Subscribe to RSS
DOI: 10.1055/s-0044-1779342
The Role of Proximal Locking Fixation in Volar Distal Radius Fracture Fixation

Abstract
Background Volar-locked plating has become a popular treatment option for unstable distal radius fractures. While using locking screws in the distal fragment increases the stability of the fixation, the biomechanical benefits of locking fixation in the proximal fragment have not been definitively established.
Purposes This study evaluated the initial mechanical behavior of the volar plating construct with different locking screw configurations in the proximal fragment.
Materials and Methods Sixteen Sawbones radius models were used. An unstable metaphyseal distal radius fracture was created and fixated with a volar-locked plate. Four different screw configurations in the proximal fragment were tested: all nonlocking screws, locking screw in the distal-most hole, locking screw in the proximal-most hole, and locking screws in both the proximal- and distal-most holes. Initial stiffness, displacement during harmonic loading, and load-to-failure were compared among the three groups.
Results The initial stiffness, displacement during harmonic loading, and load-to-failure did not significantly differ among the four proximal screw configurations (p < 0.05). Failure occurred via toggling of the screws in the configuration with all nonlocking screws and through screw breakage or locking mechanism failure in the configurations with locking screws.
Conclusion The use of locking screws in the proximal fragment did not significantly affect the initial stability of volar distal radius plating. However, failure modes differed between the nonlocking and locking configurations, consistent with known mechanical properties of locking fixation. Further mechanical studies in cadaveric models and clinical trials are warranted to determine the optimal screw configuration in volar distal radius plating.
Level of Evidence To be determined. Biomechanical study on synthetic models.
Note
This was a biomechanical study which used a synthetic model of bone and did not involve testing on human, animals, or their parts.
Publication History
Received: 07 November 2023
Accepted: 09 January 2024
Article published online:
31 January 2024
© 2024. Thieme. All rights reserved.
Thieme Medical Publishers, Inc.
333 Seventh Avenue, 18th Floor, New York, NY 10001, USA
-
References
- 1 Chung KC, Watt AJ, Kotsis SV, Margaliot Z, Haase SC, Kim HM. Treatment of unstable distal radial fractures with the volar locking plating system. J Bone Joint Surg Am 2006; 88 (12) 2687-2694
- 2 Smith DW, Henry MH. Volar fixed-angle plating of the distal radius. J Am Acad Orthop Surg 2005; 13 (01) 28-36
- 3 Arora R, Lutz M, Fritz D, Zimmermann R, Oberladstätter J, Gabl M. Palmar locking plate for treatment of unstable dorsal dislocated distal radius fractures. Arch Orthop Trauma Surg 2005; 125 (06) 399-404
- 4 Musgrave DS, Idler RS. Volar fixation of dorsally displaced distal radius fractures using the 2.4-mm locking compression plates. J Hand Surg Am 2005; 30 (04) 743-749
- 5 Orbay JL, Fernandez DL. Volar fixed-angle plate fixation for unstable distal radius fractures in the elderly patient. J Hand Surg Am 2004; 29 (01) 96-102
- 6 Ruckenstuhl P, Bernhardt GA, Sadoghi P. et al. Quality of life after volar locked plating: a 10-year follow-up study of patients with intra-articular distal radius fractures. BMC Musculoskelet Disord 2014; 15: 250
- 7 Orbay JL, Touhami A. Current concepts in volar fixed-angle fixation of unstable distal radius fractures. Clin Orthop Relat Res 2006; 445 (445) 58-67
- 8 Rozental TD, Blazar PE. Functional outcome and complications after volar plating for dorsally displaced, unstable fractures of the distal radius. J Hand Surg Am 2006; 31 (03) 359-365
- 9 Shyamalan G, Theokli C, Pearse Y, Tennent D. Volar locking plates versus Kirschner wires for distal radial fractures—a cost analysis study. Injury 2009; 40 (12) 1279-1281
- 10 Doxey SA, Huyke-Hernández FA, Robb JL, Bohn DC, Cunningham BP. Implant cost variation in surgically treated distal radius fractures. J Orthop 2023; 39: 45-49
- 11 Weninger P, Schueller M, Drobetz H, Jamek M, Redl H, Tschegg E. Influence of an additional locking screw on fracture reduction after volar fixed-angle plating-introduction of the “protection screw” in an extra-articular distal radius fracture model. J Trauma 2009; 67 (04) 746-751
- 12 Mehling I, Müller LP, Delinsky K, Mehler D, Burkhart KJ, Rommens PM. Number and locations of screw fixation for volar fixed-angle plating of distal radius fractures: biomechanical study. J Hand Surg Am 2010; 35 (06) 885-891
- 13 Willis AA, Kutsumi K, Zobitz ME, Cooney III WP. Internal fixation of dorsally displaced fractures of the distal part of the radius. A biomechanical analysis of volar plate fracture stability. J Bone Joint Surg Am 2006; 88 (11) 2411-2417
- 14 Bockmann B, Budak C, Figiel J. et al. Is there a benefit of proximal locking screws in osteoporotic distal radius fractures? - A biomechanical study. Injury 2016; 47 (08) 1631-1635
- 15 O'Toole RV, Andersen RC, Vesnovsky O. et al. Are locking screws advantageous with plate fixation of humeral shaft fractures? A biomechanical analysis of synthetic and cadaveric bone. J Orthop Trauma 2008; 22 (10) 709-715
- 16 Egol KA, Kubiak EN, Fulkerson E, Kummer FJ, Koval KJ. Biomechanics of locked plates and screws. J Orthop Trauma 2004; 18 (08) 488-493
- 17 Grawe B, Le T, Williamson S, Archdeacon A, Zardiackas L. Fracture fixation with two locking screws versus three non-locking screws: a biomechanical comparison in a normal and an osteoporotic bone model. Bone Joint Res 2012; 1 (06) 118-124