The Influence of Biomechanically Inadequate Internal Fixation on Dynamic Hip Screws
A Metallurgical Analysis in Three CasesPublication History
Publication Date:
17 May 2002 (online)
Abstract
The breakage of two plates and one lag screw of Dynamic Hip Screws (AOR) was analyzed. Radiological examination showed technically inadequate internal fixation in one case and potentially unstable internal fixation in the other despite adequate reduction. In the case with breakage of a lag screw fracture reduction was unsuccessful and the fragments were stabilized in malalignment. Metallurgical analysis showed that fatigue breakages caused by corrosion had occurred in all cases and that high local tensile stresses generated by poor internal fixation had led to implant failure. Plate breakages were found to be due to mechanical damage in the screw holes. From these crevice corrosion started. In all three cases metallurgical analysis ruled out material faults, which had originally been thought to be the underlying factor. Only metallurgical analysis can provide conclusive evidence of the actual cause of failure.
Zusammenfassung
Analysiert werden drei Fälle von Implantatbrüchen nach Stabilisierung hüftgelenksnaher Frakturen mittels dynamischer Hüftschrauben (AOR), davon zwei Platten-, ein Schenkelhalsschrauben- und ein Bruch einer Corticalisschraube. Röntgenologisch fand sich bei den Plattenbrüchen einmal eine technisch insuffiziente und einmal eine trotz anatomischer Reposition potentiell instabile Osteosynthese. Im Falle des Schenkelhalsschraubenbruches lag eine operativ fixierte Fehlrotation vor. Die metallurgische Analyse deckte in allen Fällen das Vorliegen von Korrosionsspaltbrüchen auf und die hohen lokalen Zugbeanspruchungen aufgrund biomechanisch inadäquater Osteosynthesen führten letztendlich zum Implantatversagen. Bei den Plattenbrüchen konnte nachgewiesen werden, dass geringfügige mechanische Beschädigungen in den Schraubenaufnahmelöchern als Ausgangsstellen für die Spaltkorrosion anzusprechen waren. Durch die metallurgische Untersuchung konnte bewiesen werden, dass in keinem der Fälle ein zunächst in Erwägung gezogener Materialfehler vorlag und das Implantatversagen ausschließlich durch insuffiziente Osteosynthesen bedingt wurde. Es ist daher nur eine metallurgische Analyse in der Lage über die tatsächliche Ursache eines Implantatbruches Aufschluss zu geben.
Key words
Implant breakage - Dynamic hip screws - Metallurgical analysis
Schlüsselwörter
Implantatbrüche - Dynamische Hüftschrauben - Metallurgische Analyse
References
- 1 Bannister G C, Orth M Ch, Gibson A GF, Ackroyd C E, Newman J H. The fixation and prognosis of trochanteric fractures. A randomised prospective controlled trial. Clin Orthop. 1990; 254 242-246
- 2 Bonnaire F, Götschin U, Kuner E H. Früh- und Spätergebnisse nach 200 DHS-Osteosynthesen zur Versorgung pertrochanterer Femurfrakturen. Unfallchirurg. 1992; 95 246-253
- 3 Bridle S H, Patel A D, Bircher M, Calvert P T. Fixation of intertrochanteric fractures of the femur. A randomised prospective comparison of the Gamma Nail and the Dynamic Hip Screw. J Bone Joint Surg [Br]. 1991; 73 330-334
- 4 Esser M P, Kassab J Y, Jones D HA. Trochanteric fractures of the femur. A randomised prospective trial comparing the Jewett Nail Plate with the Dynamic Hip Screw. J Bone Joint Surg [Br]. 1986; 68 557-560
-
5 Frank E, Zitter E. Metallische Implantate in der Knochenchirurgie. Springer, Wien, New York 1971
- 6 Friedl W, Göhring U, Fritz T h, Krieglstein C h. Die Gleitnagelosteosynthese. Ein neues universell einsetzbares Implantat zur Versorgung per- und subtrochanterer Femurfrakturen. Chirurg. 1998; 69 191-197
