Comparison of Flow Parameters to Analyse Bolus Kinetics of Ultrasound Contrast Enhancement in a Capillary Flow Model

Parametervergleich zur Beschreibung der Flussrate im Kapillarphantom nach Ultraschallkontrastmittel-BolusinjektionH. Cangür1 , K. Meyer-Wiethe1 , G. Seidel1
  • 1Department of Neurology, University Hospital Schleswig-Holstein, Campus Lübeck, Germany
This study was sponsored by the Ultrasonographic Monitoring and Early Diagnosis of Stroke (UMEDS) project, which is funded by the European Union (QLG1-CT-2002 - 01 518).
Weitere Informationen

Publikationsverlauf

received: 6.8.2004

accepted: 30.10.2004

Publikationsdatum:
14. Dezember 2004 (online)

Zusammenfassung

Studienziel: Die Darstellung von Ultraschall-Kontrastmitteln im Hirngewebe mittels Harmonic Imaging ist ein neues Verfahren, um Hirnperfusion darzustellen. Das Ziel unserer Untersuchungen war, den Einfluss der Flussrate auf Parameter zur Beschreibung der Boluskinetik in einem Phantommodell zu untersuchen. Methode: Der In-vitro-Versuchsaufbau bestand aus einem offenen Kartuschenflussphantom in das SonoVue® per Bolus injiziert wurde. Die Boluskinetiken wurden mittels Harmonic Imaging (SONOS 5500, 1,8/3,6 MHz, MI:1,6) und einer festen Pulsrepetitionsfrequenz (0,67 Hz) analysiert. Dabei wurden fünf unterschiedliche Flussgeschwindigkeiten (1,8 - 21,6 ml/min) untersucht. Es wurden Zeit-Intensitäts-Kurven in einer definierten Region-of-Interest (ROI) aus den Grauwertbild-Sequenzen hergestellt. Aus den Zeit-Intensitäts-Kurven wurden dann verschiedene Parameter (time-to-peak-intensity [TTP], peak-intensity [PI], area under the curve [AUC], positive gradient [PG] and full width at half maximum [FWHM]) errechnet und mit der Flussrate korreliert. Ergebnisse: Es zeigte sich eine hochsignifikante lineare Korrelation zwischen TTP und PG mit der Flussrate (r = - 0.96, p < 0,01 für TTP, bzw. r = 0,91, p = 0,03 für PG). Sowohl die AUC (r = - 0.67, p = 0,22), als auch die PI (r = 0,68, p = 0,20) und die FWHM (r = - 0.63, p = 0,37) korrelierten nicht signifikant mit der Flussrate. Schlussfolgerung: Unser Flussmodell zeigt eine hervorragende Korrelation zwischen TTP und PG mit der Flussrate, so dass diese beiden Parameter sehr gut geeignet sind, Veränderungen der Flussrate zu erfassen.

Abstract

Aim: Perfusion harmonic imaging is a new method to evaluate brain perfusion. The purpose of our study was to describe flow by analysing the kinetics of ultrasound contrast enhancement in a flow model after bolus injection of an ultrasound contrast agent (UCA). Methods: We performed an in-vitro study with an open-circuit flow model and SonoVue® given as bolus injection. Bolus kinetics was analysed by harmonic imaging (SONOS 5500, 1.8/3.6 MHz, MI: 1.6) at a fixed frame rate (0.67 Hz). Five different flow rates from 1.8 to 21.6 ml/min were tested. We generated time-intensity curves (TIC) from the grey scale loops. Five parameters (time-to-peak-intensity [TTP], peak-intensity [PI], area under the curve [AUC], positive gradient [PG] and full width at half maximum [FWHM]) were correlated with flow rate. Results: The analysis of both TTP and PG versus flow rate indicated a highly significant linear correlation (r = - 0.96, P <.01; r = 0.91, P = .03; respectively). Neither the AUC (r = - 0.67, P = .22), nor the PI (r = 0.68, P = .20) nor the FWHM (r = - 0.63, P = .37) correlated significantly with flow rate. Conclusion: The relationship between TIC parameters and flow rates indicates that, in our flow model, the TTP and PG are the most robust and reliable parameters compared to the AUC, PI and FWHM.

