CC BY 4.0 · Eur J Dent 2023; 17(04): 1029-1036
DOI: 10.1055/s-0042-1758796
Original Article

Evaluation of Fitness and Accuracy of Milled and Three-Dimensionally Printed Inlays

1   Department of Conservative Dentistry, School of Dentistry, Wonkwang University, Iksan, Republic of Korea
,
Jeong Mi Kim
2   Wonkwang University Dental Hospital, Central Dental Laboratory, Iksan, Republic of Korea
,
Yoorina Choi
1   Department of Conservative Dentistry, School of Dentistry, Wonkwang University, Iksan, Republic of Korea
,
1   Department of Conservative Dentistry, School of Dentistry, Wonkwang University, Iksan, Republic of Korea
› Institutsangaben
Funding This study was supported by Wonkwang University in 2021.
 

Abstract

Objective This article compares and evaluates the marginal and internal fitness and three-dimensional (3D) accuracy of class II inlays fabricated using Tescera (TS) resin, milling of hybrid and zirconia blocks, and 3D printing with NextDent C&B.

Materials and Methods Fifty-two mesio-occlusal inlays were fabricated using conventional method with TS, milling of Lava Ultimate (LU), milling of Zolid Fx multilayer (ZR), and 3D printing (n = 13 each). The marginal and internal fitness were evaluated at six points in the mesio-distal section of a replica under a digital microscope (160× magnification), and the accuracy was evaluated using 3D software. Analyses were conducted using t-test, one-way analysis of variance (ANOVA) and two-way ANOVA, while Duncan's multiple range test was used for post hoc analyses (α = 0.05).

Results The marginal and internal fitness of the 3D and ZR were significantly superior to that of the TS and LU. For LU, ZR, and 3D, a significant discrepancy between the marginal gap and internal gap was observed (p < 0.05). On evaluating accuracy, trueness was significantly higher in ZR than in TS and LU; precision was significantly higher in 3D and ZR than in TS and LU (p < 0.05).

Conclusion The marginal and internal fitness and the accuracy of TS, ZR, and 3D were within the clinically acceptable range. The marginal and internal fitness and accuracy of 3D were better than those of TS and LU, which are commonly used in dentistry. There is immense potential for using 3D-printed inlays in routine clinical practice.


#

Introduction

Computer-aided design/computer-aided manufacturing (CAD/CAM) was introduced in dentistry in the 1980s and has an extensive range of applications.[1] It is divided into subtractive (milling) and additive (three-dimensional [3D] printing and rapid prototyping) methods depending on the process of fabrication.[2] [3] Compared with additive method, the disadvantages of subtractive method include wastage of material, frequent replacement of equipment due to wear, and higher possibility of defects due to poor machinability when working on high-strength blocks.[1] [3] [4] Contrarily, the additive method has no limitations on the output size, can be used to fabricate complicated designs, and causes no material waste. Hence, it is preferred in dentistry.[5]

Among the 3D printing methods, selective laser sintering, selective laser melting, stereolithography (SLT), and digital light processing (DLP) are commonly used in dentistry.[6] SLT and DLP are more frequently used due to small equipment size, excellent cost-effectiveness, and ability to print high-resolution, sophisticated outputs; among these, DLP is more popular due to its faster printing speed than SLT.[1] [3] [5]

Marginal gap (MG) and internal gap (IG) are critical factors affecting the long-term success of dental prostheses. Larger MG increases the risk of leakage from dissolution and wear of cement caused by physical fatigue and chemical corrosion; the resultant subgingival plaque and bacterial infiltration may lead to hypersensitivity, secondary caries, and prosthesis failure. Poor internal fitness leads to decreased retention and support of the prosthesis and increased cement thickness, thereby lowering the fracture resistance of the prosthesis.[7] [8]

Although difficult to obtain in clinical practice, the American Dental Association Specification No. 8 stipulates that a gap of 25 to 40 µm is appropriate for optimal fitness of fixed prostheses; however, McLean and von Fraunhofer suggested a clinically acceptable limit of 120 µm.[9] Many studies have reported that an MG and IG of < 120 µm is clinically satisfactory.[10] The accuracy of 3D-printed prostheses has been reported by various studies to be within the clinically acceptable range compared with those made by milling or the traditional lost wax technique.[3] [4] One study suggested an average MG of 50 to 60 and 150 to 168 µm in ceramic crowns and mesio-occluso-distal ceramic inlays fabricated by CAD/CAM[11]; another study reported that the MG increased with more complex cavity geometry, such as with cusp capping.[12]

Nondestructive methods for evaluating MG and IG include micro-computed tomography and the replica technique.[13] Of these, the replica technique has some disadvantages, including difficulties in identifying the margins, potential errors when the silicone impression is torn or cut, and a limited number of measurement points[14] [15] [16] [17]; additionally, the accuracy may be affected by the type of silicone material, measurement method, and number of measurement points used.[18] Nevertheless, the replica technique is generally accepted as an easy, reliable, and less expensive measurement method for MG and IG.[17] [19] [20]

For evaluating accuracy, the trueness and precision of the measured data were evaluated against the reference data obtained from 3D scanning in accordance with the International Organization for Standardization 5725–1 standard, and the results were expressed as root mean square (RMS). Lower RMS values indicate superior trueness and precision. Previous studies have suggested that RMS values < 10 µm indicate excellent fitness, whereas RMS values > 50 µm indicate poor fitness.[21] [22] Although trueness does not measure the fitness, it is used to evaluate the clinical quality by comparing the difference between reference data and measured data from the fabricated prosthesis.[23] The measured data may be affected by scanner accuracy and data analysis method; however, 3D accuracy is reported to be a more reliable method than two-dimensional evaluation since a computer could be used to measure the 3D distance from the measurement points and additional random points.[21]

With an increase in the use of 3D printing in dentistry, many studies have investigated the MG and IG or accuracy of 3D-printed provisional crowns, orthodontic models, abutment models, and full dentures. Nonetheless, evidence is scarce on the fitness and accuracy of 3D-printed resin inlays.

