Semin Thromb Hemost 2008; 34(7): 593-603
DOI: 10.1055/s-0028-1104538
© Thieme Medical Publishers

Standardization of the INR: How Good Is Your Laboratory's INR and Can It Be Improved?

Emmanuel J. Favaloro1 , Dorothy M. Adcock2
  • 1Department of Haematology, ICPMR, Westmead Hospital, Westmead, Australia
  • 2Medical/Laboratory Director, Esoterix Coagulation, Englewood, Colorado
Further Information

Publication History

Publication Date:
15 December 2008 (online)

ABSTRACT

The prothrombin time (PT) assay is the most clinically ordered coagulation test and is most often used for monitoring of vitamin K antagonist (VKA) therapy (e.g., warfarin), where results are expressed as an international normalized ratio (INR). The INR is in essence the patient's PT “mathematically adjusted” to a standardized value by taking into account the peculiarities of the test system through applying two “correction factors” defined by an international sensitivity index (ISI) and mean normal prothrombin time (MNPT). Although some manufacturers provide assigned ISI values for specific PT reagents and instrumentation, it is still recommended practice that laboratories check or locally validate these ISIs, as well as estimate the MNPT based on the population being tested. Where a manufacturer does not provide an ISI, the laboratory needs to define its own (local ISI) value. Current recommendations suggest the use of commercial reference-plasma calibration sets, but there is limited information to validate their performance in laboratory practice. We report our personal experience with use of some of these materials, as well as review alternate or supplementary procedures for calibration and/or validation of ISI and for determination and validation of MNPT. In brief, our data and experience suggests that further verification checks should be performed prior to acceptance of ISI and MNPT estimates generated from commercial reference-plasma calibration sets. We detail various strategies to ensure that laboratory practices are optimized to provide INRs that accurately reflect a patient's true anticoagulant status and to thus assist their clinical therapeutic management.

