Endoscope Disinfection Using Acidic Electrolytic Water
S. Tsuji 1
, S. Kawano 1
, M. Oshita 1
, A. Ohmae 1
, Y. Shinomura 1
, Y. Miyazaki 1
, S. Hiraoka 1
, Y. Matsuzawa 1
, T. Kamada 1
, M. Hori 1
, T. Maeda 2
1 First and Second Depts. of Medicine, Osaka University School of Medicine, Suita, Japan
Background and Study Aims: The aim of the present study was to evaluate a new endoscope disinfector (WM-1) that uses acidic electrolytic water (AEW).
Materials and Methods: AEW was produced by electrolysis of a 0.05 % NaCl-water mixture, with a redox potential greater than 1000 mV and a pH lower than 2.7. In the first study, an endoscope artificially contaminated with 15 species of bacteria and four strains of viruses was treated using the WM-1. In the second study, endoscopic contamination after clinical use was examined by culture for Helicobacter pylori and other bacteria, and by polymerase chain reaction for the H. pylori urease gene and hepatitis C virus. The extent of contamination was then examined after exposing the WM-1 to AEW. The safety of AEW was examined using both in vivo and in vitro studies.
Results: All of the bacteria and viruses were destroyed or inactivated after the instrument had been exposed to AEW. Clinical contamination was detected from the instrument in 19 of 30 endoscopic procedures, whereas no bacteria or viruses were detected after five minutes' exposure to AEW. AEW was found to be nonirritant, nontoxic to cells, and nonmutagenic.
Conclusion: The WM-1 successfully and safely disinfected the endoscopes. With running costs of ¥24 per day ($0.21 per day), the WM-1 provides an effective and inexpensive alternative to conventional disinfection equipment.
References
1
Spach DH, Silverstein FE, Stamm WE.
Transmission of infection by gastrointestinal endoscopy and bronchoscopy.
Ann Intern Med.
1993;
118
117-128
5
Katner HP, Buckley RL, Smith MU, et al.
Endoscopic cleaning and disinfection procedure for preventing iatrogenic spread of human immunodeficiency virus.
J Fam Pract.
1988;
27
271-276
7
Carr-Locke DL, Clayton P.
Disinfection of upper gastrointestinal fibreoptic endoscopy equipment: an evaluation of a cetrimide chlorhexidine solution and glutaraldehyde.
Gut.
1978;
19
916-922
9
Gerding DN, Peterson LR, Vennes JA.
Cleaning and disinfection of fiberoptic endoscopes: evaluation of glutaraldehyde exposure time and forced-air drying.
Gastroenterology.
1982;
83
613-618
11
Yoshii Y, Kasugai T, Nakamura T.
Disinfectious effect of endoscopes using new washer model EW-10 [in Japanese].
Gastroenterol Endosc.
1985;
27
354-361
13
American Society for Gastrointestinal Endoscopy.
Infection control during gastrointestinal endoscopy: guidelines for clinical application.
Gastrointest Endosc.
1988;
34 (Suppl 3)
37-40S
14
Van Gossum A, Loriers M, Serruys E, et al.
Methods of disinfecting endoscopic material: result of an international survey.
Endoscopy.
1989;
21
247-250
15
British Society of Gastroenterology Working Party.
Cleaning and disinfection of equipment for gastrointestinal flexible endoscopy.
Gut.
1988;
29
1134-1151
17
Fantry GT, Zheng QX, James SP.
Conventional cleaning and disinfection techniques eliminate the risk of endoscopic transmission of Helicobacter pylori.
.
Am J Gastroenterol.
1995;
90
227-232
18
Graham DY, Alpert LC, Smith JL, et al.
Iatrogenic Campylobacter pylori infection is a cause of epidemic achlorhydria.
Am J Gastroenterol.
1988;
83
974-980
20
Katoh M, Saito D, Noda T, et al.
Helicobacter pylori may be transmitted through gastrofiberscope even after manual hyamine washing.
Jpn J Cancer Res.
1993;
84
117-119
27
Okamoto H, Sugiyama Y, Okada S, et al.
Typing hepatitis C virus by polymerase chain reaction with type-specific primers: application to clinical surveys and tracing infectious sources.
J Gen Virol.
1992;
73
673-679
28
Draize JH, Woodard G, Calvery HO.
Methods for the study of irritation and toxicity of substances applied topically to the skin and mucous membranes.
J Pharmacol Exp Ther.
1944;
82
377-390
30
Chiaramonte M, Farini R, Truscia D, et al.
Risk of hepatitis B virus infection following upper gastrointestinal endoscopy: a prospective study in an endemic area.
Hepatogastroenterology.
1983;
30
189-191
31
Hoofnagle JH, Blake J, Buskell-Bales Z, et al.
Lack of transmission of type B hepatitis by fiberoptic upper endoscopy.
J Clin Gastroenterol.
1980;
2
65-69