In this study, we compared class II inlays made by conventional method using Tescera resin (TS), milling Lava Ultimate (LU), and Zolid Fx multilayer (ZR) blocks, which are widely used in clinical practice for esthetic restoration, and 3D printing using NextDent C&B resin. The null hypotheses tested were: (1) there is no difference in the marginal and internal fitness and accuracy according to the material, and (2) there is no difference according to the measurement point between IG and MG in each group.


#

Materials and Methods

A mesial class II inlay cavity (occlusal box depth = 2.5 mm and isthmus width = 3.0 mm; proximal box with mesiodistal width = 1.2 mm, buccolingual width = 3.0 mm, and occlusogingival depth = 4.0 mm; and convergence angle = 8 degrees) was prepared on a mandibular right first molar resin tooth (Nissin Dental Products, Kyoto, Japan) using a high-speed handpiece with a diamond burr (TF-31, MANI, Tochigi, Japan). Silicon impression material (Honigum Light Body, Regular Body, DMG, Hamburg, Germany) was used to make an impression of the prepared cavity. Using high-strength dental stone (Fujirock EP, GC Europe), 52 master dies ([Fig. 1]) were made. Subsequently, 52 class II inlays were fabricated using the four different materials in accordance with the manufacturer's protocol. The experimental groups were conventional preparation using TS, milling of LU, milling of ZR, and 3D printing with NextDent C&B resin ([Table 1]). The number of inlay specimens in each group was 13.

Zoom Image
Fig. 1 Inlay cavity preparation and fabricated inlay specimen. (A) Mandibular right first molar with a prepared class II mesio-occlusal inlay cavity, (B) master die, (C) image of the master die saved as standard tessellation language file of master die, and (D) inlay specimen fabricated by 3D printing. 3D, three-dimensional.
Table 1

Materials assessed

Group

Brand name

Manufacturer

Composition

Monomer

Filler (mass%)

TS

Tescera

Bisco, Chicago, IL, USA

Ethoxylated bis-GMA,

UDMA

Glass frit

amorphous silica

< 80

< 25

LU

Lava Ultimate

3M ESPE, St. Paul, MN, USA

Bis-GMA, UDMA, Bis-EMA, TEGDMA

SiO2 (20 nm), ZrO2 (4–11 nm), aggregated ZrO2/SiO2 cluster (SiO2 = 20 nm, ZrO2 = 4–11 nm)

80

ZR

Zolid Fx multilayer

Amann Girrbach, Pforzheim, Germany

ZrO2 + HfO2 + Y2O3

Y2O3

HfO2

Al2O3

Other oxides

≥ 99.0

8.5–9.5

≤ 5

≤ 0.5

≤ 1

3D

NextDent C&B

3D Systems, Soesterberg, Netherlands

Methacrylic oligomers,

phosphine oxides

Abbreviations: 3D, three-dimensional printing; Bis-EMA, ethoxylated bisphenol-A dimethacrylate; Bis-GMA, bisphenol A diglycidylether methacrylate; LU, milling with Lava Ultimate; TEGDMA, triethylene glycol dimethacrylate; TS, Tescera resin; UDMA, urethane dimethacrylate; ZR, milling with Zolid Fx multilayer blocks.


The TS group specimens were prepared as follows: a resin separator (TS die separator, Bisco, Chicago, Illinois, United States) was applied to the cavity in the master die. After drying, TS resin (Bisco) was placed directly inside the cavity to prepare the external morphology of the inlay, followed by light curing for 2 minutes in a TS ATL Light Cup (Bisco). After adding enough water to submerge the specimen and one spoon of dissolved oxygen scavenger (TS Oxygen Scavenger Plus, Bisco) to the TS ATL Heat Cup, the specimen was placed and heat-cured for 25 minutes at 130°C in accordance with the manufacturer's protocol.

In the LU, ZR, and 3D groups, class II inlay specimens were prepared as follows: after scanning the master die using an intraoral scanner (Trios 3, 3shape, Copenhagen, Denmark), the Ceramill Mind (Amann Girrbach, Pforzheim, Germany) program was used to design the class II inlay specimens with 30 µm of cement space 1.0 mm above the margin. The specimens were designed and fabricated according to each manufacturer's protocol.

In the LU and ZR groups, CAD/CAM milling equipment (Ceramill Motion 2, Amann Girrbach) was used to mill the hybrid blocks (LU, 3M ESPE, St. Paul, Minnesota, United States) and zirconia blocks (ZR multilayer, Amann Girrbach). The milled ZR specimens were placed in a heating furnace (Ceramill Therm, Amannn Girrbach), and the temperature was increased from 20°C to 1,450°C at a rate of 12°C/min. Subsequently, the specimens were left at 1,450°C for 1 hour before being cooled.

