Methods Inf Med 2011; 50(03): 253-264
DOI: 10.3414/ME0616
Original Article
Schattauer GmbH

Feasibility Evaluation of Smart Stretcher to Improve Patient Safety during Transfers

K. Ohashi
1   Division of General Medicine, Brigham and Women’s Hospital, Boston, MA, USA
5   Department of Bioinformatics, Graduate School of Biomedical Science, Tokyo Medical and Dental University, Bunkyo-ku, Tokyo, Japan
,
Y. Kurihara
2   Department of Computer and Information Science, Faculty of Science and Technology, Seikei University, Musashino-shi Tokyo, Japan
,
K. Watanabe
3   Faculty of Engineering, Hosei University, Musashino-shi, Tokyo, Japan
,
L. Ohno-Machado
4   Division of Biomedical Informatics, University of California, San Diego, CA, USA
,
H. Tanaka
5   Department of Bioinformatics, Graduate School of Biomedical Science, Tokyo Medical and Dental University, Bunkyo-ku, Tokyo, Japan
› Institutsangaben
Weitere Informationen

Publikationsverlauf

received: 17. November 2008

accepted: 22. März 2010

Publikationsdatum:
18. Januar 2018 (online)

Summary

Objectives: The integration of noninvasive vital sign sensors and wireless sensor networks into intelligent alarm systems has the potential to improve patient safety. We developed a wireless network-based system (“Smart Stretcher”), which was designed to constantly monitor patient vital signs and detect apnea during transfers within a hospital. The system alerts medical staff in case of an emergency through a wireless network.

Methods: A small-scale technical feasibility study was conducted to assess the performance of the system in a simulated hospital environment. Smart Stretcher consists of three components: a small air-mat type pressure sensor measuring respiratory rate and detecting apnea, a patient identification system using RFID technology, and an indoor positioning system using a ZigBee wireless network. In the feasibility experiment, two nurses transferred four subjects who stopped breathing for 10 seconds, after which we calculated the accuracy of apnea detections, repeating this at varying speeds and subject positions. We alsoperformed asubjective evaluation of perceptions and expectations of Smart Stretcher by nurses.

Results: The system could detect apnea in all subjects at a rate of over 90%, patient IDs and locations were correctly detected in real time, and the system could alert medical staff. In addition, the results of nurse’s evaluations were mostly positive.

Conclusions: The technical feasibility experiment and evaluation of Smart Stretcher suggest that the system could play a key role in monitoring patients during hospital transfers.

