleren unkorrigierten Visus (logMAR) von 0,99 ± 0,35 auf 0,29 ± 0,17. Bei der KeraRing SI6-Gruppe zeigte sich in der 3-Monats-Kontrolle ein Anstieg des mittleren unkorrigierten Visus (logMAR) von 0,93 ± 0,4 auf 0,41 ± 0,34. Die K-Werte wurden in beiden Gruppen flacher. Schlussfolgerung: Durch Implantation der ICRS können die unkorrigierte und korrigierte Sehschärfe verbessert werden. Die Progression der Ektasie scheint gehemmt zu sein. Dabei sind engmaschige Nachkontrollen zur frühzeitigen Erkennung von Komplikationen notwendig. Allerdings sind größere Fallserien und eine längere postoperative Beobachtungszeit erforderlich. Die Komplikationen nach ICRS sind bei strenger Patientenauswahl und moderner Operationstechnik heute selten.

Abstract

Background and Purpose: Implantation of intracorneal ring segments (ICRS) using a femtosecond laser is a minimally invasive procedure which represents a reliable option to widen the spectrum of the stage-related therapy of keratoconus in patients with contact lens intolerance, or with post-LASIK-keratectasia or pellucide marginale degeneration (PMD). Methods: From August 2011 to September 2015, 84 eyes of 69 patients were implanted with ICRS. 74 eyes were implanted with INTACS-SK (Intacs, Addition Technology, Inc.), and 10 eyes were implanted with KeraRing SI6 (Mediaphacos Ltda.). The indication is taken in the presence of a clear central cornea and the patients had to fulfill the corneal diagnostic tests required for implantation. Tunnel creation should nowadays only be carried out by femtosecond laser, in order to avoid intra- and postoperative complications. Results: Three months after surgery, the INTACS-SK group showed an increase in uncorrected distance visual acuity (logMAR) from 0.99 ± 0.35 to 0.29 ± 0.17. The KeraRing SI6 group showed an increase in uncorrected distance visual acuity (logMAR) from 0.93 ± 0.4 to 0.41 ± 0.34. The keratometric values were reduced in both groups. Conclusion: Uncorrected and corrected distance visual acuity can be improved by implantation of the ICRS. Progression of ectasia seems to be retarded. Consistent follow-up visits at close intervals are necessary to identify complications at an early stage. However, larger case series and a longer postoperative observation period are required. Complications after ICRS implantation are rare, due to strict patient selection and modern surgical techniques.

 
  • Literatur

  • 1 El-Husseiny M, Tsintarakis T, Eppig T et al. Intrakorneale Ringsegmente beim Keratokonus. Ophthalmologe 2013; 110: 823-826
  • 2 Rabinowitz YS. INTACS for keratoconus. Int Ophthalmol Clin 2006; 46: 91-103
  • 3 Piñero DP, Alio JL. Intracorneal ring segments in ectatic corneal disease – a review. Clin Experiment Ophthalmol 2010; 38: 154-167
  • 4 Burris TE, Baker PC, Ayer CT et al. Flattening of central corneal curvature with intrastromal corneal rings of increasing thickness: an eyebank eye study. J Cataract Refract Surg 1993; 19: 182-187
  • 5 Ertan A, Colin J. Intracorneal rings for keratoconus and keratectasia. J Cataract Refract Surg 2007; 33: 1303-1314
  • 6 Oezertuerk Y, Sari ES, Kubaloglu A et al. Comparison of deep anterior lamellar keratoplasty and intrastomal corneal ring segment implantation in advanced keratoconus. J Cataract Refract Surg 2012; 38: 324-332
  • 7 Tunc Z, Helvacioglu F, Sencan S. Evaluation of intrastromal corneal ring segments for treatment of post-LASIK ectasia patients with a mechanical implantation technique. Indian J Ophthalmol 2011; 59: 437-443
  • 8 Alió JL, Artola A, Ruiz-Moreno JM et al. Changes in keratoconic corneas after intracorneal ring segment explantation and reimplantation. Ophthalmology 2004; 111: 747-751
  • 9 Miranda D, Sartori M, Francesconi C et al. Intrastromal corneal ring segments for severe keratoconus. J Refract Surg 2003; 19: 645-653
  • 10 Grabner G, Ruckhofer J, Tratter C et al. Der Intrastromale Corneale Ring (KeraVision® Ring, ICR®, ICRS®). Eine moderne Methode zur Korrektur der geringen Myopie. Wien Med Wochenschr 1997; 147: 309-321
  • 11 Ruckhofer J, Stoiber J, Alzner E et al. One year results of European Multicenter Study of intrastromal corneal ring segments. Part 2: complications, visual symptoms, and patient satisfaction. J Cataract Refract Surg 2001; 27: 287-296
  • 12 Daxer A. CISIS: Reversible chirurgische Behandlung hoher Kurzsichtigkeiten und des Keratokonus. Spektrum Augenheilkd 2009; 23: 39-41
  • 13 Chhadva P, Yesilirmak N, Cabot F et al. Intrastromal corneal ring segment explantation in patients with keratoconus: causes, technique, and outcomes. J Refract Surg 2015; 31: 392-397
  • 14 Alió JL, Vega-Estrada A, Esperanza S et al. Intrastromal corneal ring segments: how successful is the surgical treatment of keratoconus?. Middle East Afr J Ophthalmol 2014; 21: 3-9
  • 15 Fernández-Velázquez FJ, Fernández-Fidalgo MJ. Feasibility of custom-made hydrogel contact lenses in keratoconus with previous implantation of intracorneal ring segments. Cont Lens Anterior Eye 2015; 38: 351-356