In the 3D-printed group, NextDent C&B resin (3D Systems, Soesterberg, Netherlands) and a DLP 3D printer (Bio3D L12 Dental Professional, BIO3D, Seoul, Republic of Korea) were used to print the specimens with a layer thickness of 50 µm, printing rate of 10 mm/h, and building angle of 180 degrees. Subsequently, the specimens were placed inside a light-curing unit (LC-3Dprint Box, 3D Systems) and cured for 30 minutes.

The replica for the evaluation of marginal and internal fitness was prepared as follows: after injecting light body silicone (Aquasil Ultra XLV, Dentsply DeTrey, Konstanz, Germany) into the cavity of the master die ([Fig. 1A]) and inserting the inlay specimen within its position in the cavity, a universal testing machine (Instron 3345, Instron Corporation, Norwood, Massachusetts, United States) was used to apply a load of 5 N for 5 minutes until the silicone hardened.[24] On the mesiodistal portion of the replica that passed through the central fossa, the MG was measured at two points, cervical margin (A) and occlusal margin (F), and the IG was measured at four points, the midpoints of cervical floor (B), axial wall (C), pulpal wall of the occlusal box (D), and axial wall of occlusal box (E), under a digital microscope (KH-7700, Hirox, Tokyo, Japan) at 160× magnification ([Fig. 2]).

Zoom Image
Fig. 2 (A) Schematic representation of measurement points for the marginal and internal discrepancy in the sagittal section of silicone replica: marginal areas = A, F; internal areas = B, C, D, E. (B) Measurement of the internal gap of Zolid Fx multilayer by digital microscope at 160× magnification (light green: light body silicone, green: heavy body silicone).

After evaluating marginal and internal fitness, accuracy was assessed. The standard tessellation language (STL) files obtained by scanning the inlay cavity of the master die were used as the reference data for the TS group, while STL files designed by CAD software were used as reference data for the other three groups. Measurement data were generated as STL files by scanning the prepared inlay specimens with a model scanner (Identica Blue, Medit, Seoul, Republic of Korea). The generated data files were evaluated for trueness and precision using 3D software (Geomagic Verify, 3D Systems Inc., United States). Trueness was evaluated by comparing the discrepancy between the reference and measurement data files in each experimental group (n = 13). Precision was calculated by comparing pairs of measurement data files in each experimental group (n = 78).

To determine the deviation between the reference and measurement data files, two data files were superimposed to measure trueness and precision. The RMS value was used to quantify the trueness and precision, and was calculated as follows:

Zoom Image

where x 1 is the reference data at measurement point i, x 2 is the measurement data at measurement point i, and n is the total number of measurement points. A smaller RMS value indicated a higher accuracy.

For visualizing the differences between the reference and measurement data, 3D deviation image analysis was used to obtain a 3D color map. Areas where the measured data differed within ± 10 µm from the reference data were displayed in green, areas where the measured data differed greater than +10 µm were displayed in yellow to red, and areas smaller than –10 µm were displayed in blue to dark blue.

All data were evaluated using SPSS ver. 20.0 statistical software (IBM Corp., Armonk, New York, United States). The interaction effect between the material (TS, LU, ZR, and 3D) and measurement point (MG and IG) was analyzed using two-way analysis of variance (ANOVA). The difference in marginal and internal fitness according to material used was analyzed using one-way ANOVA. The discrepancy between MG and IG in each experimental group was analyzed using t-test (α = 0.05). Trueness and precision, representing accuracy, were analyzed by one-way ANOVA, and post hoc analyses were performed using Duncan's new multiple range test (α = 0.05).


#

Results

[Table 2] shows the results obtained in the evaluation of marginal and internal fitness in the four experimental groups. On comparing MG among the four experimental groups (mean of the values at A and F), the MG of ZR and 3D was statistically significantly smaller than that of TS and LU (p < 0.05). All groups were considered clinically acceptable because MG of all four groups was < 120 μm.

Table 2

Discrepancy in each group at the measurement points

Measurement

point

Material

TS

LU

ZR

3D

MG

107.65 ± 67.37a

118.54 ± 45.54a

58.35 ± 14.88b

53.77 ± 16.29b

IG

103.1 9 ± 43.80a

168.81 ± 42.67b

95.69 ± 13.34a

82.02 ± 8.32a

Total sum value

104.68 ± 48.50b

152.05 ± 36.96c

83.24 ± 11.78a

72.60 ± 7.29a

Abbreviations: 3D, three-dimensional printing; IG, internal gap; LU, Lava Ultimate; MG, marginal gap; TS, Tescera; ZR, zirconia.


Note: Values are presented as mean ± standard deviation; unit, µm. Within the same row, the same lowercase letters indicate no statistically significant difference in one-way analysis of variance (ANOVA) at α = 0.05.


On comparing the IG between the groups (mean of the values at B, C, D, and E), the IG of LU (168.81 ± 42.67) was statistically significantly larger than that of TS, ZR, and 3D and larger than the clinically acceptable range (p < 0.05). On comparing the mean MG and IG, the values were statistically significantly smaller in ZR and 3D than in TS and LU (p < 0.05).

On evaluating the discrepancy between MG and IG in each experimental group ([Table 2]), there was a statistically significant discrepancy between MG and IG in LU, ZR, and 3D (p < 0.05; [Fig. 3]).

