Nuklearmedizin 1989; 28(05): 201-207
DOI: 10.1055/s-0038-1629492
Orticle Articles
Schattauer GmbH

Fission 99Mo/99mTc Generators – A Study of their Quality[*]

E. Reich
1   From the Institute for Radiation Hygiene of the Federal Flealth Office, Neuherberg, FRG
,
K. W. Bögi
1   From the Institute for Radiation Hygiene of the Federal Flealth Office, Neuherberg, FRG
› Institutsangaben
Weitere Informationen

Publikationsverlauf

Received: 13. April 1989

26. April 1989

Publikationsdatum:
05. Februar 2018 (online)

Quality was controlled of seven different 99Mo/99mTc generators sold in the Federal Republic of Germany, which were all loaded with fission molybdenum. Aim of the work was to test the 99mTc eluates for their radionuclidic and radiochemical purity. Furthermore, the 99Mo content of the aluminium oxide column was measured. Determination of the elution efficiency, measurement of the pH-value of the eluate as well as a test to detect soluble aluminium in the eluate were also performed. In addition to the chemical and physical examinations, application, quality of accessories and their practicability were tested at all generators. The methods of analysis chosen were, among others, gamma spectrometry and thin-layer chromatography; the measuring methods were a site-sensitive proportional counter (measurement of the distribution of radioactivity of the chromatograms) and atomic absorption spectrometry (AI content of the eluates). The quality control of the 99mTc eluates had satisfying results. The eluates showed – with one exception – high and sufficient radionuclidic purity and very good radiochemical purity. A considerable overload of the columns with 99Mo at the time of reference was not found. The elution yields with values between 85 and 122% were in good agreement with the requirements. All eluates had pH-values between 5.0 and 6.5, and an aluminium content below 1 μg/ml. The generators had good performance and proved generally to be a reliable source of 99mTc-pertechnetate. The application was safe and, with some exceptions, fulfilled the requirements.

Zusammenfassung

Die Qualitätskontrolle wurde an sieben verschiedenen in der Bundesrepublik Deutschland kommerziell vertriebenen 99Mo/99mTc Generatoren, die alle mit Spaltmolybdän beladen waren, durchgeführt. Ziel dieser Arbeit war es, die gewonnenen 99mTc Eluate hinsichtlich ihrer radionuklearen und radiochemischen Reinheit zu überprüfen. Weiterhin wurde die Beladung der Aluminiumoxidsäule mit 99Mo gemessen. Bestimmung der Elutionsausbeute, Messung des pH-Wertes im Eluat sowie die Prüfung auf lösliches Aluminium im Eluat wurden ebenfalls vorgenommen. Neben den radiochemischen und radionuklearen Untersuchungen wurde außerdem an allen Generatoren deren Handhabung, Qualität des Zubehörs sowie dessen Praktikabilität getestet. Als Analysenmethoden wurden unter anderen die Gammaspektrometrie und die Dünnschichtchromatographie, als Meßverfahren ein ortsempfindlicher Proportionalzähler (Messung der Radioaktivitätsverteilung auf den Chromatogrammen) sowie die Atomabsorptionsspektrometrie (Al-Gehalt der Eluate) gewählt. Es zeigte sich, daß die Qualität der 99mTc Eluate zufriedenstellend war. Die Eluate zeigten mit einer Ausnahme hohe und ausreichende radionukleare Reinheit und sehr gute radiochemische Reinheit. Eine nennenswerte Überladung der Säule mit 99Mo zum Referenzzeitpunkt konnte nicht festgestellt werden. Die Elutionsausbeuten lagen mit Werten zwischen 85 und 122% im Rahmen der gestellten Forderungen. Alle Eluate hatten pH-Werte zwischen 5,0 und 6,5 und einen Aluminiumgehalt unter 1 μg/ml. Die Generatoren erbrachten eine gute Leistung und erwiesen sich weitgehend als zuverlässige Quelle für die Erzeugung von 99mTc-Pertechnetat. Die Handhabung war sicher und erfüllte mit wenigen Ausnahmen die gestellten Ansprüche.

* This report is based on work conducted within the scope of a research project supported by the Federal Minister for the Environment, Nature Conservation and Nuclear Safety of the Federal Republic of Germany. The present report reflects the opinions of the authors and does not necessarily express the views of the Minister.