REFERENCES

  • 1 Gallus A S. Management options for thrombophilias.  Semin Thromb Hemost. 2005;  31 118-126
  • 2 Harenberg J, Wehling M. Current and future prospects for anticoagulant therapy: inhibitors of factor Xa and factor IIa.  Semin Thromb Hemost. 2008;  34 39-57
  • 3 Medicare Australia .Australian Medicare statistics. Available at: http://www.medicareaustralia.gov.au/statistics/dyn_mbs/forms/mbs_tab4.shtml Accessed May 4, 2008
  • 4 Australian Bureau of Statistics .Australian population statistics. Available at: http://www.abs.gov.au/ Accessed May 4, 2008
  • 5 Kitchen S, Preston F E. Standardisation of prothrombin time for laboratory control of oral anticoagulant therapy.  Semin Thromb Hemost. 1999;  25 17-25
  • 6 Guidelines on oral anticoagulation: third edition.  Br J Haematol. 1998;  101 374-387
  • 7 Baglin T P, Keeling D M, Watson H G. British Committee for Standards in Haematology . Guidelines on oral anticoagulation (warfarin): third edition – 2005 update.  Br J Haematol. 2006;  132 277-285
  • 8 Adcock D M, Brien W F, Duff S L et al.. Procedures for Validation of INR and Local Calibration of PT/INR Systems; Approved Guideline. Nt H54-A, Vol. 25, No. 23 (replaces H54-P, Vol. 24, No. 30). Wayne, PA; Clinical and Laboratory Standards Institute 2005
  • 9 Adcock D M, Duff S. Enhanced standardization of the international normalized ratio through the use of plasma calibrants: a concise review.  Blood Coagul Fibrinolysis. 2000;  11 583-590
  • 10 McGlasson D L. A comparison of INRs after local calibration of thromboplastin international sensitivity indexes.  Clin Lab Sci. 2002;  15 91-95
  • 11 Kearon C, Kahn S R, Agnelli C, Goldhaber S, Rascob G, Comerota A J. Antithrombotic therapy for venous thromboembolic disease: American College of Chest Physicians Evidence-Based Practice Guidelines; 8th edition.  Chest. 2008;  133 454S-545S
  • 12 Stevenson K J, Craig S, Dufty J MK, Taberber D A. System ISI calibration: a universally acceptable scheme is possible only when coumarin plasma calibrants are used.  Br J Haematol. 1997;  96 435-441
  • 13 Expert WHO Committee on Biological Standardization .Requirements for Thromboplastins and Plasmas Used to Control Oral Anticoagulant Therapy. 33rd report, WHO Technical Report Series. Geneva, Switzerland; World Health Organization 1983 687: 81-105
  • 14 Adcock D M, Johnston M. Evaluation of frozen plasma calibrants for enhanced standardization of the international normalized ratio (INR): a multicenter study.  Thromb Haemost. 2002;  87 74-79
  • 15 Fareed J, Hoppensteadt D A, Fareed D et al.. Survival of heparins, oral anticoagulants, and aspirin after the year 2010.  Semin Thromb Hemost. 2008;  34 58-73
  • 16 Lippi G, Franchini M, Montagnana M, Guidi G C. Genomics and proteomics in venous thromboembolism: building a bridge toward a rational personalized medicine framework.  Semin Thromb Hemost. 2007;  33 759-770
  • 17 Kelton J G, Hirsh J. Bleeding associated with antithrombotic therapy.  Semin Hematol. 1980;  17 259-291
  • 18 Hirsh J. Is the dose of warfarin prescribed by American physicians unnecessarily high?.  Arch Intern Med. 1987;  147 769-771
  • 19 Poller L. International normalized ratio (INR): the first 20 years.  J Thromb Haemost. 2004;  2 849-860
  • 20 Favaloro E J, Lippi G, Adcock D. Preanalytical and postanalytical variables: the leading causes of diagnostic error in hemostasis?.  Semin Thromb Hemost. 2008;  34 612-634
  • 21 Favaloro E J, Hamdam S, McDonald J, McVicker W, Ule V. Time to think outside the box? Prothrombin time (PT), international normalised ratio (INR), international sensitivity index (ISI), mean normal prothrombin time (MNPT) and measurement of uncertainty (MU): a novel approach to standardisation.  Pathology. 2008;  40 277-287
  • 22 Poller L, Thompson J M, Taberner D A, Clarke D K. The correction of coagulometer effects on the international normalized ratio.  Br J Haematol. 1994;  86 112-117
  • 23 Ray M J, Smith I R. The dependence of the international sensitivity index on the coagulometer used to perform the prothrombin time.  Thromb Haemost. 1990;  63 424-429
  • 24 Poller L, Triplett D A, Hirsh J, Carroll J, Clarke K. A comparison of lyophilized artificially depleted plasmas and lyophilized plasmas from patients receiving warfarin in correcting for coagulometer effects on international normalized ratios.  Am J Clin Pathol. 1995;  103 366-371