Zoom Image
Fig. 3 Marginal and internal discrepancies at each measurement point according to materials. The same letters indicate no statistically significant difference by one-way analysis of variance (ANOVA) at α = 0.05. *Statistically significant difference by t-test at α = 0.05. 3D, three-dimensional printing; IG, internal gap; LU, Lava Ultimate; MG, marginal gap; TS, Tescera; ZR, zirconia.

A two-way ANOVA showed no interaction effect between material and measurement point (p > 0.05). There was a difference in the effect of materials and measurement points on fitness, as explained by 47.8%. The material (7.93%) was a more important factor in determining fitness than the measurement point (35.66%) ([Table 3]).

Table 3

Two-way ANOVA considering two factors with the interaction term

Source

Sum of

squares

df

Mean

squares

F

p

Corrected model

121461.523

7

17351.646

12.550

0.000

Material

90653.483

3

30217.828

21.857

0.000

Measurement point

20167.578

1

20167.578

14.587

0.000

Material[a]

(measurement point)

10640.463

3

3546.821

2.565

0.059

Error

132725.038

96

1382.552

Corrected total

254186.562

103

Abbreviations: ANOVA, analysis of variance; df, degrees of freedom.


Note: R 2 = 0.478.


a Significantly different by two-way ANOVA at α = 0.05.


Regarding trueness, a factor of accuracy, the RMS value of ZR was statistically significantly lower than that of TS and LU (p < 0.05), while no significant difference was identified in the trueness of 3D compared with that of TS, LU, and ZR (p > 0.05; [Table 4]).

Table 4

Trueness (RMS values) of the inner surface of the inlays made of four fabrication methods

Group

Trueness (RMS ± SD)

F

p

TS

47.06 ± 14.73b

4.360

0.009

LU

48.97 ± 17.49b

ZR

33.26 ± 7.24a

3D

40.86 ± 4.78ab

Abbreviations: 3D, three-dimensional printing; LU, Lava Ultimate; RMS, root mean square; SD, standard deviation; TS, Tescera; ZR, zirconia.


Note: Unit, µm. The same lowercase letters were not significantly different by one-way analysis of variance (ANOVA) at α = 0.05.


The 3D color deviation maps in [Fig. 4] show the 3D surface deviation between the reference and measurement data. Yellow to red areas indicate positive deviation and blue to dark blue areas indicate negative deviation. The 3D color deviation map of trueness and precision of LU showed more positive and negative deviations than the other experimental groups.

Zoom Image
Fig. 4 The three-dimensional color deviation maps for trueness and precision of four groups. Positive deviation is displayed as yellow to red colors and negative deviation as blue to dark blue. 3D, three-dimensional printing; LU, Lava Ultimate; TS, Tescera; ZR, zirconia.

On evaluating precision, another component of accuracy, the RMS value of 3D and ZR was statistically significantly lower than that of TS and LU, while the RMS value of LU (65.08 ± 13.34) was larger than the clinically acceptable range and showed a significant difference compared with the other three groups (p < 0.05; [Table 5]).

Table 5

Precision (RMS values) of the inner surface of the inlays made using the four fabrication methods

Group

Precision (RMS ± SD)

F

p

TS

45.72 ± 5.68b

101.698

0.000

LU

65.08 ± 13.34c

ZR

24.66 ± 2.35a

3D

20.05 ± 2.01a

Abbreviations: 3D, three-dimensional printing; LU, Lava Ultimate; RMS, root mean square; SD, standard deviation; TS, Tescera; ZR, zirconia.


Note: Unit, µm. The same lowercase letters indicate no statistically significant difference by one-way analysis of variance (ANOVA) at α = 0.05.



#

Discussion

This study evaluated the marginal and internal fitness and accuracy of class II inlays fabricated by conventional method using TS, milling of LU and ZR, and 3D printing with NextDent C&B resin. Differences in marginal and internal fitness and accuracy were identified among the four experimental groups (p < 0.05). ZR and 3D showed significantly better marginal and internal fitness and precision than TS and LU (p < 0.05); additionally, ZR showed better trueness than the other three groups. With TS, ZR, and 3D, the marginal and internal fitness was < 120 µm and the RMS values of trueness and precision were < 50 µm; hence, these three groups were within the clinically acceptable.[23] Therefore, the null hypotheses that there would be no differences in marginal and internal fitness and accuracy according to the material between the four groups and that there would be no differences between IG and MG in each group were rejected.

LU was introduced as a material that causes less wear of the opposing dentition, less susceptibility to fracture, better machinability, and easier repair than other porcelain materials used for esthetic restorations.[20] However, a study compared the internal fitness of inlays fabricated using four different blocks reported that LU had a mean IG of 186 µm, which was larger than that of Vita Enamic and IPS e.max (p < 0.05). That is why LU had lower hardness and flexural strength than the other two ceramic blocks, as such a greater amount of material was removed during milling, leading to its poor fitness.[20]

In this study, the LU group showed the poorest marginal and internal fitness and accuracy among all groups. Although the MG of the LU group was within the clinically acceptable range, the IG and RMS value of precision were greater than the clinically acceptable range.

