Fifty-eight non-pathological fractures of the proximal femur treated by the unreamed femoral nail with spiral blade (UFN-SB) were reviewed prospectively with focus on the postoperative mechanical complications.
Twelve patients had Sensheimer type II A fractures, five Sensheimer type II B, twenty-nine Sensheimer type II C, six Sensheimer type III A, two Sensheimer type IV, and five Sensheimer type V.
In thirteen fractures, the UFN-SB failed before bony union. Eleven mechanical failures occurred in type II C, and two were type V fractures. Implant failure was defined as loosening, breakage or bending of the spiral blade with clear influence on the fracture alignment.
Revision surgery was necessary in five patients: two dynamic hip screws, two 95°-blade plates and one hemiarthroplasty.
The data suggest that the unreamed femoral nail with the spiral blade should not be the implant of choice for proximal femoral fractures with intertrochanteric extensions of the fracture line.
Zusammenfassung
58 nicht pathologische proximale Femurfrakturen wurden durch einen von proximal eingeschlagenen unaufgebohrten Femurmarknagel mit Spiralblatt (UFN-SB) versorgt und prospektiv unter besonderer Berücksichtigung mechanischer Komplikationen beobachtet. Bei 12 Patienten lagen Sensheimer Typ II A-Frakturen vor. 5-mal bestanden Sensheimer Typ II B-Frakturen, 29-mal Sensheimer Typ II C. 6-mal Sensheimer Typ III A, 2-mal Sensheimer Typ IV und 5-mal Sensheimer Typ V.
In 13 Fällen versagte der UFN-SB-Nagel vor Abschluss der Frakturheilung. Bei der II C-Frakturen traten 11-mal mechanische Fehler auf, beim Typ V 2-mal. Das Implantatversagen bestand in Lockerung, Bruch oder Verbiegung des Spiralblattes, jeweils mit entsprechend negativem Einfluss auf die Frakturstellung. 5 Patienten mussten revidiert werden. 2-mal war eine dynamische Hüftschraube notwendig, 2-mal eine 95° Winkelplatte sowie 1-mal eine Hemirarthroplastie (Frakturendoprothese?).
Diese Ergebnisse legen es nahe, die Anzeigestellung von unaufgebohrten Femurmarknägeln mit Spiralblatt bei proximalen Femurfraktruen insbesondere mit intertrochantärem Frakturverlauf kritisch zu überdenken.
Key words
Traumatology - Implant failure - Hip fractures - Subtrochanteric fractures - Femur fractures
Schlüsselwörter
Traumatologie - Implantatversagen - Proximale Femurfraktur - Subtrochantäre Femurfraktur
References
- 1 Alho A, Stromsoe K, Ekeland A. Locked intramedullary nailing of femoral shaft fractures. J Trauma. 1991; 31 49-59
- 2 Alho A, Ekeland A, Stromsoe K. Subtrochanteric femoral shaft fracturestreated with locked intramedullary nails. Acta Orthop Scand. 1991; 62 573-576
-
3 Bonnair F. Loadability of unstable inter-subtrochanteric fractures. Personal communication, Hip fractures 2000 state of the art. International Congres May 26-27 2000, Amsterdam
- 4 Brumback R J, Uwagie-Ero S, Lakatos R P. et al . Intramedullary nailing of femoral shaft fractures. J Bone Joint Surg [Am]. 1988; 70 1453-1462
- 5 Kempf I, Grosse A, Beck G. Closed locked intramedullary nailing: its application to comminuted fractures of the femur. J Bone Joint Surg [Am]. 1985; 67 709-713
- 6 Kinast C, Bolhofner B R, Mast J W, Ganz R. Subtrochanteric fractures of the femur: results of treatment with the 95° condylar blade plate. Clin Orthop. 1989; 238 122-125
- 7 Krettek C, Rudolf J, Schandelmaier P, Guy P, Könemann B, Tscherne H. Unreamed intramedullary nailing of femoral shaft fractures: operative technique and early clinical experience with the standard locking option. Injury. 1996; 27 233-254
- 8 Kummer F J, Olsson O. et al . Intramedullary versus extramedullary fixation of subtrochanteric fractures. A biomechanical study. Acta Orthop Scand. 1998; 69 580-584
- 9 Lunsjo K, Ceder L, Stigsson L, Hauggaard A. Two-way compression along the shaft and the neck of the femur with the Medoff sliding plate. One year follow-up of 108 intertrochanteric fractures. J Bone Joint Surg [Br]. 1996; 78 387-390
-
10 Russell T A. Campbell's operative orthopaedics. Crenshaw AH (ed). Eight Edition. Mosby, St. Louis 1992; 895-971
- 11 Wheeler D L, Croy T J, Woll S T. et al . Comparison of reconstruction nails for high subtrochanteric femur fracture fixation. Clin Orthop. 1997; 338 231-239
- 12 Whitelaw G P, Segal D, Sanzone C F. et al . Unstable intertrochanteric/subtrochanteric fractures of the femur. Clin Orthop. 1990; 252 238-245