J Knee Surg
DOI: 10.1055/a-2514-9655
Special Focus Section

Patellar Overstuffing in Total Knee Arthroplasty Results in Decreased Extensor Mechanism Efficiency

Brandon R. Bukowski
1   Department of Orthopedic Surgery, Hospital for Special Surgery, New York, New York
,
Sonia K. Chandi
1   Department of Orthopedic Surgery, Hospital for Special Surgery, New York, New York
,
Pravjit Bhatti
1   Department of Orthopedic Surgery, Hospital for Special Surgery, New York, New York
,
Paola Jimenez
2   Smith and Nephew Inc., Advanced Surgical Devices, Memphis, Tennessee
,
Nathan Lenz
2   Smith and Nephew Inc., Advanced Surgical Devices, Memphis, Tennessee
,
Fred D. Cushner
1   Department of Orthopedic Surgery, Hospital for Special Surgery, New York, New York
,
Steven B. Haas
1   Department of Orthopedic Surgery, Hospital for Special Surgery, New York, New York
,
Brian P. Chalmers
1   Department of Orthopedic Surgery, Hospital for Special Surgery, New York, New York
› Author Affiliations

Abstract

Patellofemoral offset is influenced by patellar thickness in total knee arthroplasty (TKA), with effects on extensor mechanism efficiency (EME), patellar contact moment arm (PCMA), and patellar contact force (PCF). This study utilized a computational model to better quantify these measures during knee range of motion (ROM).

A computational model of a cruciate-sacrificing, posterior-stabilized TKA design was used to quantify moments at knee flexion angles from 0° to 135° when patella thickness was increased (“overstuffed”) or decreased (“understuffed”) by 2 mm and 5 mm from normal. Outcomes included changes in EME, PCMA, PCF, tibial tubercle moment arm (TTMA), and the quadriceps moment arm (QMA).

Overstuffing of the patellofemoral articulation by 2 mm and 5 mm decreased EME up to 12 and 28%, respectively, with an 8% and 16% increase in EME with understuffing that peaked at 25° knee flexion. The QMA increased up to 5 and 12% with overstuffing by 2 mm and 5 mm and decreased up to 5% and 14% with understuffing, respectively. There was an increase in the PCMA/PCF, which opposed the QMA, up to 8%/17% and 16%/39% with overstuffing by 2 mm and 5 mm, respectively. The PCMA/PCF decreased up to 3%/12% and 8%/33% with understuffing by 2 mm and 5 mm, respectively.

In this computational model, increased patellofemoral offset resulting in overstuffing of the patella resulted in decreased EME and increased PCMA and PCF during knee ROM. This may contribute to abnormal patellofemoral mechanics and a potential etiology for anterior knee pain and quadriceps weakness after TKA.

Note

This investigation was performed at the Hospital for Special Surgery, New York, NY.




Publication History

Received: 26 February 2024

Accepted: 10 January 2025

Article published online:
11 February 2025

© 2025. Thieme. All rights reserved.

Thieme Medical Publishers, Inc.
333 Seventh Avenue, 18th Floor, New York, NY 10001, USA

 
  • References

  • 1 Hamilton WG, Ammeen DJ, Parks NL, Goyal N, Engh GA, Engh Jr CA. Patellar cut and composite thickness: the influence on postoperative motion and complications in total knee arthroplasty. J Arthroplasty 2017; 32 (06) 1803-1807
  • 2 Hsu HC, Luo ZP, Rand JA, An KN. Influence of patellar thickness on patellar tracking and patellofemoral contact characteristics after total knee arthroplasty. J Arthroplasty 1996; 11 (01) 69-80
  • 3 Kim AD, Shah VM, Scott RD. The effect of patellar thickness on intraoperative knee flexion and patellar tracking in patients with arthrofibrosis undergoing total knee arthroplasty. J Arthroplasty 2016; 31 (05) 1011-1015
  • 4 Mihalko W, Fishkin Z, Krackow K. Patellofemoral overstuff and its relationship to flexion after total knee arthroplasty. Clin Orthop Relat Res 2006; 449: 283-287
  • 5 Oishi CS, Kaufman KR, Irby SE, Colwell Jr CW. Effects of patellar thickness on compression and shear forces in total knee arthroplasty. Clin Orthop Relat Res 1996; (331) 283-290
  • 6 Star MJ, Kaufman KR, Irby SE, Colwell Jr CW. The effects of patellar thickness on patellofemoral forces after resurfacing. Clin Orthop Relat Res 1996; (322) 279-284
  • 7 Tanikawa H, Tada M, Harato K, Okuma K, Nagura T. Influence of total knee arthroplasty on patellar kinematics and patellofemoral pressure. J Arthroplasty 2017; 32 (01) 280-285
  • 8 Bengs BC, Scott RD. The effect of patellar thickness on intraoperative knee flexion and patellar tracking in total knee arthroplasty. J Arthroplasty 2006; 21 (05) 650-655
  • 9 Matz J, Howard JL, Morden DJ, MacDonald SJ, Teeter MG, Lanting BA. Do changes in patellofemoral joint offset lead to adverse outcomes in total knee arthroplasty with patellar resurfacing? A radiographic review. J Arthroplasty 2017; 32 (03) 783-787.e1
  • 10 Lenz N. Comparing ligament strain in total knee arthroplasty designs using a computational model. Paper presented at: 62nd Annual Meeting of the Orthopaedic Research Society. 2016.
  • 11 Bracey DN, Brown ML, Beard HR. et al. Effects of patellofemoral overstuffing on knee flexion and patellar kinematics following total knee arthroplasty: a cadaveric study. Int Orthop 2015; 39 (09) 1715-1722
  • 12 Abolghasemian M, Samiezadeh S, Sternheim A, Bougherara H, Barnes CL, Backstein DJ. Effect of patellar thickness on knee flexion in total knee arthroplasty: a biomechanical and experimental study. J Arthroplasty 2014; 29 (01) 80-84
  • 13 Browne C, Hermida JC, Bergula A, Colwell Jr CW, D'Lima DD. Patellofemoral forces after total knee arthroplasty: effect of extensor moment arm. Knee 2005; 12 (02) 81-88
  • 14 D'Lima DD, Poole C, Chadha H, Hermida JC, Mahar A, Colwell Jr CW. Quadriceps moment arm and quadriceps forces after total knee arthroplasty. Clin Orthop Relat Res 2001; (392) 213-220
  • 15 Reuben JD, McDonald CL, Woodard PL, Hennington LJ. Effect of patella thickness on patella strain following total knee arthroplasty. J Arthroplasty 1991; 6 (03) 251-258
  • 16 Huberti HH, Hayes WC. Patellofemoral contact pressures. The influence of q-angle and tendofemoral contact. J Bone Joint Surg Am 1984; 66 (05) 715-724
  • 17 Rosenberg AG, Andriacchi TP, Barden R, Galante JO. Patellar component failure in cementless total knee arthroplasty. Clin Orthop Relat Res 1988; (236) 106-114
  • 18 Wilson JM, Sullivan MH, Pagnano MW, Trousdale RT. AAHKS Surgical Techniques & Technologies Award: resurfacing the thin native patella: is it safe?. J Arthroplasty 2023; 38 (7 Suppl 2): S9-S14