Although the LU and ZR specimens were fabricated by a subtractive method using the same milling machine in this study, ZR showed significantly better marginal and internal fitness and accuracy than TS and LU (p < 0.05), with the lowest RMS value of trueness among all four experimental groups. Zirconia materials were designed and milled with 25% extra size to compensate for the 20 to 25% shrinkage after sintering,[2] [25] thereby having better fitness than other milled materials.[7] Zirconia blocks were generally overmilled compared with the reference design, which could be helpful with the insufficient compensation of shrinkage.[23] And the pre-sintering density of zirconia is 40% of the final density,[26] which is softer than the LU block. Furthermore, latest CAD software programs were reportedly able to accurately compensate for shrinkage of the block during sintering.[2]

Factors that may affect the fitness of the prosthesis fabricated by milling included the material of the block, milling equipment, milling burr size, and degree of wear of the burr.[8] [20] [27] Spitznagel et al mentioned that the power and torque of the milling device should be modified according to the type of block during milling.[28] Many dental technicians fabricate dozens of crowns without replacing the burr because there are no definitive manufacturer guidelines about changing the milling tool except following tool breakdown and while milling different materials; the prosthesis fabricated without burr replacement is also considered clinically acceptable.[21] In this study, milling of LU blocks and ZR blocks was performed without burr replacement.

When designing the STL files for ZR, LU, and 3D with the CAD software, a cement space of 30 µm at 1 mm above the margin was provided as recommended by the manufacturer. However, for TS, the scanned data of the master die was converted to the STL files and used as the reference data, and the specimen was fabricated without cement space.

In this study, unlike the other three experimental groups, no cement space was provided for TS, nevertheless it showed larger MG and IG than 3D and ZR (p < 0.05). In TS, the RMS value of trueness was also larger than that of ZR, while precision was significantly larger than that of ZR and 3D (p < 0.05). TS may have shown poorer marginal and internal fitness and accuracy than ZR and 3D, fabricated by the CAD/CAM method, due to the accumulation of errors during wax blockout and manual fabrication of the specimens.[29]

On comparing MG and IG in the four experimental groups, three of the experimental groups, except TS in which fabrication was performed without cement space, showed a statistically significant discrepancy between MG and IG (p < 0.05).

In 3D, the MG and IG were the smallest among all four experimental groups and significantly smaller than that of LU and TS (p < 0.05). The precision of 3D was statistically significantly better than that of LU and TS, with the smallest RMS value (p < 0.05). The present study supports previous studies that reported better internal fitness of 3D-printed restorations than milled from resin blocks.[3] [30] [31] The excellent marginal and internal fitness may have been because the specimens were designed and printed to a larger size to compensate for resin polymerization shrinkage. The excellent precision may have been due to fewer errors among the specimens which were 3D printed simultaneously.

Flexural strengths of TS resin, Synfony resin, and LU block, which are commonly used in clinical practice for aesthetic inlays, were 84.6, 76.8, and 170.5 MPa, respectively.[32] [33] Although the flexural strength of NextDent C&B (114–119 MPa, 2.3–2.4 GPa, respectively) specimens reported by Keßler et al was lower than that of LU block, it was higher than that of TS or Synfony resin.[5] We expect to improve not only the flexural strength but also other physical property of 3D-printed materials.

This study had several limitations while evaluating the marginal and internal fitness and accuracy of each experimental group. The same burr was repeatedly used while milling ZR and LU specimens. Additionally, the TS specimens were fabricated with no cement space. Although resins for 3D printing have been approved for midterm use for up to 2 years owing to physical properties,[34] additional long-term studies on the properties of 3D-printed resin are needed to implement the use of 3D-printed inlays for permanent restoration.


#

Conclusion

Within the limitations of this current study, it was concluded that:

  1. The MG of all experimental groups was within the clinically acceptable range (< 120 µm). The trueness and precision RMS value of the ZR, 3D, and TS groups was within the clinically acceptable range < 50 µm.

  2. The marginal and internal fitness and accuracy of 3D-printed resin inlays were within the clinically acceptable range and those of 3D-printed inlays were statistically significantly better than those of inlays fabricated by TS and LU, which are widely used in dentistry; 3D-printed resin inlays have high potential for use in routine clinical practice for esthetic restoration.


#
#

Conflict of Interest

The authors declare no competing interests.

Acknowledgments

None.

Financial Interest

The authors do not have any financial interest in the companies whose materials are included in this article.