References

  • 1 Bos L J, Piek J J, Spaan J AE. Effects of Shadowing on the time-intensity curves in contrast echocardiography: a phantom study.  Ultrasound Med Biol. 1996;  22 217-227
  • 2 Claassen L, Seidel G, Algermissen C. Quantification of flow rates using harmonic grey-scale imaging and an ultrasound contrast agent: an in vitro and in vivo study.  Ultrasound Med Biol. 2001;  27 83-88
  • 3 Cosgrove D, Eckersley R, Blomley M. et al . Quantification of blood flow.  Eur Radiol. 2001;  11 1338-1344
  • 4 Devvoye L. Non-Uniform Random Variate Generation. New York; Springer Verlag 1986
  • 5 Federlein J, Postert T, Meves S. et al . Ultrasonic evaluation of pathological brain perfusion in acute stroke using second harmonic imaging.  J Neural Neurosurg Psychiatry. 2000;  69 616-622
  • 6 Harrer J U, Klötzsch C. Second harmonic imaging of the human brain: the practicability of coronal insonation Planes and alternative perfusion parameters.  Stroke. 2002;  33 1530-1535
  • 7 Kaul S, Kelly P, Oliner J D. et al . Assessment of regional myocardial blood flow with myocardial contrast two-dimensional echocardiography.  J Am Coll Cardiol. 1989;  13 468-482
  • 8 Lucidarme O, Kono Y, Corbeil J. et al . Validation of ultrasound contrast destruction imaging for flow quantification.  Ultrasound Med Biol. 2003;  29 1697-1704
  • 9 Pinard E, Nallet H, MacKenzie E T. et al . Penumbral microcirculatory changes associated with peri-infarct depolarizations in the rat.  Stroke. 2002;  33 606-612
  • 10 Postert T, Hoppe P, Federlein J. et al . Contrast agent specific imaging modes for the ultrasonic assessment of parenchymal cerebral echo contrast enhancement.  J Cereb Blood Flow Metab. 2000;  20 1709-1716
  • 11 Seidel G, Vidal-Langwasser M, Algermissen C. et al . The influence of doppler system settings on the clearance kinetics of different ultrasound contrast agents.  European J Ultrasound. 1999;  9 167-175
  • 12 Seidel G, Algermissen C, Christoph A. et al . Harmonic imaging of the human brain: visualization of brain perfusion with ultrasound.  Stroke. 2000;  31 151-154
  • 13 Seidel G, Albers T, Meyer K. et al . Perfusion harmonic imaging in acute middle cerebral artery infarction.  Ultrasound Med Biol. 2003;  29 1245-1251
  • 14 Thompson H K, Starmer F, Whalen R E. et al . Indicator transit time considered as a gamma variate.  Circ Res. 1964;  14 502-515
  • 15 Ugolini P, Delouche A, Herment A. et al . In vitro flow quantification with contrast power doppler imaging.  Ultrasound Med Biol. 2002;  26 113-120
  • 16 Veltmann C, Lohmaier S, Schlosser T. et al . On the design of a capillary flow phantom for the evaluation of ultrasound contrast agents at very low flow velocities.  Ultrasound Med Biol. 2002;  28 625-634
  • 17 Villringer A, Them A, Lindauer U. et al . Capillary perfusion of the rat brain cortex. An in vivo confocal microscopy study.  Circ Res. 1994;  75 55-62
  • 18 Wiesmann M, Seidel G. Ultrasound perfusion imaging of the human brain.  Stroke. 2000;  31 2421-2425

Priv.-Doz. Dr. med. Günter Seidel

University Hospital Schleswig Holstein, Campus Lübeck, Department of Neurology

Ratzeburger Allee 160

23538 Lübeck, Germany

Telefon: ++ 49/4 51/5 00 33 34

Fax: ++ 49/4 51/5 00 24 89

eMail: guenter.seidel@neuro.uni-luebeck.de

    >