- 7 Höntzsch D, Weller S, Karnatz N. Die dynamische Hüftschraube (DHS) im Vergleich zur Ender-Nagelung. Akt Traumatol. 1990; 20 14-19
- 8 Kroczek H, Karim R, Pfister U. Drei Jahre Erfahrung mit der dynamischen Hüftkompressionsschraube (DHS). Akt Traumatol. 1988; 18 187-191
- 9 Kukla C, Heinz T, Berger G, Kwasny O, Rosenberger A, Vecsei V. Gamma Nail vs. Dynamic Hip Screw in 120 patients over 60 years. A randomized trial. Acta Chir Austriaca. 1997; 5 290-293
- 10 Kyle R F, Wright T M, Burstein A H. Biomechanical analysis of the sliding characteristics of compression hip screws. J Bone Joint Surg [Am]. 1980; 62 1308-1314
- 11 Leung K S, So W S, Shen W Y, Hui P W. Gamma nails and dynamic hip screws for peritrochanteric fractures. A randomised prospective study in elderly patients. J Bone Joint Surg [Br]. 1992; 74 345-351
- 12 Lustenberger A, Bekic J, Ganz R. Rotationsstabilität trochantärer Femurfrakturen fixiert mit der DHS. Eine radiologische Analyse. Unfallchirurg. 1995; 98 514-517
- 13 Manner M, Ruf W. Die dynamische Hüftschraube. Lösung aller Frakturprobleme des trochanteren Femurbereiches. Unfallchirurg. 1988; 91 299-306
- 14 McLaren C AN, Buckley J R, Rowley D I. Intertrochanteric fractures of the femur: A randomized prospective trial comparing the Pugh nail with the dynamic hip screw. Injury. 1991; 22 193-196
- 15 Mommsen U, Schmidt K D, Schumpelik V, Jungbluth K H. Erfahrungen mit der Endernagelung und der dynamischen Kompressionsschraube der AO bei der Versorgung pertrochanterer Frakturen. Hefte Unfallheilkunde. 1983; 165 266-268
- 16 Muller B, Bonnaire F, Heckl T, Jaeger J H, Kempf I, Kuner E H. Ender-Nägel mit Verriegelung oder dynamische Hüftschraube bei pertrochanteren Frakturen? Eine prospektive, grenzüberschreitende Studie. Unfallchirurgie. 1994; 20 18-29
- 17 Müller-Färber J, Wittner B, Reichel R. Spätergebnisse nach Versorgung pertrochanterer Femurfrakturen des alten Menschen mit der DHS. Unfallchirurg. 1988; 91 341-350
- 18 Ortner F, Wagner M, Trojan E. Operative Versorgung der pertrochanteren Frakturen mit der dynamischen Hüftschraube (DHS) der AO. Unfallchirurg. 1989; 92 274-281
- 19 Penschuk C E. Möglichkeiten der operativen Versorgung von per- und subtrochantären Femurfrakturen unter besonderer Berücksichtigung der DHS-Schraube. Akt Chir. 1986; 21 72-75
- 20 Radford P J, Needoff M, Webb J K. A prospective randomised comparison of the Dynamic Hip Screw and the Gamma Locking Nail. J Bone Joint Surg [Br]. 1993; 75 789-793
-
21 Regazzoni P, Rüedi T, Allgöwer M. The dynamic hip screw implant system. Springer, Berlin, Heidelberg, New York 1985
- 22 Siebler G, Bonnaire F, Kuner E H. Intraoperative und frühe postoperative Komplikationen bei der Osteosynthese pertrochanterer Femurfrakturen mit der DHS. Unfallchirurg. 1987; 90 407-411
- 23 Soerensen J L, Varmerken J E, Boemler J. Internal fixation of femoral neck fractures. Dynamic Hip and Gouffon screws compared in 73 patients. Acta Orthop Scand. 1992; 63 288-292
- 24 Sperner G, Wanitschek P, Benedetto K P, Glötzer W. Verfahrensfehler und Frühkomplikationen bei der Osteosynthese pertrochanterer Oberschenkelfrakturen mit der dynamischen Hüftschraube. Unfallchirurg. 1989; 92 571-576
- 25 Thomas A P. Dynamic hip screws that fail. Injury. 1991; 22 45-46
- 26 Wu C-C, Shih C-H. Biomechanical analysis of the dynamic hip screw in the treatment of intertrochanteric fractures. Arch Orthop Trauma Surg. 1991; 110 307-310
Ernst SimM.D.
Unfallkrankenhaus Wien-Meidling
Kundratstraße 37
1120 Vienna
Austria
Phone: +43-1-60150
Fax: +43-1-60150-3075