  • References

  • 1 Reymus M, Fabritius R, Keßler A, Hickel R, Edelhoff D, Stawarczyk B. Fracture load of 3D-printed fixed dental prostheses compared with milled and conventionally fabricated ones: the impact of resin material, build direction, post-curing, and artificial aging-an in vitro study. Clin Oral Investig 2020; 24 (02) 701-710
  • 2 Bousnaki M, Chatziparaskeva M, Bakopoulou A, Pissiotis A, Koidis P. Variables affecting the fit of zirconia fixed partial dentures: a systematic review. J Prosthet Dent 2020; 123 (05) 686-692.e8
  • 3 Lee WS, Lee DH, Lee KB. Evaluation of internal fit of interim crown fabricated with CAD/CAM milling and 3D printing system. J Adv Prosthodont 2017; 9 (04) 265-270
  • 4 Patzelt SB, Bishti S, Stampf S, Att W. Accuracy of computer-aided design/computer-aided manufacturing-generated dental casts based on intraoral scanner data. J Am Dent Assoc 2014; 145 (11) 1133-1140
  • 5 Keßler A, Hickel R, Ilie N. In vitro investigation of the influence of printing direction on the flexural strength, flexural modulus and fractographic analysis of 3D-printed temporary materials. Dent Mater J 2021; 40 (03) 641-649
  • 6 Park SM, Park JM, Kim SK, Heo SJ, Koak JY. Flexural strength of 3D-printing resin materials for provisional fixed dental prostheses. Materials (Basel) 2020; 13 (18) 3970
  • 7 Cunali RS, Saab RC, Correr GM. et al. Marginal and internal adaptation of zirconia crowns: a comparative study of assessment. Braz Dent J 2017; 28 (04) 467-473
  • 8 Goujat A, Abouelleil H, Colon P. et al. Marginal and internal fit of CAD-CAM inlay/onlay restorations: a systematic review of in vitro studies. J Prosthet Dent 2019; 121 (04) 590-597.e3
  • 9 McLean JW, von Fraunhofer JA. The estimation of cement film thickness by an in vivo technique. Br Dent J 1971; 131 (03) 107-111
  • 10 Kang SY, Lee HN, Kim JH, Kim WC. Evaluation of marginal discrepancy of pressable ceramic veneer fabricated using CAD/CAM system: additive and subtractive manufacturing. J Adv Prosthodont 2018; 10 (05) 347-353
  • 11 Reich S, Gozdowski S, Trentzsch L, Frankenberger R, Lohbauer U. Marginal fit of heat-pressed vs. CAD/CAM processed all-ceramic onlays using a milling unit prototype. Oper Dent 2008; 33 (06) 644-650
  • 12 Park SH, Yoo YJ, Shin YJ, Cho BH, Baek SH. Marginal and internal fit of nano-composite CAD/CAM restorations. Restor Dent Endod 2016; 41 (01) 37-43
  • 13 Alajaji NK, Bardwell D, Finkelman M, Ali A. Micro-CT evaluation of ceramic inlays: comparison of the marginal and internal fit of five and three axis cam systems with a heat press technique. J Esthet Restor Dent 2017; 29 (01) 49-58
  • 14 Homsy FR, Özcan M, Khoury M, Majzoub ZAK. Comparison of fit accuracy of pressed lithium disilicate inlays fabricated from wax or resin patterns with conventional and CAD-CAM technologies. J Prosthet Dent 2018; 120 (04) 530-536
  • 15 Paul N, Raghavendra Swamy KN, Dhakshaini MR, Sowmya S, Ravi MB. Marginal and internal fit evaluation of conventional metal-ceramic versus zirconia CAD/CAM crowns. J Clin Exp Dent 2020; 12 (01) e31-e37
  • 16 Wu J, Xie H, Sadr A, Chung KH. Evaluation of internal fit and marginal adaptation of provisional crowns fabricated with three different techniques. Sensors (Basel) 2021; 21 (03) 740
  • 17 Laurent M, Scheer P, Dejou J, Laborde G. Clinical evaluation of the marginal fit of cast crowns–validation of the silicone replica method. J Oral Rehabil 2008; 35 (02) 116-122
  • 18 Al Hamad KQ, Al Quran FA, AlJalam SA, Baba NZ. Comparison of the accuracy of fit of metal, zirconia, and lithium disilicate crowns made from different manufacturing techniques. J Prosthodont 2019; 28 (05) 497-503
  • 19 Svanborg P. A systematic review on the accuracy of zirconia crowns and fixed dental prostheses. Biomater Investig Dent 2020; 7 (01) 9-15
  • 20 Goujat A, Abouelleil H, Colon P. et al. Mechanical properties and internal fit of 4 CAD-CAM block materials. J Prosthet Dent 2018; 119 (03) 384-389
  • 21 Cho JH, Yoon HI, Han JS, Kim DJ. Trueness of the inner surface of monolithic crowns fabricated by milling of a fully sintered (Y, Nb)-TZP block in chairside CAD-CAM system for single-visit dentistry. Materials (Basel) 2019; 12 (19) 3253
  • 22 Peters MC, Delong R, Pintado MR, Pallesen U, Qvist V, Douglas WH. Comparison of two measurement techniques for clinical wear. J Dent 1999; 27 (07) 479-485
  • 23 Al Hamad KQ, Al-Rashdan RB, Al-Rashdan BA, Baba NZ. Effect of milling protocols on trueness and precision of ceramic crowns. J Prosthodont 2021; 30 (02) 171-176
  • 24 Kim KB, Kim JH, Kim WC, Kim HY, Kim JH. Evaluation of the marginal and internal gap of metal-ceramic crown fabricated with a selective laser sintering technology: two- and three-dimensional replica techniques. J Adv Prosthodont 2013; 5 (02) 179-186
  • 25 Denkena B, Breidenstein B, Busemann S, Lehr CM. Impact of hard machining on zirconia based ceramics for dental applications. Procedia CIRP 2017; 65: 248-252
  • 26 Denry I, Kelly JR. State of the art of zirconia for dental applications. Dent Mater 2008; 24 (03) 299-307
  • 27 Bayrak A, Akat B, Ocak M, Kιlιçarslan MA, Özcan M. Micro-computed tomography analysis of fit of ceramic inlays produced with different CAD software programs. Eur J Prosthodont Restor Dent 2021; 29 (03) 29
  • 28 Spitznagel FA, Boldt J, Gierthmuehlen PC. CAD/CAM ceramic restorative materials for natural teeth. J Dent Res 2018; 97 (10) 1082-1091
  • 29 Türk AG, Sabuncu M, Ünal S, Önal B, Ulusoy M. Comparison of the marginal adaptation of direct and indirect composite inlay restorations with optical coherence tomography. J Appl Oral Sci 2016; 24 (04) 383-390
  • 30 Alharbi N, Alharbi S, Cuijpers VMJI, Osman RB, Wismeijer D. Three-dimensional evaluation of marginal and internal fit of 3D-printed interim restorations fabricated on different finish line designs. J Prosthodont Res 2018; 62 (02) 218-226
  • 31 Alharbi N, Osman R, Wismeijer D. Effects of build direction on the mechanical properties of 3D-printed complete coverage interim dental restorations. J Prosthet Dent 2016; 115 (06) 760-767
  • 32 Lauvahutanon S, Takahashi H, Shiozawa M. et al. Mechanical properties of composite resin blocks for CAD/CAM. Dent Mater J 2014; 33 (05) 705-710
  • 33 Kim KS, Yoon TH, Song KY, Ahn SG. Comparison of mechanical properties in 4 indirect composite resin. J Korean Acad Prosthodont 2007; 45: 21-33
  • 34 Revilla-León M, Özcan M. Additive manufacturing technologies used for processing polymers: current status and potential application in prosthetic dentistry. J Prosthodont 2019; 28 (02) 146-158

