Retrospective Comparison of the Myopia Master and the Lenstar LS900 Axial Length Measurements in Children with Myopia
Retrospektiver Vergleich der Axiallängenmessungen mit Myopia Master und Lenstar LS900 bei Kindern mit MyopieAbstract
Background This study is a retrospective analysis to compare ocular biometry measurements of axial length in children with myopia using Myopia Master (OCULUS, Wetzlar, Germany) and Lenstar LS900 (HAAG-STREIT AG, Köniz, Switzerland).
Patients and Methods Axial length measurements obtained with both instruments within a 3-week period were collected retrospectively. Measurements were visualized with a Bland-Altman plot. For statistical evaluation, a paired t-test was applied, and the Pearson correlation coefficient (r) was established.
Results Sixty-one eyes from 31 myopic patients (59% male, 41% female) with a mean age of 11.34 ± 3.25 years (range: 6 – 18 years) were identified. Mean axial length was 24.7 mm (SD 1.29) with the Myopia Master and 24.69 mm (SD 1.30) with the Lenstar LS900 (Pearson correlation: r = 0.9991). The average difference of the axial length measurement between the two biometers was 0.00 064 mm ± 0.056 SD (p = 0.9293).
Conclusion The axial length measured by Myopia Master and Lenstar LS900 did not differ significantly. Thus, previous values obtained with the Lenstar LS900 can be applied to assess myopia progression.
Zusammenfassung
Hintergrund Diese Studie vergleicht die Axiallängenmessungen zweier Biometriegeräte, Myopia Master (OCULUS, Wetzlar, Deutschland) und Lenstar LS900 (HAAG-STREIT AG, Köniz, Schweiz), die bei Kindern mit Myopie erhoben wurden.
Patienten und Methoden Retrospektiv wurden Axiallängenmessungen, die jeweils mit den beiden Geräten innerhalb von 3 Wochen ermittelt worden sind, gesammelt. Zur optischen Beurteilung der Messdaten wurde ein Bland-Altman-Diagramm erstellt. Die Differenz der beiden Messungen wurde mittels t-Differenztest verglichen und der lineare Zusammenhang anhand des Pearson Korrelationskoeffizient (r) bestimmt.
Ergebnisse Insgesamt konnten 61 Augen von 31 myopen Patienten (59% männlich, 41% weiblich) mit einem Durchschnittsalter von 11.34 ± 3.25 Jahren (Spannweite: 6 – 18 Jahre) in die Studie eingeschlossen werden. Der Durchschnitt der Axiallängen ergab 24.7 mm (SD 1.29) mit dem Myopia Master und 24.69 mm (SD 1.30) mit dem Lenstar LS900 (Pearson Korrelationskoeffizient: r = 0.9991). Die mittlere Differenz der Axiallängenmessungen zwischen den beiden Biometriegeräten betrug 0.00 064 mm ± 0.056 SD (p = 0.9293).
Schlussfolgerung Die Axiallängen, die mit dem Myopia Master und dem Lenstar LS900 gemessen wurden, unterscheiden sich statistisch nicht signifikant. Demzufolge können zuvor mit dem Lenstar LS900 erhobene Messungen für die Berechnung der Myopie Progression hinzugezogen werden.
Key words
Myopia Master - Lenstar LS900 - axial length - axial length measurement - myopia - myopia in childrenSchlüsselwörter
Myopia Master - Lenstar LS900 - Axiallänge - Axiallängenmessung - Myopie - Myopie bei KindernPublication History
Received: 16 October 2022
Accepted: 12 January 2023
Article published online:
25 April 2023
© 2023. Thieme. All rights reserved.
Georg Thieme Verlag KG
Rüdigerstraße 14, 70469 Stuttgart, Germany
-
References
- 1 Flitcroft DI. The complex interactions of retinal, optical and environmental factors in myopia aetiology. Prog Retin Eye Res 2012; 31: 622-660 DOI: 10.1016/j.preteyeres.2012.06.004.
- 2 Tideman JW, Snabel MC, Tedja MS. et al. Association of Axial Length With Risk of Uncorrectable Visual Impairment for Europeans With Myopia. JAMA Ophthalmol 2016; 134: 1355-1363 DOI: 10.1001/jamaophthalmol.2016.4009.
- 3 Huang J, McAlinden C, Huang Y. et al. Meta-analysis of optical low-coherence reflectometry versus partial coherence interferometry biometry. Sci Rep 2017; 7: 43414 DOI: 10.1038/srep43414.
- 4 Montés-Micó R, Pastor-Pascual F, Ruiz-Mesa R. et al. Ocular biometry with swept-source optical coherence tomography. J Cataract Refract Surg 2021; 47: 802-814 DOI: 10.1097/j.jcrs.0000000000000551.
- 5 Huang J, Chen H, Li Y. et al. Comprehensive Comparison of Axial Length Measurement With Three Swept-Source OCT-Based Biometers and Partial Coherence Interferometry. J Refract Surg 2019; 35: 115-120 DOI: 10.3928/1081597x-20190109-01.
- 6 Haag-Streit AG. INSTRUCTIONS FOR USE Biometer Lenstar LS900® . Accessed February 28, 2023 at: https://www.manualslib.com/manual/1202086/Lenstar-Ls-900.html#manual
- 7 Ramdass S. http://
- 8 Ye Y, Zhao Y, Han T. et al. Accuracy of axial length, keratometry, and refractive measurement with Myopia Master in children with ametropia. BMC Ophthalmol 2022; 22: 468 DOI: 10.1186/s12886-022-02672-9.
- 9 Sel S, Stange J, Kaiser D. et al. Repeatability and agreement of Scheimpflug-based and swept-source optical biometry measurements. Cont Lens Anterior Eye 2017; 40: 318-322 DOI: 10.1016/j.clae.2017.03.007.