 
  • REFERENCES

  • 1 Ayres R L, Lull R, Ice R. Radioactivity standardization of99mTc and 99Mo. Int J Appl Rad Isotopes 1982; 33: 835-41.
  • 2 Billinghurst M W, Hreczuch F W. Contamination from 131I. 103Ru and 239Np in the cluate of 99Mo/99mTe generators loaded with (n-gamma)-produccd 99Mo. J Nuel Med 1974; 15: 840-3.
  • 3 British Pharmacopoeia Vol III. London: The Pharmaceutical Press; 1980
  • 4 DIN 6854. Technetium Generatoren. Anforderungen und Betrieb, Normenausschuß, Radiologie (NAR) im DIN Deutsches Institut für Normung e.V. Berlin: Beuth; 1985
  • 5 European Pharmacopoeia, 2nd ed., part II. Monograph 124, Council of Europe. Maisonneuve: Sainte-Ruffine; 1982
  • 6 Feigl F. Spot tests in inorganic analysis. Amsterdam: Elsevier; 1958
  • 7 Filthuth H. Radioscanning of TLC. Advances in thin-layer chromatography. New York: J Wiley; 1982
  • 8 Hammermaier A, Reich E, Bögl K W. Comparison of different thin-layer detection techniques to determine the radiochemical purity of radiopharmaceuticals. NUC-Comp 1985; 16: 200-6.
  • 9 Hammermaier A, Reich E, Bögl K W. Investigations on the quality of eluates from seven fission 99Mo/99mTc generators. Institut für Strahlenhygiene des Bundesgesundheitsamtes. Neuherberg. ISH-Heft 64: 1985.
  • 10 Hammermaier A, Reich E Bögl K W. Qualitätskontrolle von in-vivo Radiopharmaka - Ein Handbuch für das nuklearmedizinische Labor zur Analytik radiochemischer Verunreinigungen. Institut für Strahlenhygicne des Bundesgesundheitsamtes. Neuherberg. ISH-Hcft 94: 1986.
  • 11 Hammermaier A, Reich E, Bögl K W. Vergleichende Untersuchung von verschiedenen Meßverfahren an Radiochromatogrammen zur Beurteilung der radiochemisehen Reinheit von zehn Radiopharmaka. Institut für Strahlenhygicne des Bundesgesundheitsamtes. Neuherberg. ISH-Heft 36: 1984.
  • 12 Kocher D. Radioactive decay data tables. A handbook for application to radiation dosimetry and radiological assessments. Technical Information Center. U.S. Department of Energy. 1981
  • 13 Lederer C M, Shirley V S. Tables of isotopes. 7th ed. New York: Wiley; 1978
  • 14 McAuley R, Lull R, Ice R. Bromine-82 contamination in fission product99mTc generator elute. Eur J Nucl Med 1985; 10: 60-2.
  • 15 Molter M. Prinzipielle Probleme bei der Verwendung von 99mTc Generatoren. NUC- Comp 1979; 10: 30-2.
  • 16 Näsman P, Väyrynen T. Impurities of 99mTc generators. Eur J Nucl Med 1983; 08: 26-9.
  • 17 Pfeiffer G. Quality control of radiopharmaceulicals. Nucl-Mcd 1984; 23: 4-8.
  • 18 Robbins P J. Chromatography of tcchnetium-99m radiopharmaceuticals - A practical guide. New York: The Society of Nuclear Medicine; 1984
  • 19 Shukla S K, Manni G B, Cipriani C. Effect of aluminium impurities in the generator- produced pertechnetate-99m ion on thyroid scintigrams. Eur J Nucl Med 1977; 02: 137-41.
  • 20 The United States Pharmacopeia. 20th revision. Easton: Mack Publishing Co; 1980
  • 21 Väyrynen T, Näsman P, Kiviniitty K. Residual activity of Tc-generators. Eur J Nucl Med 1981; 06: 269-71.
  • 22 Vinberg N, Kristensen K. Fission 99Mo /99mTc generators. A study of their performance and quality. Eur J Nucl Med 1980; 05: 435-8.
  • 23 Weinstein M B, Smoak W M. Technical difficulties in 99mTc labelling of erythrocytes. J Nucl Med 1970; 11: 41-2.
  • 24 Williams C, Kerciakes J G, Grossman L W. The accuracy of molybdenum-99 assays in tcchnctium-99m solutions. Radiology 1981; 138: 445-8.