Address for correspondence

Sujung Park, DDS, PhD
Department of Conservative Dentistry, Wonkwang University School of Dentistry and Oral Science Research Center
460 Iksandae-ro, Iksan 54538
Republic of Korea   

Publikationsverlauf

Artikel online veröffentlicht:
04. Januar 2023

© 2023. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution License, permitting unrestricted use, distribution, and reproduction so long as the original work is properly cited. (https://creativecommons.org/licenses/by/4.0/)

Thieme Medical and Scientific Publishers Pvt. Ltd.
A-12, 2nd Floor, Sector 2, Noida-201301 UP, India

  • References

  • 1 Reymus M, Fabritius R, Keßler A, Hickel R, Edelhoff D, Stawarczyk B. Fracture load of 3D-printed fixed dental prostheses compared with milled and conventionally fabricated ones: the impact of resin material, build direction, post-curing, and artificial aging-an in vitro study. Clin Oral Investig 2020; 24 (02) 701-710
  • 2 Bousnaki M, Chatziparaskeva M, Bakopoulou A, Pissiotis A, Koidis P. Variables affecting the fit of zirconia fixed partial dentures: a systematic review. J Prosthet Dent 2020; 123 (05) 686-692.e8
  • 3 Lee WS, Lee DH, Lee KB. Evaluation of internal fit of interim crown fabricated with CAD/CAM milling and 3D printing system. J Adv Prosthodont 2017; 9 (04) 265-270
  • 4 Patzelt SB, Bishti S, Stampf S, Att W. Accuracy of computer-aided design/computer-aided manufacturing-generated dental casts based on intraoral scanner data. J Am Dent Assoc 2014; 145 (11) 1133-1140
  • 5 Keßler A, Hickel R, Ilie N. In vitro investigation of the influence of printing direction on the flexural strength, flexural modulus and fractographic analysis of 3D-printed temporary materials. Dent Mater J 2021; 40 (03) 641-649
  • 6 Park SM, Park JM, Kim SK, Heo SJ, Koak JY. Flexural strength of 3D-printing resin materials for provisional fixed dental prostheses. Materials (Basel) 2020; 13 (18) 3970
  • 7 Cunali RS, Saab RC, Correr GM. et al. Marginal and internal adaptation of zirconia crowns: a comparative study of assessment. Braz Dent J 2017; 28 (04) 467-473
  • 8 Goujat A, Abouelleil H, Colon P. et al. Marginal and internal fit of CAD-CAM inlay/onlay restorations: a systematic review of in vitro studies. J Prosthet Dent 2019; 121 (04) 590-597.e3
  • 9 McLean JW, von Fraunhofer JA. The estimation of cement film thickness by an in vivo technique. Br Dent J 1971; 131 (03) 107-111
  • 10 Kang SY, Lee HN, Kim JH, Kim WC. Evaluation of marginal discrepancy of pressable ceramic veneer fabricated using CAD/CAM system: additive and subtractive manufacturing. J Adv Prosthodont 2018; 10 (05) 347-353
  • 11 Reich S, Gozdowski S, Trentzsch L, Frankenberger R, Lohbauer U. Marginal fit of heat-pressed vs. CAD/CAM processed all-ceramic onlays using a milling unit prototype. Oper Dent 2008; 33 (06) 644-650
  • 12 Park SH, Yoo YJ, Shin YJ, Cho BH, Baek SH. Marginal and internal fit of nano-composite CAD/CAM restorations. Restor Dent Endod 2016; 41 (01) 37-43
  • 13 Alajaji NK, Bardwell D, Finkelman M, Ali A. Micro-CT evaluation of ceramic inlays: comparison of the marginal and internal fit of five and three axis cam systems with a heat press technique. J Esthet Restor Dent 2017; 29 (01) 49-58
  • 14 Homsy FR, Özcan M, Khoury M, Majzoub ZAK. Comparison of fit accuracy of pressed lithium disilicate inlays fabricated from wax or resin patterns with conventional and CAD-CAM technologies. J Prosthet Dent 2018; 120 (04) 530-536
  • 15 Paul N, Raghavendra Swamy KN, Dhakshaini MR, Sowmya S, Ravi MB. Marginal and internal fit evaluation of conventional metal-ceramic versus zirconia CAD/CAM crowns. J Clin Exp Dent 2020; 12 (01) e31-e37
  • 16 Wu J, Xie H, Sadr A, Chung KH. Evaluation of internal fit and marginal adaptation of provisional crowns fabricated with three different techniques. Sensors (Basel) 2021; 21 (03) 740
  • 17 Laurent M, Scheer P, Dejou J, Laborde G. Clinical evaluation of the marginal fit of cast crowns–validation of the silicone replica method. J Oral Rehabil 2008; 35 (02) 116-122
  • 18 Al Hamad KQ, Al Quran FA, AlJalam SA, Baba NZ. Comparison of the accuracy of fit of metal, zirconia, and lithium disilicate crowns made from different manufacturing techniques. J Prosthodont 2019; 28 (05) 497-503