 
  • References

  • 1 Wicks AM, Visich JK, Li S. Radio frequency identification applications in hospital environments. Hosp Top 2006; 84 (03) 3-8.
  • 2 Booth P, Frisch PH, Miodownik S. Application of RFID in an integrated healthcare environment. Conf Proc IEEE Eng Med Biol Soc 2006; 1: 117-119.
  • 3 Fisher JA, Monahan T. Tracking the social dimensions of RFID systems in hospitals. Int J Med Inform 2008; 77 (03) 176-183.
  • 4 Sutherland JV, van den Heuvel WJ, Ganous T, Burton MM, Kumar A. Towards an intelligent hospital environment: OR of the future. Stud Health Technol Inform 2005; 118: 278-312.
  • 5 Lin BS, Chou NK, Chong FC, Chen SJ. RTWPMS: a real-time wireless physiological monitoring system. IEEE Trans Inf Technol Biomed 2006; 10 (04) 647-656.
  • 6 Yuce MR, Ng PC, Lee CK, Khan JY, Liu W. A wireless medical monitoring over a heterogeneous sensor network. Conf Proc IEEE Eng Med Biol Soc 2007 pp 5895-5899.
  • 7 Fry EA, Lenert LA. MASCAL: RFID tracking of patients, staff and equipment to enhance hospital response to mass casualty events. AMIA Annu Symp Proc 2005 pp 261-265.
  • 8 Broom MA, Slater J, Ure DS. An observational study of practice during transfer of patients from anaesthetic room to operating theatre. Anaesthesia 2006; 61 (10) 943-945.
  • 9 Kikuchi A, Sakaue G, Abe T, Kyakuno M, Nosaka S, Maeda S. Examination of criminal suits on patients and operation sites. The Journal of Japan Society for Clinical Anesthesia 2004; 24 (09) 562-567.
  • 10 Makino Y, Kikuchi Matsumura Y, Kyakuno M, Ebara K, Nosaka S. Status of Inadequate Postoperative Respiratory Management in Legal Cases. The Journal of Japan Society for Clinical Anesthesia 2008; 28 (07) 986-992.
  • 11 Sandlin D. SurgiChip – new technology for prevention of wrong site, wrong procedure, wrong person surgery. J Perianesth Nurs 2005; 20 (02) 144-146.
  • 12 Greenberg CC, Regenbogen SE, Studdert DM, Lipsitz SR, Rogers SO, Zinner MJ. et al. Patterns of communication breakdowns resulting in injury to surgical patients. J Am Coll Surg 2007; 204 (04) 533-540.
  • 13 Michaels RK, Makary MA, Dahab Y, Frassica FJ, Heitmiller E, Rowen LC. et al. Achieving the National Quality Forum’s “Never Events”: prevention of wrong site, wrong procedure, and wrong patient operations. Ann Surg 2007; 245 (04) 526-532.
  • 14 Clarke JR, Johnston J, Finley ED. Getting surgery right. Ann Surg 2007; 246 (03) 395-403.
  • 15 Ohashi K, Kurihara Y, Watanabe K, and Tanaka H. Safe patient transfer system with monitoring of location and vital signs. J Med Dent Sci 2008; 55: 33-41.
  • 16 Jepico Cooperation.. Available at: http://www.jepico.co.jp/english/index.html. Accessed March 12 2009
  • 17 Watanabe T, Watanabe K. Noncontact method for sleep stage estimation. IEEE Trans Biomed Eng 2004; 51 (10) 1735-1748.
  • 18 Watanabe K, Watanabe T, Watanabe H, Ando H, Ishikawa T. Noninvasive Measurement of Heartbeat, Respiration, Snoring and Body Movements of a Subject in Bed via a Pneumatic Method. IEEE Trans Inf Technol Biomed 2005; 52 (12) 2100-2107.
  • 19 Kurihara Y, Matsushima M, Watanabe K, Ohashi K, Tanaka H. Respiratory Arrest Detection of a Patient Transferred by a Stretcher. Frontiers of Medical and Biological Engineering 2007; 45 (03) 208-215.
  • 20 Kurihara Y, Watanabe K, Kobayashi K, Tanaka H. Unconstrained Respiration Measurement and Respiratory Arrest Detection Method by Dynamic Threshold in Transferring Patients by Stretchers. IEEJ Transactions on Electronics, Information and Systems 2008; 128 (11) 1641-1648.
  • 21 Zigbee alliance.. Wireless control that simply works. Available at: http://www.zigbee.org. Accessed April 25 2008
  • 22 ZigBee sensor node.. Airsense, Hitach ltd. Available at: http://www.hitachi.com. Accessed March 14 2009
  • 23 RFID reader, Takaya Cooperation.. Available at: http://www.takaya.co.jp. Accessed March 12 2009
  • 24 Lee MJ, Zheng J, Ko YB, Shrestha DM. Emerging standards for wireless mesh technology. IEEE Wirel Commun 2006; 13 (02) 56-63.
  • 25 Neil R. On a roll. RFID moves toward patient safety. Mater Manag Health Care 2005; 14 (03) 20-23.
  • 26 Macario A, Morris D, Morris S, Weed L. Initial clinical evaluation of a handheld device for detecting retained surgical gauze sponges using radiofrequency identification technology. Arch Surg 2006; 141 (07) 659-662.
  • 27 Chen CI, Liu CY, Li YC, Chao CC, Liu CT, Chen CF. et al. Pervasive Observation Medicine: The Application of RFID to Improve Patient Safety in Observation Unit of Hospital Emergency Department. Stud Health Technol Inform 2005; 116: 311-315.
  • 28 Meyer M, Fairbrother P, Egan M, Chueh H, Sand-berg WS. Using location tracking data to assess efficiency in established clinical workflows. AMIA Annu Symp Proc 2006 p 1031
  • 29 Clarke M, Bratan T, Kulkarni S, Jones R. The impact of remote patient monitoring in managing silent myocardial infarction in a residential home setting. Anadolu Kardiyol Derg 2007; 7 (01) 186-188.
  • 30 Sum KW, Zheng YP, Mak AF. Vital sign monitoring for elderly at home: development of a compound sensor for pulse rate and motion. Stud Health Technol Inform 2005; 117: 43-50.
  • 31 Lin YH, Jan IC, Ko PC, Chen YY, Wong JM, Jan GJ. A wireless PDA-based physiological monitoring system for patient transport. IEEE Trans Inf Technol Biomed 2004; 8 (04) 439-447.
  • 32 Tia Gao, Tammara Massey, Leo Selavo, David Crawford, Bor-rong Chen, Konrad Lorincz, et al. The Advanced Health and Disaster Aid Network: A Light-weight Wireless Medical System for Triage. IEEE Trans Cir and Sys 2007; 1 (03) 203-216.
  • 33 Curtis DW, Pino EJ, Bailey JM, Shih EI, Waterman J, Vinterbo SA. et al. SMART – an integrated wireless system for monitoring unattended patients. JAMIA 2008; 15 (01) 44-53.