  • 25 van Rijn J L, Schmidt N A, Rutten W P. Correction of instrument and reagent based differences in determination of the international normalized ratio (INR) for monitoring anticoagulant therapy.  Clin Chem. 1989;  35 840-843
  • 26 Ebert R F. PTs, ISIs and INRs: a primer on prothrombin time reporting.  Clin Hemost Rev. 1994;  7 1-20
  • 27 World Health Organization .Annex 3 – Guidelines for Thromboplastins and Plasma Used to Control Anticoagulant Therapy. WHO Technical Report Series No. 889. Geneva, Switzerland; World Health Organization 1999: 64-93
  • 28 Loeliger E A, van den Besselar A MPH, Lewis S M. Reliability and clinical impact of the normalization of the prothrombin times in oral anticoagulant control.  Thromb Haemost. 1985;  53 148-154
  • 29 Cunningham M T, Johnson G F, Pennel B J, Olsen J D. The reliability of manufacturer-determined instrument-specific international sensitivity values for calculating the international normalized ratio.  Am J Clin Pathol. 1994;  102 128-133
  • 30 Ng V L, Levin J, Corash L, Gottfried E L. Failure of the international normalized ratio to generate consistent results within a local medical community.  Am J Clin Pathol. 1993;  99 689-694
  • 31 Clarke K, Taberner D A, Thomsom J M, Morris J A, Poller L. Assessment of the value of calibrated lyophilized plasmas to determine international sensitivity index for coagulometers.  J Clin Pathol. 1992;  45 58-60
  • 32 Fairweather R B, Ansell J, van den Besselaar A MHP et al.. College of American Pathologists Conference XXXI on laboratory monitoring of oral anticoagulant therapy.  Arch Pathol Lab Med. 1998;  122 768-781
  • 33 Barowcliffe T W, ven den Besselaar A MHP et al. Guidelines for Preparation, Calibration and Use of Calibrated Plasmas for ISI and INR Determination. ISTH Subcommittee on Control of Anticoagulation Maastricht, The Netherlands; International Society of Thrombosis and Hemostasis 2000
  • 34 Johnston M, Brigden M. A cross–Canada survey of prothrombin time testing. Does the establishment of local ISI values improve the accuracy of international normalized ratio reporting?.  Am J Clin Pathol. 1998;  110 683-690
  • 35 Brien W F, Baskerville J C, Taberner D A, Crawford L. Calculation vs. calibration curve for INR determination. Results of an interlaboratory proficiency scheme.  Am J Clin Pathol. 1999;  111 193-201
  • 36 Chantarangkul V, Tripodi A, Cesana B M, Manucci P M. Calibration of local systems with lyophilized calibrant plasmas improves interlaboratory variability of the INR in the Italian external quality assessment scheme.  Thromb Haemost. 1999;  82 1621-1626
  • 37 van den Besselaar A M. Comparison of lyophilized plasmas with fresh plasmas for calibration of thromboplastin reagents in oral anticoagulant control.  Br J Haematol. 1996;  93 437-444
  • 38 Dufty J MK, Craig S, Stevenson K J, Taberner D A. Calculation of system international sensitivity index: how many calibrant plasmas are required?.  J Clin Pathol. 1997;  50 40-44
  • 39 Hubbard A R, Margetts S MA, Barrowcliffe T W. International normalized ratio determination using calibrated reference plasmas.  Br J Haematol. 1997;  98 74-78
  • 40 Duncan E M, Casey C R, Duncan B M, Lloyd J V. Effect of concentration trisodium citrate on calculation of the international normalized ratio and the international sensitivity index.  Thromb Haemost. 1994;  72 84-88
  • 41 Arkin C F. Do we need the geometric mean as the mean normal value in INR determination?.  J Thromb Hemost. 2005;  3(Suppl 1) , abstract 1903
  • 42 Marlar R A. Determining the model for end-stage liver disease with better accuracy: neutralizing the international normalized ratio pitfalls.  Hepatology. 2007;  46 295-296
  • 43 Tripodi A, Chantarangkul V, Primignani M et al.. The international normalized ratio calibrated for cirrhosis (INR(liver)) normalizes prothrombin time results for model for end-stage liver disease calculation.  Hepatology. 2007;  46 520-527
  • 44 Bellest L, Eschwège V, Poupon R, Chazouillères O, Robert A. A modified international normalized ratio as an effective way of prothrombin time standardization in hepatology.  Hepatology. 2007;  46 528-534
  • 45 Tripodi A. How to implement the modified international normalized ratio for cirrhosis (INR(liver)) for model for end-stage liver disease calculation.  Hepatology. 2008;  47 1423-1424

Emmanuel J FavaloroPh.D. M.A.I.M.S. 

Department of Haematology, Institute of Clinical Pathology and Medical Research (ICPMR)

Westmead Hospital, SWAHS, Westmead, NSW 2145, Australia

Email: emmanuel.favaloro@swahs.health.nsw.gov.au