  • 19 Svanborg P. A systematic review on the accuracy of zirconia crowns and fixed dental prostheses. Biomater Investig Dent 2020; 7 (01) 9-15
  • 20 Goujat A, Abouelleil H, Colon P. et al. Mechanical properties and internal fit of 4 CAD-CAM block materials. J Prosthet Dent 2018; 119 (03) 384-389
  • 21 Cho JH, Yoon HI, Han JS, Kim DJ. Trueness of the inner surface of monolithic crowns fabricated by milling of a fully sintered (Y, Nb)-TZP block in chairside CAD-CAM system for single-visit dentistry. Materials (Basel) 2019; 12 (19) 3253
  • 22 Peters MC, Delong R, Pintado MR, Pallesen U, Qvist V, Douglas WH. Comparison of two measurement techniques for clinical wear. J Dent 1999; 27 (07) 479-485
  • 23 Al Hamad KQ, Al-Rashdan RB, Al-Rashdan BA, Baba NZ. Effect of milling protocols on trueness and precision of ceramic crowns. J Prosthodont 2021; 30 (02) 171-176
  • 24 Kim KB, Kim JH, Kim WC, Kim HY, Kim JH. Evaluation of the marginal and internal gap of metal-ceramic crown fabricated with a selective laser sintering technology: two- and three-dimensional replica techniques. J Adv Prosthodont 2013; 5 (02) 179-186
  • 25 Denkena B, Breidenstein B, Busemann S, Lehr CM. Impact of hard machining on zirconia based ceramics for dental applications. Procedia CIRP 2017; 65: 248-252
  • 26 Denry I, Kelly JR. State of the art of zirconia for dental applications. Dent Mater 2008; 24 (03) 299-307
  • 27 Bayrak A, Akat B, Ocak M, Kιlιçarslan MA, Özcan M. Micro-computed tomography analysis of fit of ceramic inlays produced with different CAD software programs. Eur J Prosthodont Restor Dent 2021; 29 (03) 29
  • 28 Spitznagel FA, Boldt J, Gierthmuehlen PC. CAD/CAM ceramic restorative materials for natural teeth. J Dent Res 2018; 97 (10) 1082-1091
  • 29 Türk AG, Sabuncu M, Ünal S, Önal B, Ulusoy M. Comparison of the marginal adaptation of direct and indirect composite inlay restorations with optical coherence tomography. J Appl Oral Sci 2016; 24 (04) 383-390
  • 30 Alharbi N, Alharbi S, Cuijpers VMJI, Osman RB, Wismeijer D. Three-dimensional evaluation of marginal and internal fit of 3D-printed interim restorations fabricated on different finish line designs. J Prosthodont Res 2018; 62 (02) 218-226
  • 31 Alharbi N, Osman R, Wismeijer D. Effects of build direction on the mechanical properties of 3D-printed complete coverage interim dental restorations. J Prosthet Dent 2016; 115 (06) 760-767
  • 32 Lauvahutanon S, Takahashi H, Shiozawa M. et al. Mechanical properties of composite resin blocks for CAD/CAM. Dent Mater J 2014; 33 (05) 705-710
  • 33 Kim KS, Yoon TH, Song KY, Ahn SG. Comparison of mechanical properties in 4 indirect composite resin. J Korean Acad Prosthodont 2007; 45: 21-33
  • 34 Revilla-León M, Özcan M. Additive manufacturing technologies used for processing polymers: current status and potential application in prosthetic dentistry. J Prosthodont 2019; 28 (02) 146-158

Zoom Image
Fig. 1 Inlay cavity preparation and fabricated inlay specimen. (A) Mandibular right first molar with a prepared class II mesio-occlusal inlay cavity, (B) master die, (C) image of the master die saved as standard tessellation language file of master die, and (D) inlay specimen fabricated by 3D printing. 3D, three-dimensional.
Zoom Image
Fig. 2 (A) Schematic representation of measurement points for the marginal and internal discrepancy in the sagittal section of silicone replica: marginal areas = A, F; internal areas = B, C, D, E. (B) Measurement of the internal gap of Zolid Fx multilayer by digital microscope at 160× magnification (light green: light body silicone, green: heavy body silicone).
Zoom Image
Zoom Image
Fig. 3 Marginal and internal discrepancies at each measurement point according to materials. The same letters indicate no statistically significant difference by one-way analysis of variance (ANOVA) at α = 0.05. *Statistically significant difference by t-test at α = 0.05. 3D, three-dimensional printing; IG, internal gap; LU, Lava Ultimate; MG, marginal gap; TS, Tescera; ZR, zirconia.
Zoom Image
Fig. 4 The three-dimensional color deviation maps for trueness and precision of four groups. Positive deviation is displayed as yellow to red colors and negative deviation as blue to dark blue. 3D, three-dimensional printing; LU, Lava Ultimate; TS, Tescera; ZR, zirconia.