CC BY 4.0 · Aorta (Stamford) 2013; 01(01): 71-77
DOI: 10.12945/j.aorta.2013.13.005
New Technology
Thieme Medical Publishers 333 Seventh Avenue, New York, NY 10001, USA.

A Circular Surgical Stapler Designed to Anastomose Aorta and Dacron Tube Graft

Validation of the Concept and Comparison to Hand-Sewn Anastomosis in Bench Experiments
Syed T. Raza
1   Division of Cardiothoracic Surgery, Columbia University Medical Center and the New York Presbyterian Hospital, New York, New York
› Institutsangaben
Weitere Informationen

Publikationsverlauf

16. Januar 2013

13. Februar 2013

Publikationsdatum:
28. September 2018 (online)

Abstract

Background: A circular aortic stapler has been developed to anastomose the open end of the aorta to a size-matched Dacron tube graft in one quick motion and without having to pull sutures through the aortic wall.

Methods: A prototype was developed, and its design and function were tested in bench experiments and compared with hand-sewn anastomosis. The basic design of the stapler is a central rod (anvil) surrounded by 10 stapling limbs, which can be closed over the anvil in a full circle, with staples extruded by turning a knob at the back. To test its function, a Dacron tube graft was inserted in the middle of a length of bovine aorta. One side was anastomosed with the stapler and the other hand-sewn in each of 10 experiments. Bovine blood was infused under increasing pressure.

Results: It took considerably less time to complete the stapled anastomosis than the hand-sewn side (3 minutes, 46 seconds versus 15 minutes, 42 seconds). Initial leak occurred at low pressures on the hand-sewn side (mean pressure 40 mm Hg) compared with the stapled side (mean pressure 70 mm Hg). In 7 of 10 experiments, the leak became too brisk on the hand-sewn side to sustain pressure, compared with 3 of 10 with stapled anastomoses. The stapling device performed well in all cases except when the bovine aorta was too thick for the staples (two cases) or when there was a missed branch at the anastomotic site (one case).

Conclusions: These experiments validate the concept and the design of this aortic stapler. There are some limitations in the current design, which will need to be modified before its use in live animals or clinically.

 
  • References

  • 1 Williams JB, Peterson ED, Zhao Y, O'Brien SM, Anderson ND, Miller DC. , et al. Contemporary results for proximal aortic replacement in North America. J Am Coll Cardiol 2012; 60 (13) 1156-1162
  • 2 Trimarchi S, Nienaber CA, Rampoldi V, Myrmel T, Suzuki T, Mehta RH. , et al. Contemporary results of surgery in acute type A aortic dissections: The International Registry of Acute Aortic Dissection experience. J Thorac Cardiovasc Surg 2005; 129 (01) 112-122
  • 3 Hoefer D, Ruttman E, Riha M, Shobersberger W, Mayr A, Laufer G, Bonatti J. Factors influencing intensive care unit length of stay after surgery for acute aortic dissection type A. Ann Thorac Surg 2002; 73 (03) 714-718
  • 4 Davies RR, Coe MP, Mandapati D, Gallo A, Botta DM, Elefteriades JA, Coady MA. What is the optimal management of late-presenting survivors of acute type A aortic dissection?. Ann Thorac Surg 2007; 83 (05) 1593-1602
  • 5 Svensson LG, Crawford ES, Hess KR, Coselli JS, Raskin S, Shenaq SA, Safi HJ. Deep hypothermia with circulatory arrest: Determinants of stroke and early mortality in 656 patients. J Thorac Cardiovasc Surg 1993; 106 (01) 19-28